GE Healthcare CARESCAPE B850

Error: 0xHOST1002 Internal voltage out of spec

On this page

Asset Type

Patient Monitor

Manufacturer

GE Healthcare

Model

CARESCAPE B850

What This Guide Helps With

Error 0xHOST1002 indicating internal voltage out of specification, potentially causing instability, unexpected shutdowns, or unreliable monitoring performance.

Step-by-Step Troubleshooting

1. Ensure Patient Safety First

2. Verify the Reported Error and Behavior

3. Check External Power Source (Most Common Cause)

4. Inspect Power Cord and Connections

5. Check Battery Status and Seating

6. Review Internal Voltage via Webmin (If Accessible)

7. Power Cycle the Monitor

8. Isolate Environmental Factors

If the Problem Persists

If Error 0xHOST1002 continues after all external checks, internal causes have been ruled out. This indicates a likely internal power regulation or power supply issue.

Clinical Use Tip

Never troubleshoot on an active patient. Move the patient to a backup monitor before addressing voltage-related errors.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Monitor displaying Error 0xHOST1002 with intermittent rebooting during use"

Cause

What was observed during troubleshooting.

Example:
"Internal voltage out of specification; external power, cord, and battery verified good"

Resolution

What action was taken.

Example:
"Removed from service, labeled Out of Service, and sent for repair evaluation"

Helpful Details to Include (If Known)

Final Thought

Voltage-related errors should always be taken seriously. Start with external checks, but recognize that persistent faults indicate internal failure requiring escalation and repair.

That is successful troubleshooting.