GE Healthcare Aisys

Alarm: COM ERROR VENT TO ACB

On this page

Asset Type

Anesthesia Machine

Manufacturer

GE Healthcare

Model

Aisys

What This Guide Helps With

This guide helps troubleshoot the COM ERROR VENT TO ACB alarm on a GE Healthcare Aisys anesthesia machine. This alarm indicates a communication problem between the ventilator side of the system and the Anesthesia Controller Board, which may cause a system malfunction and affect ventilator control or reliability.

Step-by-Step Troubleshooting

1. Ensure Patient Safety First

2. Verify the Reported Alarm and Behavior

3. Understand What the Alarm Means

4. Reboot the System

5. Check Power Source and Startup Conditions

6. Check for Obvious External or Physical Issues

7. Review Alarm Timing and Event History

8. Check Cabling and Connections

9. Isolate Whether the Issue Appears Cable-Related or Device-Related

10. Escalate to Ventilator Interface Board Evaluation

11. Escalate to Anesthesia Controller Board Evaluation if Needed

If the Problem Persists

If COM ERROR VENT TO ACB continues after rebooting, checking the power source, inspecting physical condition, reviewing alarm history, and checking cabling, the common external causes have been ruled out.

A persistent alarm may indicate an internal communication failure involving the ventilator communication path, Ventilator Interface Board, Anesthesia Controller Board, or related harnessing.

Do not return the anesthesia machine to patient use until the alarm is cleared, checkout passes, and ventilator operation is verified.

Clinical Use Tip

Never troubleshoot this alarm on an active patient. A system malfunction involving ventilator communication can affect ventilation reliability. Move the patient to a backup anesthesia machine, manual ventilation, or another appropriate support method before troubleshooting.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"GE Healthcare Aisys anesthesia machine displayed COM ERROR VENT TO ACB with System Malfunction alarm. Staff reported ventilator communication fault during startup or use and removed machine from service."

Cause

What was observed during troubleshooting.

Example:
"Verified COM ERROR VENT TO ACB alarm. Rebooted system and checked power source, external condition, alarm history, and accessible cabling. Alarm indicates loss of communication between ventilator side of system and Anesthesia Controller Board after communication had been established."

Resolution

What action was taken.

Example:
"Removed machine from patient use. Rebooted system, inspected power and cabling, and reviewed alarm behavior. If alarm cleared, completed checkout and verified ventilator operation before returning to service. If alarm persisted, labeled Out of Service and escalated for bench repair with VIB and ACB evaluation as needed."

Helpful Details to Include (If Known)

Final Thought

COM ERROR VENT TO ACB should be treated as a serious anesthesia machine communication fault. Start with patient safety, confirm the alarm, reboot the system, check power and cabling, then escalate logically toward VIB and ACB evaluation if the issue persists. Clear documentation shows what was verified, what was ruled out, and why the device was either returned to service or removed for repair.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide