Fresenius Kabi Agilia Infusion Pump

Keypad / Button Unresponsive

On this page

Asset Type

Infusion Pump

Manufacturer

Fresenius Kabi

Model

Agilia

What This Guide Helps With

This guide assists Clinical Engineering in troubleshooting a Fresenius Kabi Agilia infusion pump when the physical keypad or buttons fail to respond. Symptoms may include:

Most causes are external and simple to verify before considering internal repair.

Step-by-Step Troubleshooting

Verify Power Source

Inspect the Keypad Surface

Restart the Pump

Check for Locked or Inactive Mode

Test with Another User/Setup

Observe for Error Codes or Alarms

If the Problem Persists

Common external causes have been ruled out; the issue is likely internal.

Reinforce: Stopping here avoids patient risk and unnecessary damage.

Clinical Use Tip

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
“Keypad unresponsive; cannot enter infusion parameters.”

Cause

What was observed during troubleshooting.

Example:
“Observed no button response; power verified; keypad free of debris; pump restarted with no effect.”

Resolution

What action was taken.

Example:
“Removed pump from service; labeled Out of Service; sent for bench evaluation.”

Helpful Details to Include

Final Thought

Clinical Engineering troubleshooting prioritizes patient safety, logical stepwise checks, and proper escalation. Documenting CCR ensures clarity and supports cost avoidance by preventing unnecessary interventions. Knowing when to stop is as important as knowing what to check.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide