On this page
Asset Type
Manufacturer
Model
What This Guide Helps With
Error 0xHOST1100 indicating depleted CPU battery causing incorrect time/date, potential data loss, and possible EMR integration or system stability issues.
Step-by-Step Troubleshooting
Ensure Patient Safety First
- Remove the monitor from active patient use if time/date errors or instability are present
- Transfer monitoring to a backup unit
- Ensure continuous patient monitoring is maintained
Verify the Reported Error and Behavior
- Power on the monitor and confirm Error: 0xHOST1100 is displayed
- Observe when the error appears (startup vs during use)
- Check system time/date accuracy
- Confirm if time resets or is incorrect
Check System Time and Date (Quick Verification)
- Navigate to system settings
- Verify current date and time
- Manually correct if incorrect
- Power cycle and observe if settings are retained
Perform a Full Power Cycle
- Power OFF the monitor
- Disconnect AC power
- Wait 30–60 seconds
- Reconnect power and restart
- Observe if the error returns
Inspect External Conditions
- Verify stable and consistent power source
- Confirm no recent software resets or updates
- Check for abnormal system behavior (reboots, data loss)
- Ensure no environmental factors are contributing
Identify the Root Cause
- Confirm time resets after power loss
- Verify error persists after reboot
- Rule out all external causes
- Determine internal CPU timekeeper battery is end-of-life
If the Problem Persists
If all external causes have been ruled out and the error continues:
- Remove the device from service
- Label the unit as Out of Service
- Send for repair or controlled battery replacement per facility policy
Replacing the CPU battery may affect system configuration or integration. Post-repair validation, including network and EMR connectivity, may be required.
Clinical Use Tip
Never troubleshoot on an active patient. Move the patient to a backup monitor before addressing system errors. Do not rely on a monitor with incorrect time/date, as it can affect clinical documentation and alarm accuracy.
Work Order Documentation (CCR Method)
CCR = Complaint, Cause, Resolution
Complaint
What was reported by the clinical staff.
Example:
"Monitor displaying Error 0xHOST1100 and time resets after power cycle."
Cause
What was observed during troubleshooting.
Example:
"CPU timekeeper battery depleted and unable to retain system time when powered off."
Resolution
What action was taken.
Example:
"Removed from service and sent for battery replacement and system verification."
Helpful Details to Include (If Known)
- Error occurs at startup
- Time/date resets after power cycle
- Power source verified
- No external accessories involved
- No additional alarms present
- Final device status
Final Thought
Error codes like 0xHOST1100 indicate internal component failure. Focus on confirming symptoms, ruling out external causes, and escalating appropriately to maintain patient safety and data integrity.
That is successful troubleshooting.