Elekta Versa HD

Workstation Freezes or Loses Communication

On this page

Asset Type

Radiation Therapy System

Manufacturer

Elekta

Model

Versa HD

What This Guide Helps With

A Versa HD workstation freezes, becomes unresponsive, or loses communication because of power, cables, network connectivity, peripheral, software, or infrastructure conditions.

Step-by-Step Troubleshooting

1. Protect the Patient and Treatment Workflow

Do not continue treatment when the workstation or communication path required for safe treatment delivery is frozen or unreliable.

Follow departmental procedures to place the patient and equipment in a safe state.

Expected outcome: No clinical treatment depends on an unreliable workstation or communication link.

2. Identify the Affected Workstation or Link

Determine which workstation is affected and whether the problem is a frozen user interface, loss of communication with the treatment system, network disconnection, or failure of a connected peripheral.

Record displayed messages and the last successful action.

Expected outcome: The failure is narrowed to a workstation, network path, peripheral, or wider system issue.

3. Check Basic Power and Indicators

Verify that the workstation, monitors, network devices accessible to Clinical Engineering, and required peripherals have power.

Inspect for abnormal power indicators or unexpected shutdowns.

Expected outcome: All required external components are powered normally.

4. Check Keyboard, Mouse, and Display Operation

Confirm that the issue is not limited to a failed input device or display.

Use a known-good compatible peripheral if approved and readily available.

Expected outcome: A peripheral problem is distinguished from an actual workstation freeze.

5. Inspect Network and Data Connections

Check accessible network and system communication cables for loose connectors, damage, disconnected patch cords, or disturbed cable routing.

Verify link indicators where they are normally visible.

Expected outcome: External communication connections are intact.

6. Determine Whether the Problem Is Local or Widespread

Check whether other Versa HD workstations or connected systems can communicate normally.

If multiple systems are affected, investigate infrastructure availability before assuming a local workstation problem.

Expected outcome: The fault is localized to one workstation or identified as an infrastructure-level event.

7. Check for Recent Infrastructure Changes

Determine whether network maintenance, switch work, power interruption, software deployment, or other system changes occurred near the time of failure.

Expected outcome: Any external change potentially related to the outage is identified and routed to the appropriate support group.

8. Perform an Approved Restart

If the patient is safe and site procedures permit, perform a normal restart of the affected workstation or software environment.

Do not repeatedly hard-cycle power or restart active treatment-system components without an approved process.

Expected outcome: The workstation restarts, reconnects, and remains responsive.

9. Verify Communication and Workflow

Confirm that the workstation communicates with required treatment, imaging, record, or network systems and that the affected workflow can be completed using an appropriate nonclinical verification.

Expected outcome: Communication remains stable and the required functions work normally. Troubleshooting can stop after applicable verification.

10. Escalate Persistent Communication Failure

If the workstation continues to freeze or lose communication after power, peripherals, network connections, infrastructure, and approved restart checks, stop troubleshooting.

Expected outcome: The affected system is withheld from clinical use until service-level evaluation is completed.

If the Problem Persists

External causes have been ruled out. Remaining possibilities may include workstation hardware, operating-system or application issues, internal treatment-system communications, network configuration, server availability, interface services, or other service-level conditions.

The Versa HD should be:

Knowing when to stop external troubleshooting and escalate is proper troubleshooting.

Clinical Use Tip

Verify the complete treatment communication path after a workstation recovery; a responsive screen alone does not prove that treatment-system communications are restored.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Radiation Oncology reported that a Versa HD workstation became unresponsive and stopped communicating with the treatment system."

Cause

What was observed during troubleshooting.

Example:
"Clinical Engineering found a loose external network connection at the affected workstation."

Resolution

What action was taken.

Example:
"Clinical Engineering secured the connection, restarted the workstation using the approved process, and verified stable communication and normal workflow before return to service."

Helpful Details to Include (If Known)

Final Thought

Separate peripheral, workstation, network, and infrastructure problems methodically. Confirm the entire communication path before return to service and escalate persistent freezes rather than repeatedly restarting an unreliable system.

That is successful troubleshooting.