Fresenius Kabi Agilia

Network or Wireless Communication Failure

On this page

Asset Type

Infusion Pump

Manufacturer

Fresenius Kabi

Model

Agilia

What This Guide Helps With

Troubleshooting missing Wi-Fi connection, failed data transfer, server communication errors, or intermittent network connectivity caused by external configuration or coverage issues.

Step-by-Step Troubleshooting

1. Ensure Patient Safety First

Do not interrupt an active infusion solely to troubleshoot a wireless or network communication problem.

Expected outcome: Patient therapy remains safe while the communication issue is evaluated separately.

Agilia wireless communication is not required for the pump to continue infusing. Pending data may be stored and retransmitted when communication becomes available.

2. Confirm the Pump Has Wireless Capability

Inspect the device identification label and configuration to verify that the affected pump is a Wi-Fi-equipped Agilia model.

Agilia configurations may include:

Do not assume every Agilia pump contains wireless hardware.

Expected outcome: The device is confirmed to be a network-capable model assigned to the facility’s wireless system.

If the pump is not a Wi-Fi model, stop troubleshooting the network connection and correct the equipment assignment or work order description.

3. Identify the Exact Communication Failure

Record what is being reported or observed:

Also record any displayed communication message or error code.

Expected outcome: The failure is narrowed to the individual pump, wireless coverage, data transfer, or server-side communication.

4. Check Whether the Problem Is Location-Specific

Move the pump, when clinically safe, to a known area with reliable Agilia wireless connectivity.

Compare its connection behavior with another known-good Agilia Wi-Fi pump in the same location.

Expected outcome: The issue is identified as either location-related or isolated to the pump.

5. Verify Normal Power and Startup

Connect the pump to a known-good AC power source and confirm:

A low battery, incomplete startup, or system fault may prevent normal communication behavior.

Expected outcome: The pump is fully powered and operating without unrelated system errors.

If normal operation is restored and wireless communication returns, perform a functional check, document the result, and stop.

6. Inspect the Pump Externally

Check the pump for:

Do not open the housing or attempt antenna or board-level repair during floor troubleshooting.

Expected outcome: No external physical condition is found that could indicate internal communication-hardware damage.

If damage, heat, odor, or fluid intrusion is present, remove the pump from service immediately.

7. Check Wireless Status Indicators

Observe the pump’s wireless or communication indicator.

Determine whether it shows:

Allow sufficient time after startup or relocation for the pump to reconnect. Agilia wireless transactions are device-initiated and periodic, so communication may not appear instantaneous.

Expected outcome: The displayed status identifies whether the pump is failing to associate with Wi-Fi or failing to communicate after association.

8. Perform a Controlled Restart

When the pump is not connected to a patient and clinical use has ended:

Do not repeatedly power-cycle a pump that displays a persistent system error or enters a restart loop.

Expected outcome: A temporary software or network-association problem clears and communication returns.

If communication is restored, confirm that the device appears in the appropriate management application before returning it to service.

9. Compare With a Known-Good Pump

Place a known-good Agilia Wi-Fi pump beside the affected device and compare:

Use pumps assigned to the same facility, profile, and network environment whenever possible.

Expected outcome: The comparison confirms whether the problem follows the pump or affects the surrounding network.

If both devices fail, escalate the network or server issue rather than replacing parts in the pump.

10. Confirm Device Assignment and Network Enrollment

Using approved Clinical Engineering or system-management resources, verify that:

Agilia Wi-Fi models may use IEEE 802.11 wireless networking with TCP/IP, IPv4, DHCP, and supported enterprise or pre-shared-key security configurations.

Do not change network credentials or security settings without authorization and approved configuration tools.

Expected outcome: Administrative records and network enrollment match the physical device.

11. Check for a Broader Network or Server Problem

Determine whether:

Coordinate with the facility’s network, cybersecurity, pharmacy informatics, or infusion-system administrator as appropriate.

Expected outcome: A system-wide infrastructure issue is identified or ruled out before the pump is sent for repair.

12. Verify Communication After Recovery

When the connection returns, confirm the expected function rather than relying only on the wireless icon.

Verify, as applicable:

Do not simulate clinical documentation or transmit patient-associated data unless the approved test environment and procedure permit it.

Expected outcome: Communication is stable and the pump is correctly recognized by the connected system.

If the issue is resolved, complete the required operational inspection, document the repair, and return the pump to service according to facility policy.

If the Problem Persists

If wireless communication remains unavailable after power, location, coverage, device assignment, server availability, and network conditions have been checked, common external causes have been ruled out.

The problem may involve the pump’s internal wireless hardware, stored configuration, security credentials, software, or communication circuitry.

The device should be:

Do not open the housing, replace communication components, or alter protected network settings without the approved service documentation, training, and authorization.

Knowing when to stop and escalate is proper troubleshooting.

Clinical Use Tip

A network failure does not automatically mean the Agilia pump has stopped infusing. Confirm therapy and local pump alarms first. Never delay medication delivery solely to restore wireless communication.

Do not troubleshoot the pump while it is being used on an active patient. Move the patient to a verified backup device first when troubleshooting could affect therapy, and maintain therapy continuity throughout the process.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Clinical staff reported that the Fresenius Kabi Agilia pump showed no wireless connection and was not appearing in the infusion-management system."

Cause

What was observed during troubleshooting.

Example:
"The pump connected normally in another department, while a known-good pump also failed in the original room, indicating a location-specific wireless network issue."

Resolution

What action was taken.

Example:
"Verified normal infusion operation, reported the affected location to the network team, and returned the pump to service after stable communication was confirmed in a known-good coverage area."

Helpful Details to Include (If Known)

Final Thought

Protect the patient and confirm the pump’s primary infusion function before investigating connectivity. Use location and known-good-device comparisons to separate pump failures from infrastructure problems, and document the final status clearly.

That is successful troubleshooting.