GE Healthcare Aisys

Alarm: ACMains POWER FAIL

On this page

Asset Type

Anesthesia Machine

Manufacturer

GE Healthcare

Model

Aisys

What This Guide Helps With

This guide helps troubleshoot the ACMains POWER FAIL alarm on a GE Healthcare Aisys anesthesia machine. This alarm indicates the system is running on battery because AC mains power has been lost, unstable, disconnected, or not detected by the machine.

Loss of AC mains power can affect how long the anesthesia machine can safely operate before battery depletion. If AC power is not restored, the condition may progress to low battery, battery empty, or system shutdown alarms.

Step-by-Step Troubleshooting

1. Ensure Patient Safety First

2. Verify the Reported Alarm and Behavior

3. Confirm the Power Cord Is Plugged In

4. Restore AC Mains Power

5. Check the Wall Outlet

6. Inspect the Power Cord and Plug

7. Check for Facility Power or Circuit Issues

8. Monitor Battery Status

9. Check for Repeat or Intermittent Alarm Behavior

10. Stop When Resolved

If the Problem Persists

If ACMains POWER FAIL remains active after confirming a known-good outlet, secure power cord connection, acceptable cord condition, and no facility power issue, external and common causes have been ruled out.

Persistent AC mains power failure under known-good power conditions may indicate a device power entry issue, power detection problem, internal power supply path issue, or related internal connection fault.

Clinical Use Tip

Never troubleshoot this alarm on an active patient in a way that risks interruption of ventilation, monitoring, or anesthesia delivery. If the unit remains on battery or battery time is decreasing, move the patient to a backup device or alternate anesthesia setup first, following clinical direction and facility policy.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"GE Aisys anesthesia machine displayed ACMains POWER FAIL alarm with message Plug in power cable. On battery. Staff reported unit was operating on battery power."

Cause

What was observed during troubleshooting.

Example:
"AC mains power was not detected by the machine. Power cord was loose at the outlet and the unit was operating from battery backup."

Resolution

What action was taken.

Example:
"Verified alarm condition, reseated power cord, tested known-good AC outlet, restored AC mains power, monitored battery status, and confirmed alarm cleared. Unit passed operational checks and was returned to service."

Helpful Details to Include (If Known)

Final Thought

ACMains POWER FAIL usually means the Aisys correctly detected loss of AC mains power and switched to battery backup. Clinical Engineering should confirm patient safety, restore or verify AC power, check simple external causes first, and escalate only when the machine fails to recognize known-good power.

Clear CCR documentation helps show whether the event was a resolved power-source issue or a device-level fault requiring repair.

That is successful troubleshooting.