Fresenius Kabi Agilia

Free-Flow Protection or Anti-Bolus Concern

On this page

Asset Type

Infusion Pump

Manufacturer

Fresenius Kabi

Model

Agilia

What This Guide Helps With

Troubleshooting suspected uncontrolled flow, free-flow protection failure, post-occlusion bolus, or unexpected fluid delivery caused by tubing, loading, clamps, or setup.

Step-by-Step Troubleshooting

1. Ensure Patient Safety First

Do not investigate suspected free flow or unintended bolus delivery while the pump is connected to an active patient.

Expected outcome: The patient is protected from additional uncontrolled delivery.

A pump associated with suspected free flow or unintended bolus delivery must not remain in clinical service until evaluated.

2. Clarify the Reported Behavior

Determine whether the complaint involved:

Record the medication, programmed rate, approximate unintended volume, alarm messages, and whether the line was connected to a patient.

Expected outcome: The suspected free-flow or anti-bolus event is clearly defined before testing begins.

3. Identify the Agilia Pump Type and Approved Disposable

Confirm whether the device is an Agilia volumetric pump, such as an Agilia VP or VP MC, or an Agilia syringe pump.

Free-flow protection on Agilia volumetric pumps depends on the compatible administration set and its anti-free-flow device. Fresenius Kabi describes the Agilia VP MC SafeClip as preventing free flow when the pump door is open.

Verify that:

Expected outcome: The pump is being evaluated with the correct, undamaged disposable.

If the wrong set was used, replace it with an approved set and repeat the functional check. If operation is normal, document the disposable mismatch and stop.

4. Close the Manual Clamp Before Opening the Door

Before opening the pump door:

Do not depend solely on the pump’s automatic free-flow protection while investigating a reported failure.

Expected outcome: Fluid cannot reach a patient or spill uncontrollably during set removal.

5. Inspect the Administration Set’s Free-Flow Protection

Remove the administration set and visually inspect the SafeClip or equivalent anti-free-flow component.

Check for:

Do not lubricate, bend, reshape, or repair the disposable component.

Expected outcome: The free-flow mechanism is clean, intact, and able to close the tubing when removed from the pump.

If the component is damaged or questionable, discard the set and test with a new approved set.

6. Inspect the Pump Door and Tubing Channel

With the pump removed from service and powered off, inspect externally accessible areas for:

Clean only according to the manufacturer-approved process. Do not insert tools into sensors or internal mechanisms.

Expected outcome: The door, latch, guides, and set-loading area are clean and visibly intact.

If mechanical damage is found, stop troubleshooting and send the pump for repair evaluation.

7. Load a New Approved Administration Set

Using a new compatible set:

Fresenius Kabi’s Volumat administration sets use color-coded connectors and a silicone pumping segment intended to support correct loading and pumping performance.

Expected outcome: The set loads without force, the door closes normally, and no loading alarm appears.

If the door requires excessive force or the set will not remain properly seated, remove the pump from service.

8. Perform a Controlled Free-Flow Check

Perform this test without a patient connection and in accordance with facility policy.

The set’s anti-free-flow mechanism should prevent continuous gravity flow when properly engaged.

Expected outcome: No sustained uncontrolled flow occurs with the door open.

A small isolated drop caused by tubing movement may not represent continuous free flow. Sustained flow, however, is unacceptable.

If sustained flow occurs:

9. Check Stop and Pause Behavior

Reload the set, close the door, and program a low-risk test infusion using appropriate test fluid.

Verify that:

Expected outcome: Pumping stops promptly and the line remains controlled.

If delivery continues after a stop command, remove the pump from service immediately.

10. Evaluate the Reported Anti-Bolus Concern

Anti-bolus concerns commonly involve fluid released after downstream pressure has accumulated during an occlusion.

Using an approved infusion-device analyzer or controlled test fixture:

Do not release a pressurized line toward a patient. Clamp or isolate the line before clearing an occlusion in clinical use.

Expected outcome: The pump alarms appropriately, and any post-occlusion volume remains within approved performance limits.

If an analyzer is unavailable, do not estimate anti-bolus performance by watching the drip chamber. Send the device for bench evaluation.

11. Inspect Setup Conditions That Can Mimic a Pump Failure

Review whether the reported event could have resulted from:

Expected outcome: External setup factors are either identified and corrected or ruled out.

If correcting the setup resolves the concern and the pump passes all functional tests, return it to service only according to facility policy.

12. Review Available Event Information

Record or review, when available:

Agilia Partner maintenance software can support biomedical configuration, testing, and pump log analysis where available.

Expected outcome: Available evidence either supports an external setup problem or identifies the need for technical evaluation.

Do not clear logs or alter configuration before preserving relevant event information.

If the Problem Persists

If the pump permits sustained free flow with multiple new approved sets, fails to stop delivery, does not detect an occlusion correctly, or produces excessive post-occlusion delivery, common external causes have been ruled out.

The device should be:

Retain the administration set, packaging, lot number, and event details when a disposable-device failure is suspected.

Knowing when to stop and escalate is proper troubleshooting.

Clinical Use Tip

Never open an infusion pump door or clear an occlusion while the line remains open to a patient. Clamp and isolate the line first, then transfer therapy to a verified device when necessary.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Clinical staff reported that fluid continued flowing after the Agilia pump door was opened and expressed concern about free-flow protection."

Cause

What was observed during troubleshooting.

Example:
"Testing found that the administration set’s SafeClip was damaged and did not fully occlude the tubing when removed from the pump."

Resolution

What action was taken.

Example:
"Removed the defective set, verified pump operation with two new approved sets, completed a controlled free-flow and delivery test, and returned the pump to service."

Helpful Details to Include (If Known)

Final Thought

Suspected free flow or unintended bolus delivery requires immediate attention because even a small uncontrolled volume may be clinically significant. Protect the patient first, verify the disposable and loading technique, perform controlled testing, and escalate any repeatable failure. Thorough CCR documentation supports patient-safety review and helps distinguish pump failure from administration-set or setup problems.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide