GE Healthcare Aisys

Alarm: VENT SUSTAINED PAW SDOWN

On this page

Asset Type

Anesthesia Machine

Manufacturer

GE Healthcare

Model

Aisys

What This Guide Helps With

This guide helps troubleshoot VENT SUSTAINED PAW SDOWN on a GE Healthcare Aisys anesthesia machine. This alarm can occur when high airway pressure causes a ventilator failure condition and the machine enters monitoring-only operation.

This condition may be associated with Vent Fail. Monitoring Only. and airway pressure remaining above the expected safe range. Because ventilation may be affected, this alarm should be treated as a serious patient safety issue.

Step-by-Step Troubleshooting

1. Ensure Patient Safety First

2. Verify the Reported Alarm and Behavior

3. Check for Obvious External High-Pressure Causes

4. Check the Breathing System Setup

5. Review Ventilation Settings and Alarm Context

6. Power Cycle the System After Patient Is Safe

7. Perform a Functional Check

8. Inspect Airway Pressure Reading Behavior

9. Isolate External Setup Issue vs Device Fault

10. Return to Service Only If Fully Resolved

If the Problem Persists

If VENT SUSTAINED PAW SDOWN continues after reboot, known-good circuit setup, external restriction checks, and functional testing, remove the anesthesia machine from service.

At that point, external causes such as kinked hoses, blocked filters, incorrect breathing circuit setup, scavenging restriction, and improper ventilation settings have been ruled out.

Label the unit Out of Service and send it for repair or bench evaluation. A repeating high Paw shutdown alarm may indicate an internal pressure sensing, ventilator control, or pressure measurement issue that should not be cleared casually.

Clinical Use Tip

Never troubleshoot this alarm on an active patient. If the machine displays Vent Fail. Monitoring Only., move the patient to a backup device or alternate safe ventilation method before beginning Clinical Engineering evaluation.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"GE Healthcare Aisys anesthesia machine reported VENT SUSTAINED PAW SDOWN with Vent Fail. Monitoring Only. during use. Staff reported high airway pressure alarm and ventilator shutdown."

Cause

What was observed during troubleshooting.

Example:
"Found breathing circuit restriction from kinked hose causing excessive airway pressure. No internal device fault found during checkout."

Resolution

What action was taken.

Example:
"Removed kinked circuit, installed known-good circuit, rebooted system, completed anesthesia machine checkout and test lung ventilation. Paw readings normal. Alarm did not return. Returned unit to service."

Helpful Details to Include (If Known)

Final Thought

VENT SUSTAINED PAW SDOWN should be treated as a serious ventilator failure condition, not just a nuisance alarm. Start with patient safety, confirm the exact alarm, rule out external pressure causes, reboot only after the patient is safe, and escalate if the Paw signal or ventilator behavior remains abnormal.

Good troubleshooting shows whether the event was caused by setup, obstruction, or a likely internal failure.

That is successful troubleshooting.