Fujifilm Sonosite Edge II

Battery Charging Issues

On this page

Asset Type

Ultrasound System

Manufacturer

Fujifilm Sonosite

Model

Edge II

What This Guide Helps With

This guide assists Clinical Engineering in troubleshooting situations where a Fujifilm Sonosite Edge II battery is not charging properly or has reduced runtime. Common symptoms include:

The guide focuses on simple external checks before assuming internal battery or board failure.

Step-by-Step Troubleshooting

Verify AC Power Source

Check Power Cord and Adapter

Confirm Battery Seating

Inspect Battery for Damage

Observe System Indicators

Test with a Known-Good Battery

Power Cycle the System

If the Problem Persists

External causes (outlet, cord, adapter, battery seating) have been ruled out. The internal charging circuit or battery management system may be faulty.

Knowing when to stop prevents unsafe troubleshooting and protects patient care.

Clinical Use Tip

Do not attempt battery charging troubleshooting while the device is connected to an active patient.

If patient imaging is required, move the patient to another operational ultrasound system first.

Only perform troubleshooting in a safe environment where patient care is not impacted.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
“Battery not charging, system only works on AC power.”

Cause

What was observed during troubleshooting.

Example:
Observed that the AC outlet and adapter are functioning, battery seated correctly, but the system does not charge. Battery indicator remains static.

Resolution

What action was taken.

Example:
Removed device from service, labeled Out of Service, and sent for repair evaluation.

Helpful Details to Include

Final Thought

Battery issues in portable ultrasound systems can appear minor but may compromise system readiness. Logical troubleshooting and early escalation protect patient safety and reduce risk. Accurate CCR documentation ensures clear communication and supports future maintenance decisions.

That is successful troubleshooting.