Enthermics DC Series (Blanket Warmer)

Erratic or False “High Temp” / Overheat Alarms

On this page

Asset Type

Blanket Warmer

Manufacturer

Enthermics

Model

DC Series (Blanket Warmer)

What This Guide Helps With

This guide addresses situations where an Enthermics DC Series (Blanket Warmer) displays intermittent or persistent “High Temp” or overheat alarms, but the unit does not appear excessively hot.

The focus is on ruling out common external causes—such as airflow restriction, sensor obstruction, loading practices, and power quality—before assuming internal thermostat, control board, or sensor failure.

Step-by-Step Troubleshooting

1. Ensure the Unit Is Not in Active Patient Use

2. Verify Set Temperature vs. Displayed Temperature

3. Check for Overloading of Blankets

4. Inspect Internal Airflow Path

5. Verify Door Seal and Closure

6. Check Ambient Environment

7. Inspect Power Source

8. Observe Alarm Behavior Pattern

9. Perform Power Reset

If the Problem Persists

If:

The issue is likely internal, such as:

Do not bypass safety thermostats or override temperature protections.

Knowing when to stop is proper troubleshooting.

Clinical Use Tip

Never troubleshoot a blanket warmer while it is actively supporting patient care.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
“Unit alarming ‘High Temp’ intermittently during use. Staff reports blankets do not feel excessively hot.”

Cause

What was observed during troubleshooting.

Example:
“Airflow restriction noted due to overpacked blankets blocking rear vent channels. Chamber temperature spiking during heat cycle.”

Resolution

What action was taken.

Example:
“Removed excess blankets, cleared airflow path, verified normal heating cycle. Unit tested for 30 minutes with no further alarms. Returned to service.”

“Removed from service, labeled Out of Service, sent for bench evaluation.”

Helpful Details to Include

Final Thought

Blanket warmers are simple devices, but airflow and loading practices greatly influence temperature stability. Most erratic high-temperature alarms are caused by external factors, not internal failure.

Start with airflow, loading, and environment. Escalate only after logical external causes are ruled out. Document clearly and protect patient safety at every step.

That is successful troubleshooting.