GE Healthcare CARESCAPE B850

Error: 0xHOST1100 Timekeeper Battery Depleted

On this page

Asset Type

Patient Monitor

Manufacturer

GE Healthcare

Model

CARESCAPE B850

What This Guide Helps With

Error 0xHOST1100 indicates a depleted CPU timekeeper battery, which can cause incorrect system time/date, loss of stored settings, and potential communication or documentation issues with EMR systems.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Error and Behavior

Check System Time and Date

Perform Power Cycle and Recheck

Inspect for External Factors

Replace Timekeeper Battery (If Authorized)

Reconfigure System Settings if Needed

Verify Resolution

If the Problem Persists

If the error remains after battery replacement and configuration checks:

Clinical Use Tip

Never troubleshoot on an active patient. Always move the patient to a backup monitor before addressing system time or internal battery issues.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Monitor displaying Error 0xHOST1100 and incorrect system time."

Cause

What was observed during troubleshooting.

Example:
"Internal CPU timekeeper battery depleted; unable to retain time/date after power cycle."

Resolution

What action was taken.

Example:
"Replaced internal timekeeper battery, reset system time, and verified retention and network communication."

Helpful Details to Include (If Known)

Final Thought

Timekeeper battery failures are simple to identify but can have broader system impacts. Always verify not only that the error is cleared, but that time retention and system integration are fully restored before returning the device to service.

That is successful troubleshooting.