GE Aisys CS2 Anesthesia Machine

Leak Test Failure

On this page

Asset Type

Anesthesia Machine

Manufacturer

GE Healthcare

Model

Aisys CS2

What This Guide Helps With

This guide addresses situations where the Aisys CS2 anesthesia machine fails its leak test, which may prevent the machine from entering “ready for use” status. Leak test failures can indicate:

The focus is on verifying external connections and setup before assuming internal leaks or module failure.

Step-by-Step Troubleshooting

Verify Machine Power & Status

Why: Some leak test failures are triggered if the machine detects other unresolved faults.

Inspect the Breathing Circuit

Why: Loose or improperly seated hoses are a common cause of leak test failure.

Check the CO₂ Absorber Canister & Vaporizers

Why: Misalignment can create small leaks that prevent the test from passing.

Examine the Reservoir Bag and Mask/Endotracheal Tube

Why: The leak test is performed without patient devices; attached masks or tubes will cause failure.

Run a Manual Leak Test with the Circuit Closed

Why: Confirms that external components are functioning correctly.

Inspect for Visible Damage or Moisture

Why: Physical damage or condensation can cause leaks that prevent test completion.

If the Problem Persists

If all external checks pass and the machine still fails the leak test:

Reinforce that stopping at this point prevents patient risk and unnecessary component damage.

Clinical Use Tip

Never attempt to troubleshoot leaks on a patient-connected circuit.

Move the patient to another verified anesthesia machine if the leak test fails.

Only perform troubleshooting on a safe, unoccupied circuit.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
“Leak test fails on startup; machine will not enter ready status.”

Cause

What was observed during troubleshooting.

Example:
“External circuit components inspected; inspiratory/expiratory limbs and vaporizers seated correctly; no obvious damage observed.”

Resolution

What action was taken.

Example:
“Machine removed from service and labeled Out of Service for GE Healthcare evaluation.”

Helpful Details to Include

Final Thought

Patient safety is the top priority. Systematically checking external connections often resolves leak test failures. Knowing when to escalate prevents further risk and preserves the machine’s integrity. Documenting every step ensures continuity and accountability.

That is successful troubleshooting.