GE Healthcare Aisys

Alarm: ACB UNEXPECTED RESET

On this page

Asset Type

Anesthesia Machine

Manufacturer

GE Healthcare

Model

Aisys

What This Guide Helps With

This guide helps troubleshoot the ACB UNEXPECTED RESET alarm on a GE Healthcare Aisys anesthesia machine. This alarm indicates an unexpected Anesthesia Control Board reset, which may cause system malfunction, unexpected reboot behavior, loss of normal screen function, or interruptions to anesthesia machine control.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Alarm and Behavior

Check AC Power Source

Check Power Cord and External Power Condition

Check for Recent Events Before the Alarm

Reboot the System

Run or Repeat System Checkout

Check Whether the Alarm Is Repeatable

Isolate External Causes Before Calling the ACB Failed

Stop When Resolved

If the Problem Persists

If the ACB UNEXPECTED RESET alarm returns after reboot and external power issues have been ruled out, the condition may indicate an internal device failure involving the Anesthesia Control Board or related control circuitry.

Clinical Use Tip

Never troubleshoot an anesthesia machine with an unresolved system malfunction alarm on an active patient. Move the patient to backup equipment first, then continue troubleshooting only when it is safe to do so.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"GE Healthcare Aisys anesthesia machine reported ACB UNEXPECTED RESET alarm. Staff reported system malfunction and unexpected reset during startup or checkout."

Cause

What was observed during troubleshooting.

Example:
"Alarm confirmed. External AC power and power cord checked. System reboot performed. Alarm cleared after reboot, or alarm returned after external power issues were ruled out."

Resolution

What action was taken.

Example:
"Rebooted system, verified AC power connection, and repeated system checkout. If checkout passed and alarm did not return, device was returned to service. If alarm returned, device was removed from service and sent for repair or ACB evaluation."

Helpful Details to Include (If Known)

Final Thought

An ACB UNEXPECTED RESET alarm should be treated as a control system reliability issue. Start with patient safety, verify the alarm, rule out simple power problems, reboot the system, and confirm proper operation through checkout. If the alarm persists, remove the machine from service and escalate for repair. Good troubleshooting protects the patient, the clinical team, and the equipment record.

That is successful troubleshooting.