GE Healthcare Aisys Anesthesia Machine

Gas or Agent Delivery Errors

On this page

Asset Type

Anesthesia Machine

Manufacturer

GE Healthcare

Model

Aisys

What This Guide Helps With

This guide addresses situations where the GE Healthcare Aisys anesthesia machine:

This guide focuses on logical, external checks first—such as gas supply, vaporizers, circuit setup, and configuration—before assuming internal flow sensor or module failure.

Step-by-Step Troubleshooting

Ensure the Patient Is Safely Supported First

Check Pipeline Gas Supply Pressures

Why this matters: Gas delivery errors are commonly caused by supply pressure issues rather than internal machine faults.

Verify Cylinder Status (If Installed)

Confirm Vaporizers Are Properly Seated and Locked

An improperly seated vaporizer can trigger agent delivery or concentration errors.

Check Fresh Gas Flow Settings

Inspect the Breathing Circuit

A significant circuit leak may appear as a delivery issue.

Run the System Self-Test (If Not in Active Case)

Check for Recent Service or Configuration Changes

Observe for Physical Signs

These may indicate internal valve or sensor failure.

If the Problem Persists

If gas supply, vaporizers, circuit integrity, and configuration have been verified and the machine continues to display gas or agent delivery errors, the issue is likely internal. Possible causes include:

At this point:

Knowing when to stop prevents unnecessary risk and protects patient safety.

Clinical Use Tip

Never troubleshoot gas delivery issues while the machine is connected to a patient unless transitioning them safely to manual ventilation or a backup machine. Gas delivery errors directly affect oxygenation and anesthetic depth—patient safety always comes first.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
“OR staff reported the Aisys would not reach set Sevoflurane concentration and displayed low agent delivery alarm during induction.”

Cause

What was observed during troubleshooting.

Example:
Pipeline O₂ pressure was found below normal range due to loose wall hose connection. Machine operating normally once pressure restored.

Resolution

What action was taken.

Example:
Re-secured pipeline hose connection, verified proper supply pressure, completed system checkout test, returned unit to service.

Helpful Details to Include in Documentation

Final Thought

Anesthesia machine gas delivery issues must be approached logically—starting with supply and setup before assuming internal failure. Clear thinking, controlled escalation, and proper documentation protect both patients and providers. The goal is safe function, not unnecessary disassembly.

That is successful troubleshooting.