GE Healthcare Aespire

PEEP Valve or APL Valve Pressure Is Inaccurate

On this page

Asset Type

Anesthesia Machine

Manufacturer

GE Healthcare

Model

Aespire

What This Guide Helps With

Troubleshoots inaccurate PEEP or APL pressure caused by circuit setup, leaks, valve positioning, accessories, pressure measurement, contamination, or external configuration.

Step-by-Step Troubleshooting

1. Protect the Patient and Establish Alternate Ventilation

Do not troubleshoot questionable PEEP or APL pressure while a patient depends on the machine.

If airway pressure cannot be trusted, provide ventilation using another verified anesthesia machine or appropriate alternate method.

Expected outcome: Patient support is maintained independently of the equipment being tested.

2. Confirm Which Pressure-Control Function Is Affected

Determine whether the complaint involves:

Test under controlled non-patient conditions.

Expected outcome: The affected function and reproducible symptom are clearly identified.

3. Inspect the Breathing Circuit

Inspect the breathing circuit for:

Correct or replace questionable disposable components.

Expected outcome: The circuit is complete, unobstructed, and leak-free externally. If correcting the circuit restores expected pressure behavior, troubleshooting can stop after verification.

4. Verify the APL Valve Position and Movement

For an APL-related complaint, confirm that the valve is set appropriately for the intended test and that the external control moves smoothly through its normal operating range.

Inspect for:

Do not disassemble the valve.

Expected outcome: The APL control moves freely and the observed pressure response changes appropriately with external adjustment.

5. Verify PEEP Settings and Ventilation Mode

For a PEEP-related complaint, confirm that:

Do not change protected configuration parameters unless authorized.

Expected outcome: The operating mode and user-adjustable settings match the intended test condition.

6. Check for Breathing-System Leaks

Perform the appropriate manufacturer-approved leak or pre-use test.

Inspect accessible seals, circuit interfaces, absorber connections, bag connection, and other external components when a leak is indicated.

Expected outcome: The breathing system passes the applicable leak evaluation. If an external leak is found and corrected, repeat the pressure test.

7. Use a Known-Good Test Lung and Circuit When Appropriate

Substitute a compatible known-good test lung, breathing circuit, reservoir bag, or other external accessory to isolate questionable disposables or accessories.

Expected outcome: Pressure behavior becomes stable with known-good accessories, identifying the replaced external component as the cause.

8. Verify Pressure With Approved Test Equipment

If the displayed or delivered pressure remains questionable, evaluate it using appropriate calibrated test equipment and the applicable manufacturer procedure.

Do not infer accuracy from display appearance alone.

Expected outcome: Independent testing confirms whether the machine's pressure behavior is acceptable or genuinely abnormal.

9. Inspect Accessible Valve Areas for External Contamination or Damage

Inspect externally accessible valve surfaces and adjacent breathing-system components for:

Follow approved cleaning and assembly instructions without performing deep disassembly.

Expected outcome: Accessible components are clean, properly assembled, and free from obvious mechanical restriction.

10. Perform Final Functional Verification

After any external correction, repeat the appropriate breathing-system test and verify:

Expected outcome: Pressure behavior is stable and verified with the appropriate test method. If successful, troubleshooting is complete.

11. Remove From Service if Pressure Cannot Be Verified

If pressure remains inaccurate, unstable, or inconsistent with the selected settings, stop troubleshooting and remove the machine from clinical use.

Expected outcome: Equipment with uncertain airway-pressure control is not returned to patient care.

If the Problem Persists

External circuit assembly, accessories, user-adjustable valve positioning, ventilation settings, leaks, and test setup have been ruled out.

The remaining condition may involve an internal valve mechanism, pressure-sensing system, pneumatic pathway, calibration issue, or other service-level fault.

The device should be:

After service, complete the required pressure, ventilation, alarm, and anesthesia-machine return-to-service testing before clinical use.

Knowing when to stop external troubleshooting is proper troubleshooting.

Clinical Use Tip

Any anesthesia machine with questionable airway-pressure control should be replaced before patient use rather than evaluated during active ventilation.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Anesthesia staff reported that manual ventilation pressure on the Aespire did not respond normally to APL valve adjustment."

Cause

What was observed during troubleshooting.

Example:
"Clinical Engineering found the reservoir bag connection partially loose, producing an external breathing-system leak and unstable pressure."

Resolution

What action was taken.

Example:
"Reseated the reservoir bag connection, repeated the breathing-system leak and pressure checks, and verified normal APL response during functional testing."

Helpful Details to Include (If Known)

Final Thought

Airway-pressure concerns require verified testing, not assumptions. Start with circuit assembly, leaks, accessories, controls, and settings, confirm the pressure using appropriate test equipment, and escalate when reliable pressure control cannot be demonstrated.

That is successful troubleshooting.