Fresenius Kabi Agilia

Drug Library Download or Synchronization Failure

On this page

Asset Type

Infusion Pump

Manufacturer

Fresenius Kabi

Model

Agilia

What This Guide Helps With

Troubleshooting missing, outdated, incomplete, or failed Agilia drug-library updates caused by connectivity, assignment, power, server, or configuration issues.

Step-by-Step Troubleshooting

1. Ensure Patient Safety First

Do not troubleshoot, restart, reconfigure, or update an Agilia pump while it is actively delivering medication to a patient.

Expected outcome: Patient therapy continues without relying on an unverified or outdated drug library.

If the pump’s drug-library status cannot be confirmed, remove it from clinical use until the issue is resolved.

2. Clarify the Reported Failure

Determine whether:

Record any displayed message, synchronization status, library version, profile name, date, and time.

Expected outcome: The problem is narrowed to an individual pump, connection method, profile assignment, or system-wide deployment issue.

3. Confirm That the Correct Library Was Released

Contact the pharmacy, medication-safety, or drug-library administrator and verify:

Vigilant Master Med is used to create, customize, and distribute drug libraries, while Vigilant Centerium supports remote library distribution to connected Agilia pumps.

Expected outcome: The expected library is confirmed as valid, released, and assigned to the affected pump population.

If the library was not released or assigned, stop troubleshooting the pump and refer the issue to the drug-library administrator.

4. Verify the Pump’s Installed Library and Profile

With the pump removed from patient use:

Agilia configurations may support multiple customized profiles, so an incorrect profile assignment can resemble a failed library download.

Expected outcome: The installed version and selected clinical profile match the approved deployment.

If the correct library is present but the wrong profile is selected, follow the facility-approved profile-selection process and retest. Stop if resolved.

5. Check Power and Allow the Update to Complete

Confirm that:

A download may be delayed or interrupted if the pump powers down or loses communication before the transfer is complete.

Expected outcome: The pump remains continuously powered and available for the full update process.

If the library completes after stable power is restored, verify the version and return the device according to facility policy.

6. Inspect the Wireless or Rack Connection

For wireless Agilia Connect installations:

Wireless Agilia pumps can receive drug-library updates through the Agilia Connect system and Centerium infrastructure.

Expected outcome: The pump establishes normal communication or the failure is isolated to the pump, rack position, or network location.

If reseating the pump or moving it to a verified coverage area resolves the problem, confirm the downloaded version before returning it to service.

7. Compare With a Known-Good Pump and Location

Use a known-good Agilia pump of the same type and configuration.

Expected outcome:

8. Verify Pump Registration and Assignment

Have the appropriate system administrator confirm that:

Agilia Partner is intended for biomedical configuration and maintenance support, while Centerium manages communication between connected pumps and software applications.

Expected outcome: The pump is correctly recognized and targeted by the drug-library management system.

If an assignment or registration error is corrected, allow the pump to synchronize and verify the final library version.

9. Check for a Fleet-Wide or Server-Side Problem

If several pumps are affected:

Expected outcome: The problem is identified as a central deployment, server, network, or configuration issue rather than multiple simultaneous pump failures.

Escalate infrastructure problems to the responsible IT and Fresenius Kabi support teams.

10. Inspect a Direct Cable Connection When Used

For facilities that update standalone pumps through Vigilant Drug’Lib:

Fresenius Kabi documentation identifies approved Agilia USB and RS-232 cables for standalone pump transfers and notes that the USB driver must be installed before connection.

Expected outcome: The pump appears as a connected device and the approved library transfer completes.

If a known-good cable and workstation communicate with other pumps but not the affected unit, remove the pump from service for evaluation.

11. Perform One Controlled Restart

When the pump is off the patient and no download is actively progressing:

Do not repeatedly cycle power because repeated interruptions may prevent a pending update from completing.

Expected outcome: The pump reconnects and receives the assigned library.

If the update completes, verify the library version, profile, date, and operational status before return to service.

12. Verify the Completed Update

After the synchronization reports completion:

Expected outcome: The correct approved library and profile are installed and the pump passes operational verification.

If the expected library cannot be verified, do not return the pump to clinical use.

If the Problem Persists

If power, profile selection, release status, device assignment, wireless coverage, rack seating, approved cables, and server availability have been checked, the problem may involve internal communication hardware, corrupted configuration, incompatible software, or a system-management fault.

The device should be:

Do not perform unapproved software loading, database modification, internal disassembly, or component-level repair.

Knowing when to stop and escalate is proper troubleshooting.

Clinical Use Tip

Never assume that a drug library is current because drug names appear on the screen. Before clinical use, verify the approved library version and profile according to facility policy.

Do not troubleshoot synchronization problems while the pump is attached to an active patient. Move the patient’s infusion to another verified device first.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Clinical staff reported that the Fresenius Kabi Agilia pump did not receive the newly released drug library and continued displaying the previous version."

Cause

What was observed during troubleshooting.

Example:
"The pump was registered in Centerium under the incorrect clinical profile group and was not targeted by the current library deployment."

Resolution

What action was taken.

Example:
"Corrected the device assignment with the system administrator, completed synchronization, verified the approved library version and profile, and returned the pump to service after functional testing."

Helpful Details to Include (If Known)

Final Thought

Drug-library failures require a structured check of patient safety, release status, pump assignment, connectivity, power, and deployment infrastructure. Confirm the approved version before clinical use, escalate internal or system-level failures appropriately, and clearly document the final device status.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide