On this page
Asset Type
Manufacturer
Model
What This Guide Helps With
Troubleshoots inaccurate, drifting, or unstable arterial and venous pressure readings caused by bloodline setup, transducer protectors, moisture, connections, or calibration concerns.
Step-by-Step Troubleshooting
1. Protect the Patient and Maintain Monitoring Continuity
Do not troubleshoot unreliable arterial or venous pressure monitoring while a patient depends on those readings for safe dialysis treatment.
If pressure values are implausible, unstable, or no longer correspond with the clinical circuit, transfer treatment to a verified machine or follow the facility's approved dialysis safety procedure before technical evaluation.
Expected outcome: The patient is no longer dependent on questionable pressure monitoring, and the machine can be evaluated safely.
2. Confirm Which Pressure Channel Is Affected
Determine whether the problem involves the arterial channel, venous channel, or both. Clarify whether staff observed a fixed offset, slow drift, erratic fluctuation, failure to return near baseline, or readings inconsistent with the extracorporeal circuit.
Record the exact behavior and whether it occurred during setup or treatment.
Expected outcome: The affected channel and symptom are clearly identified.
3. Inspect the External Bloodline Pressure Path
Using a non-patient test setup when applicable, inspect the arterial and venous pressure monitoring lines for kinks, clamping, loose connections, fluid contamination, trapped liquid, damaged tubing, or incorrect routing.
A distorted pressure signal can originate in the disposable circuit rather than the machine.
Expected outcome: The pressure-monitoring tubing is correctly routed and unobstructed. If correcting an external tubing problem restores stable readings, troubleshooting can stop after verification.
4. Inspect Transducer Protectors
Check the applicable arterial and venous transducer protectors for moisture, blood, saline, discoloration, blockage, damage, or incorrect installation.
If fluid has crossed a transducer protector or contamination of the machine-side pressure port is suspected, remove the machine from service and follow the appropriate contamination and service procedure rather than continuing routine troubleshooting.
Expected outcome: Transducer protectors are dry, intact, and correctly installed. Any contamination concern triggers appropriate removal from service.
5. Inspect Machine-Side Pressure Connections
Inspect accessible pressure connection ports for residue, damage, loose fittings, or evidence that a disposable connection is not seating correctly.
Do not insert objects into pressure ports or apply unapproved pressure sources.
Expected outcome: Pressure connections are clean, intact, and provide a secure external interface.
6. Compare Behavior With a Known-Good Disposable Setup
If safe and permitted, install a known-good compatible disposable circuit or pressure line arrangement and observe the channel again.
This helps separate machine-side drift from a defective transducer protector or disposable pressure line.
Expected outcome: The pressure signal becomes stable with the known-good setup, identifying the original disposable pathway as the likely cause, or the problem remains machine-specific.
7. Verify Baseline or Zero Behavior
With the machine configured for an appropriate non-patient test condition, observe whether the affected pressure channel returns to the expected baseline condition according to approved procedures.
Do not perform unauthorized calibration adjustments based solely on a displayed offset.
Expected outcome: The channel behaves consistently at baseline. Persistent offset or drift indicates the need for approved calibration or service evaluation.
8. Test With Approved Pressure Measurement Equipment
If within Clinical Engineering scope, use approved calibrated pressure test equipment and the applicable Fresenius service procedure to compare displayed pressure with the applied test condition.
Preserve the distinction between testing and calibration. A failed comparison identifies a performance problem; it does not authorize an unapproved adjustment.
Expected outcome: Displayed pressure corresponds appropriately with the approved test input. If the channel fails the approved test, remove the machine from service for calibration or repair.
9. Perform Final Functional Verification
After correcting an external cause or completing authorized calibration, verify both arterial and venous pressure channels through the applicable functional test.
Confirm stable readings, appropriate response to test pressure changes, absence of drift, and normal alarm functionality when included in the approved return-to-service process.
Expected outcome: Both pressure channels operate consistently and pass applicable functional verification. Troubleshooting can stop.
If the Problem Persists
If the disposable pressure path, transducer protectors, accessible ports, and known-good setup are verified but pressure readings remain inaccurate or drift, common external causes have been ruled out.
Remaining causes may involve pressure sensing components, pneumatic pathways, calibration, internal tubing, wiring, analog measurement circuitry, or control-system processing.
The machine should be:
- Removed from service.
- Labeled Out of Service.
- Sent for repair or bench evaluation.
- Evaluated using appropriate Fresenius Medical Care service documentation and calibrated test equipment.
- Calibrated or repaired only by qualified personnel.
- Subjected to applicable pressure-channel, alarm, electrical safety, and functional return-to-service testing.
Stopping before unauthorized internal adjustment is proper troubleshooting.
Clinical Use Tip
Treat unexplained arterial or venous pressure drift as a patient-safety issue because pressure monitoring is part of extracorporeal circuit surveillance.
Work Order Documentation (CCR Method)
CCR = Complaint, Cause, Resolution
Complaint
What was reported by the clinical staff.
Example:
"Dialysis staff reported that the venous pressure reading on the 2008T slowly drifted upward even when the external circuit condition remained unchanged."
Cause
What was observed during troubleshooting.
Example:
"Clinical Engineering found moisture in the venous transducer protector causing an unstable pressure transmission path."
Resolution
What action was taken.
Example:
"Clinical Engineering replaced the affected disposable protector and test setup, verified stable venous pressure response with approved test equipment, and confirmed normal pressure-channel operation before return to service."
Helpful Details to Include (If Known)
- Arterial, venous, or both channels affected
- Reported pressure behavior
- Whether issue occurred during setup or treatment
- Pressure tubing condition
- Transducer protector condition
- Evidence of moisture or contamination
- Known-good disposable comparison
- Baseline behavior
- Approved pressure test results
- Calibration performed, if authorized
- Alarm verification result
- Final device status
Final Thought
Pressure-channel complaints should be approached from the disposable circuit inward. Protect the patient, inspect pressure lines and protectors, verify the measurement with approved equipment, escalate persistent drift appropriately, and document the verified cause and final test results.
That is successful troubleshooting.