On this page
Asset Type
Manufacturer
Model
What This Guide Helps With
Troubleshoots unavailable needle-visualization or optimization functions caused by transducer selection, imaging mode, presets, workflow, or configuration.
Step-by-Step Troubleshooting
1. Protect the Procedure
If a procedure depends on an imaging-assistance feature that is unavailable, clinicians should use an approved alternate imaging method or another verified ultrasound system rather than delay or compromise patient care.
Expected outcome: The procedure continues only with adequate visualization and appropriate clinical capability.
2. Confirm the Missing Function
Identify the exact feature that is unavailable and determine whether it is missing, disabled, grayed out, or selectable but ineffective.
Expected outcome: The reported behavior is clearly defined.
3. Verify the Connected Transducer
Confirm that the intended compatible transducer is recognized correctly and produces a normal baseline image.
Expected outcome: The transducer is correctly detected and baseline imaging is normal.
4. Verify the Imaging Mode
Confirm that the system is operating in the normal imaging mode in which the requested feature is expected to be available. Exit incompatible modes or workflows using standard controls.
Expected outcome: The feature becomes available when the appropriate imaging context is selected, or the problem is confirmed to persist.
5. Check the Exam Type or Preset
Review the selected exam type or preset through normal clinical controls. Compare with a known-working setup or another Sonosite LX when available.
Expected outcome: The intended exam configuration supports the expected feature.
6. Restore Normal User-Accessible Settings
If a prior user setting appears to have disabled or changed the feature, return it to the approved clinical configuration without entering service menus or changing protected system configuration.
Expected outcome: The feature becomes selectable and operates normally.
7. Compare With Another Compatible Transducer or Preset
Where appropriate, test another compatible transducer or known-good clinical preset to determine whether availability follows a specific transducer or configuration.
Expected outcome: The issue is isolated to a particular configuration, transducer, or system-wide feature.
8. Perform Final Functional Verification
Activate the feature under appropriate nonpatient test conditions and confirm that its expected visual effect or optimization is present and stable.
Expected outcome: The function is available and performs consistently. Troubleshooting can stop.
If the Problem Persists
Transducer recognition, imaging mode, exam preset, and accessible settings have been checked. Persistent unavailability may involve software options, licensing, configuration, application software, transducer compatibility, or a service-level system condition.
If the feature is required for intended clinical use and cannot be verified, remove the system from that application and label it Out of Service when appropriate. Further evaluation should use Fujifilm Sonosite documentation and authorized service tools. Configuration or software changes should be performed only by qualified personnel.
Verify the feature and associated imaging functions after correction before return to service. Knowing when to stop external troubleshooting is proper troubleshooting.
Clinical Use Tip
Do not assume an advanced imaging feature should appear with every transducer, exam type, or operating mode; verify the complete clinical configuration.
Work Order Documentation (CCR Method)
CCR = Complaint, Cause, Resolution
Complaint
What was reported by the clinical staff.
Example:
"Clinical staff reported that the needle-visualization function was unavailable during procedure setup on the Sonosite LX."
Cause
What was observed during troubleshooting.
Example:
"Clinical Engineering found that the system was operating under a different exam preset in which the requested function was not available."
Resolution
What action was taken.
Example:
"The approved clinical preset was selected, the feature became available, and its operation was functionally verified."
Helpful Details to Include (If Known)
- Feature reported unavailable
- Transducer connected
- Exam type or preset
- Imaging mode
- Whether the control was missing or disabled
- Comparison configuration used
- Known-good transducer results
- Settings observed before correction
- Functional result after correction
- Final system status
Final Thought
Feature availability often depends on the complete imaging configuration. Verify transducer, mode, preset, and authorized settings before assuming software or hardware failure, then escalate unresolved conditions appropriately.
That is successful troubleshooting.