On this page
Asset Type
Manufacturer
Model
What This Guide Helps With
Error 0xHOST1100 indicates a depleted CPU timekeeper battery, which can cause incorrect system time/date, loss of stored settings, and potential communication or documentation issues with EMR systems.
Step-by-Step Troubleshooting
Ensure Patient Safety First
- Remove the monitor from active patient use if time/date accuracy or system stability is affected
- Transfer monitoring to a backup device
- Confirm patient monitoring continues without interruption
Verify the Reported Error and Behavior
- Power on the monitor and confirm Error: 0xHOST1100 is displayed
- Observe when the error appears (startup or during operation)
- Check if the system time/date is incorrect or reset
Check System Time and Date
- Navigate to system settings
- Manually set the correct time and date
- Power cycle the monitor after setting
Perform Power Cycle and Recheck
- Power OFF the monitor
- Disconnect AC power briefly
- Power back ON and check if time/date is retained
Inspect for External Factors
- Verify stable power source
- Check for recent power interruptions
- Confirm no recent system configuration changes
Replace Timekeeper Battery (If Authorized)
- Follow facility policy for internal battery replacement
- Replace the CPU timekeeper battery
- Reassemble and power on the monitor
Reconfigure System Settings if Needed
- Verify network configuration settings
- Confirm communication with central station and EMR
- Check patient data transmission functionality
Verify Resolution
- Power cycle the monitor again
- Confirm time/date remains correct
- Ensure the error no longer appears
If the Problem Persists
If the error remains after battery replacement and configuration checks:
- Remove the monitor from service
- Label the device as Out of Service
- Send for repair or bench evaluation
Clinical Use Tip
Never troubleshoot on an active patient. Always move the patient to a backup monitor before addressing system time or internal battery issues.
Work Order Documentation (CCR Method)
CCR = Complaint, Cause, Resolution
Complaint
What was reported by the clinical staff.
Example:
"Monitor displaying Error 0xHOST1100 and incorrect system time."
Cause
What was observed during troubleshooting.
Example:
"Internal CPU timekeeper battery depleted; unable to retain time/date after power cycle."
Resolution
What action was taken.
Example:
"Replaced internal timekeeper battery, reset system time, and verified retention and network communication."
Helpful Details to Include (If Known)
- Alarm behavior
- Accessories swapped
- Power behavior
- Environmental factors
- Indicator lights
- Final device status
Final Thought
Timekeeper battery failures are simple to identify but can have broader system impacts. Always verify not only that the error is cleared, but that time retention and system integration are fully restored before returning the device to service.
That is successful troubleshooting.