On this page
Asset Type
Manufacturer
Model
What This Guide Helps With
Troubleshoots incomplete startup, repeated boot failure, or storage-related messages caused by power, connected accessories, software state, or internal storage problems.
Step-by-Step Troubleshooting
1. Protect Clinical Operations
Do not use a system that cannot reliably boot or access required patient data. Move imaging to another verified ultrasound system.
Expected outcome: Clinical care continues without relying on an unstable device.
2. Record the Exact Startup Condition
Observe where startup stops and record any exact message, blank screen, restart behavior, or storage warning. Do not repeatedly cycle power without purpose.
Expected outcome: The failure pattern is documented for comparison and escalation.
3. Verify Stable External Power
Connect the system to a verified AC outlet using its approved power equipment. Inspect external power connections for looseness or damage.
Expected outcome: Stable power is supplied to the system.
4. Disconnect Nonessential External Accessories
With the system powered down, disconnect nonessential USB devices, external displays, recording devices, or other accessories that are safe to remove.
Expected outcome: A connected accessory is ruled in or out as a contributor to startup failure.
5. Perform One Controlled Restart
Use the normal shutdown/startup procedure or approved recovery from the current state without entering service menus or performing unauthorized reset procedures.
Expected outcome: The system completes startup and remains stable. If successful, proceed to functional verification.
6. Check Available Storage Through Normal Menus
If the system boots, review user-accessible storage status and identify whether storage is unusually full or unavailable. Do not delete clinical studies unless permitted by facility policy and data-retention procedures.
Expected outcome: Storage is accessible and its status can be evaluated without compromising patient records.
7. Verify Basic Imaging and Data Functions
Acquire an appropriate test image and verify that it can be saved and retrieved through normal operation.
Expected outcome: Image acquisition and local storage function normally. If boot and storage remain stable, troubleshooting can stop after final verification.
8. Observe for Recurrence
Restart the system once more if appropriate and confirm the problem does not immediately recur.
Expected outcome: The system repeatedly completes startup without storage errors or unexpected restarts.
If the Problem Persists
External power, nonessential accessories, and normal restart procedures have been ruled out. A persistent boot or internal-storage problem may involve software corruption, storage hardware, operating-system files, internal connections, or other service-level electronics.
Remove the Sonosite LX from service, label it Out of Service, and send it for repair or bench evaluation. Use Fujifilm Sonosite service documentation and approved diagnostic/test equipment. Do not attempt unsupported data recovery, operating-system modifications, internal storage replacement, or unauthorized service procedures.
After repair, confirm boot reliability, imaging, data storage/retrieval, and applicable safety functions before return to service. Knowing when to stop external troubleshooting is proper troubleshooting.
Clinical Use Tip
Protect patient data as well as equipment function; do not delete studies or reinitialize storage simply to clear a startup problem.
Work Order Documentation (CCR Method)
CCR = Complaint, Cause, Resolution
Complaint
What was reported by the clinical staff.
Example:
"Clinical staff reported that the Sonosite LX repeatedly stopped during startup and displayed a storage-related warning."
Cause
What was observed during troubleshooting.
Example:
"Clinical Engineering verified stable external power and reproduced the failure with nonessential accessories disconnected, indicating a service-level boot/storage problem."
Resolution
What action was taken.
Example:
"The system was removed from service and referred for manufacturer-supported bench evaluation with the startup condition documented."
Helpful Details to Include (If Known)
- Exact startup or storage message
- Point where startup stops
- Restart behavior
- AC-power verification
- Battery behavior
- External accessories connected
- Storage status if accessible
- Whether test images can be saved
- Frequency of recurrence
- Final system status
Final Thought
Boot and storage failures can affect both availability and patient data. Verify power and external influences first, avoid destructive recovery attempts, and escalate persistent failures with the exact observed symptoms documented.
That is successful troubleshooting.