GE Healthcare Discovery XR656

Image Not Sending to PACS or Worklist Not Updating

On this page

Asset Type

Radiographic System

Manufacturer

GE Healthcare

Model

Discovery XR656

What This Guide Helps With

Troubleshooting PACS send failures, missing studies, delayed image transfer, or worklist refresh issues caused by network, configuration, or workflow problems.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Confirm the patient exam is complete or that imaging can safely continue before troubleshooting the Discovery XR656.

Action:

Expected outcome: Patient care continues safely while image transfer or worklist issues are investigated.

Why it matters: PACS and worklist problems can delay reporting, but troubleshooting should not interrupt safe patient imaging.

Verify the Reported Problem

Confirm whether the issue is with image sending, worklist updating, or both.

Action:

Expected outcome: The issue is narrowed to PACS transfer, worklist communication, patient/exam data, or a larger network problem.

If the issue is isolated to one patient or one accession, focus on order/data issues first.

Confirm Images Are Stored Locally

Verify the acquired images are still present on the Discovery XR656 workstation.

Action:

Expected outcome: Images are confirmed to be locally available and can potentially be resent.

If the images are missing locally, stop and escalate according to imaging downtime or data recovery policy.

Check the Send Queue or Network Queue

Look for failed, pending, or stuck DICOM transfers.

Action:

Expected outcome: Failed studies either resend successfully or provide useful error behavior for documentation.

If resend works, confirm image receipt in PACS and stop.

Verify Network Connection at the Room

Check the easiest physical network causes before assuming a software or PACS issue.

Action:

Expected outcome: The workstation has a stable physical network connection.

If a loose cable or disconnected jack is found and corrected, retry worklist refresh or PACS send.

Check Power and Status of Related Network Equipment

Confirm nearby network-dependent components are powered and normal.

Action:

Expected outcome: No obvious power or startup issue is affecting communication.

If the workstation is frozen or partially responsive, follow approved restart workflow or escalate before forcing shutdown.

Verify Modality Worklist Search Criteria

A worklist may appear “down” when the search filters are too narrow or incorrect.

Action:

Expected outcome: Valid orders appear once correct filters and location criteria are used.

If the order appears after correcting filters, the device is functioning and the issue is workflow/search-related.

Confirm Patient and Accession Data

Incorrect or incomplete exam data can prevent successful PACS association or routing.

Action:

Expected outcome: Patient and accession data match the active order.

If the image was acquired under incorrect data, follow radiology/PACS correction workflow. Do not guess or manually alter patient data outside approved policy.

Compare With Other Modalities or Rooms

Determine whether the issue is local to the Discovery XR656 or part of a larger system problem.

Action:

Expected outcome: The problem is identified as either device-specific or enterprise/network-wide.

If multiple systems are affected, escalation should go to PACS/RIS/network support rather than treating the XR656 as failed.

Check Destination Selection

Images may fail to send if routed to the wrong destination or if the expected destination is disabled.

Action:

Expected outcome: Images are routed to the correct PACS destination.

If the wrong destination was selected and the correct destination resend succeeds, confirm image receipt and stop.

Perform a Controlled Worklist Refresh

Refresh the worklist without disrupting active patient workflow.

Action:

Expected outcome: Worklist updates with expected scheduled exams.

If worklist remains unavailable but other network functions work, suspect RIS/worklist server, AE title, scheduled station mapping, or interface issue.

Perform a Controlled Resend Test

Use an existing completed study or approved test workflow to confirm PACS communication.

Action:

Expected outcome: The study transfers successfully and appears in PACS.

If the queue shows success but PACS cannot find the image, suspect routing, reconciliation, patient matching, or PACS-side issue.

Restart Only When Appropriate

A restart may clear a software communication hang, but it should be done carefully.

Action:

Expected outcome: Network services reload and normal communication returns.

If restart resolves the issue, document the behavior and final successful send/worklist update.

Escalate Configuration or DICOM Issues

If basic checks pass, the issue may involve DICOM configuration, AE title, IP address, port, routing rules, firewall rules, RIS mapping, or PACS interface behavior.

Action:

Expected outcome: The issue is escalated with enough information to avoid unnecessary troubleshooting delays.

If the Problem Persists

If local images are present, cables are connected, worklist search criteria are correct, patient/order data is valid, and resend attempts still fail, common external causes have been ruled out.

The Discovery XR656 should be:

Knowing when to stop is proper troubleshooting. PACS and worklist failures often involve shared systems, DICOM configuration, or network infrastructure that should not be changed casually at the device.

Clinical Use Tip

Do not troubleshoot PACS or worklist failures on an active patient if it delays imaging or creates confusion. Complete the exam safely, confirm images are stored locally, and move the patient or exam workflow to another system if needed. Patient care and therapy continuity should not depend on a system with unresolved image-transfer or worklist communication problems.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Radiology reported the GE Discovery XR656 was not sending completed images to PACS and the modality worklist was not updating."

Cause

What was observed during troubleshooting.

Example:
"Images were present locally, but the DICOM send queue showed failed transfers and the room network connection was found loose at the workstation."

Resolution

What action was taken.

Example:
"Reseated the network cable, confirmed link status, refreshed the worklist, resent the pending study to PACS, and verified image receipt with radiology staff."

Helpful Details to Include (If Known)

Final Thought

PACS and worklist issues should be approached logically. Protect the patient first, confirm the images are safe locally, check simple network and workflow causes, then escalate when the issue points beyond the radiographic system. Clear CCR documentation helps separate device problems from PACS, RIS, network, and order-entry issues.

That is successful troubleshooting.