GE Healthcare Aisys

Alarm: ACB SW ERROR

On this page

Asset Type

Anesthesia Machine

Manufacturer

GE Healthcare

Model

Aisys

What This Guide Helps With

This guide helps troubleshoot an ACB SW ERROR alarm on a GE Healthcare Aisys anesthesia machine. This alarm may appear with system malfunction behavior, startup failure, intermittent reset behavior, checkout failure, freezing, or unsafe anesthesia machine operation.

An ACB software-related alarm should be treated seriously because the Anesthesia Control Board is involved in core system control. The machine should not be trusted for patient use until the alarm is resolved and the unit passes the required checkout.

Step-by-Step Troubleshooting

1. Ensure Patient Safety First

2. Verify the Reported Alarm and Behavior

3. Power Down and Reboot the System

4. Check AC Power Source

5. Check Battery and Power Status

6. Inspect External Connections and Modules

7. Review Recent Service or Software Activity

8. Repeat Startup and Checkout

9. Reload ACB Software if the Alarm Persists

10. Determine Whether the Fault Is Device-Related

If the Problem Persists

If ACB SW ERROR continues after reboot, power checks, battery checks, external inspection, and ACB software reload, the basic external causes have been ruled out.

Persistent ACB SW ERROR strongly indicates an internal control board or software integrity problem. Knowing when to stop is proper troubleshooting.

Clinical Use Tip

Never troubleshoot an ACB SW ERROR on an active patient. Move the patient to a backup anesthesia machine or approved backup ventilation method first. A system malfunction on an anesthesia machine should be treated as a patient safety issue until the machine is verified safe, stable, and able to pass checkout.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"GE Healthcare Aisys anesthesia machine reported ACB SW ERROR. Staff reported system malfunction during startup or checkout. Unit removed from patient use for evaluation."

Cause

What was observed during troubleshooting.

Example:
"Confirmed ACB SW ERROR on restart. External power, battery status, and visible external connections were checked. Alarm persisted after reboot, indicating likely ACB software or control board fault."

Resolution

What action was taken.

Example:
"Removed unit from service and labeled Out of Service. Reboot performed and external checks completed. ACB software reload required. If alarm persists after reload, ACB replacement or repair evaluation required before return to service."

Helpful Details to Include (If Known)

Final Thought

ACB SW ERROR should be handled as a real anesthesia machine system malfunction, not a nuisance alarm. Start with patient safety, confirm the alarm, check power and external causes, reboot appropriately, then escalate to software reload or ACB replacement when the fault persists. Clear CCR documentation helps show what was checked, what was ruled out, and why the device was removed from service.

That is successful troubleshooting.