GE Healthcare Aisys Anesthesia Machine

Ventilator Not Delivering Set Volume

On this page

Asset Type

Anesthesia Machine

Manufacturer

GE Healthcare

Model

Aisys

What This Guide Helps With

This guide addresses situations where the GE Healthcare Aisys anesthesia machine ventilator:

This guide focuses on logical, external checks first—especially circuit integrity, gas supply, settings, and user setup—before assuming internal ventilator failure.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify Ventilation Mode and Settings

Check Fresh Gas Flow and Gas Supply

Inspect the Patient Circuit for Leaks

Check the Adjustable Pressure Limiting (APL) Valve (if in Manual Mode)

Inspect the CO₂ Absorber and Breathing System

Verify Airway Integrity

Perform System Leak / Compliance Test (if patient safely off circuit)

Check Bellows or Piston Movement (Visual Observation Only)

Power Cycle if Appropriate

If the Problem Persists

If external leaks, setup issues, and gas supply problems have been ruled out and volume delivery remains inconsistent:

At this point, the issue is likely internal (flow sensors, ventilator drive system, control valves, or internal leak) and requires controlled bench diagnostics.

Knowing when to stop and escalate is proper troubleshooting.

Clinical Use Tip

Never troubleshoot a ventilator volume issue on an active surgical patient.

Always transition the patient to manual ventilation or an alternate anesthesia machine before performing diagnostic checks.

Patient oxygenation and ventilation take priority over equipment evaluation.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
“OR staff reported anesthesia ventilator not delivering set tidal volume of 500 mL; exhaled volume reading approximately 250 mL with low volume alarm.”

Cause

What was observed during troubleshooting.

Example:
“Found loose connection at CO₂ absorber canister causing circuit leak during positive pressure ventilation.”

Resolution

What action was taken.

Example:
“Reseated and locked absorber canister. Performed leak test—passed. Verified delivered and exhaled volumes matched set tidal volume. Returned to service.”

Helpful Details to Include

Final Thought

An anesthesia ventilator volume issue is most commonly caused by circuit leaks or setup errors—not internal failure. Logical troubleshooting protects the patient, prevents unnecessary downtime, and avoids premature vendor escalation. Always prioritize patient ventilation, follow a structured process, and document findings clearly.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide