GE Healthcare CARESCAPE VC150

Temperature Probe Errors

On this page

Asset Type

Vital Signs Monitor

Manufacturer

GE Healthcare

Model

CARESCAPE VC150

What This Guide Helps With

This guide assists Clinical Engineering in troubleshooting temperature probe errors on the CARESCAPE VC150. Symptoms may include “probe error,” “sensor disconnected,” inaccurate readings, or inability to display patient temperature. The focus is on external factors such as probe condition, connections, and monitor settings before assuming internal hardware failure.

Step-by-Step Troubleshooting

Verify Patient Safety First

Check Probe Connection

Inspect Probe for Physical Damage

Swap with a Known-Good Probe

Check Monitor Port and Settings

Check for Software or Alarm Messages

If the Problem Persists

Common external causes have been ruled out (probe, connection, settings).

Device should be removed from service, labeled Out of Service, and sent for repair or bench evaluation.

Internal temperature sensor or circuit failure is possible; do not attempt internal repair on-site.

Clinical Use Tip

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
“Temperature probe error displayed; unable to measure patient temperature.”

Cause

What was observed during troubleshooting.

Example:
“Probe connector pins were slightly bent; swapping with a known-good probe restored function.”

Resolution

What action was taken.

Example:
“Replaced probe with verified working unit; monitor reading confirmed accurate. Device returned to service.”

Helpful Details to Include

Final Thought

Proper troubleshooting prioritizes patient safety, systematically rules out external causes, and escalates appropriately. Documentation ensures trends are tracked, future errors are easier to resolve, and devices remain safe for patient care.

That is successful troubleshooting.