GE Healthcare Aisys

Alarm: BATTERY FAIL

On this page

Asset Type

Anesthesia Machine

Manufacturer

GE Healthcare

Model

Aisys

What This Guide Helps With

This guide helps troubleshoot the BATTERY FAIL alarm on a GE Healthcare Aisys anesthesia machine. This alarm may appear with a No Battery Backup message and can indicate low battery voltage, failed charging, a weak battery, or a battery charge circuit issue.

A BATTERY FAIL condition should be treated as a backup power problem until proven otherwise. If AC power is lost, the machine may not have reliable battery support.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Alarm and Behavior

Check AC Power Source First

Leave the System Plugged In to Charge

Check Recent Power or Storage History

Inspect External Battery-Related Conditions

Repeat Startup or Checkout After Charging

Review Battery and Charging Information Through Approved Service Procedures

Replace the Battery if Indicated

Stop When Resolved

If the Problem Persists

If the BATTERY FAIL alarm continues after verifying AC power, allowing the battery to charge, checking external conditions, reviewing recent power history, and replacing the battery if indicated, external and common battery-related causes have been ruled out.

A persistent BATTERY FAIL alarm may indicate an internal charging circuit, power management, wiring, or control issue.

The device should be removed from service, labeled Out of Service, and sent for repair or bench evaluation using approved service procedures and Service Software.

Do not keep an anesthesia machine in service with unresolved battery backup failure unless your facility has formally assessed and accepted the risk.

Clinical Use Tip

Never troubleshoot on an active patient. Move the patient to a backup device first. An anesthesia machine with no battery backup should not be relied on during a case where loss of AC power could affect ventilation, monitoring, or anesthetic delivery.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"GE Healthcare Aisys anesthesia machine reported BATTERY FAIL alarm with No Battery Backup message during startup or checkout."

Cause

What was observed during troubleshooting.

Example:
"BATTERY FAIL alarm confirmed. AC power source and power cord were checked. Battery was allowed to charge, but alarm persisted, indicating failed battery or possible charging circuit issue."

Resolution

What action was taken.

Example:
"Removed unit from patient use. Connected to known-good AC power, allowed battery to charge, restarted system, reviewed battery charge status as needed, and replaced battery if indicated. Unit passed checkout and was returned to service, or remained Out of Service for further repair."

Helpful Details to Include (If Known)

Final Thought

BATTERY FAIL on the GE Healthcare Aisys should be treated as a real backup power issue until proven otherwise. Start with patient safety, confirm the displayed alarm, check AC power and charging time, then use approved service procedures only after simple causes have been ruled out. If the alarm persists, remove the machine from service and escalate appropriately.

That is successful troubleshooting.