GE Healthcare Aisys CS2

High Airway Pressure Alarm

On this page

Asset Type

Anesthesia Machine

Manufacturer

GE Healthcare

Model

Aisys CS2

What This Guide Helps With

Troubleshooting elevated airway pressure caused by circuit restrictions, incorrect settings, patient-side accessories, expiratory obstruction, scavenging problems, or breathing-system components.

Step-by-Step Troubleshooting

1. Ensure Patient Safety First

Do not perform extended troubleshooting while the Aisys CS2 is supporting an active patient.

If a high-airway-pressure alarm occurs during clinical use:

Expected outcome: Patient care no longer depends on equipment with unresolved elevated airway pressure.

Continue equipment troubleshooting only after the machine is safely removed from patient use.

2. Confirm the Exact Alarm Condition

Record the complete alarm message and determine:

Review the alarm history when available.

Expected outcome: The failure pattern is clearly identified before components are changed.

3. Check Ventilator and Alarm Settings

Verify that the selected ventilation parameters are appropriate for testing, including:

Excessive tidal volume, inspiratory pressure, PEEP, or inspiratory time can cause airway pressure to reach the configured alarm limit.

Use a test lung and approved test settings rather than changing clinical settings on a patient.

Expected outcome: Pressure remains below the configured limit during controlled testing.

If correcting an inappropriate setting resolves the alarm, complete the required system checkout and stop.

4. Inspect the Patient Breathing Circuit

Examine the entire circuit from the machine to the test lung for:

Straighten the circuit, drain moisture safely, or replace the circuit with a known-good compatible set.

Expected outcome: Gas flows freely through the complete breathing circuit without excessive resistance.

If replacing the circuit resolves the alarm, perform a checkout and stop.

5. Remove External Accessories

Temporarily remove nonessential patient-side accessories, including:

Inspect each accessory for contamination, liquid accumulation, blockage, or incorrect orientation.

Reconnect only the minimum components needed for testing.

Expected outcome: Airway pressure remains stable with unnecessary resistance removed.

If one accessory recreates the alarm, replace that accessory and stop after successful verification.

6. Check the Test Lung and Simulated Airway

Confirm that the test lung:

Test with a second known-good test lung when available.

Expected outcome: The machine ventilates a verified test load without reaching the high-pressure limit.

7. Inspect the Inspiratory and Expiratory Connections

Verify that the inspiratory and expiratory hoses are connected to the correct ports and are fully seated.

Inspect the external breathing-system ports for:

Do not insert tools into valves or internal gas pathways.

Expected outcome: Both breathing-system pathways are correctly connected and externally unobstructed.

8. Check the Expiratory Path and Scavenging System

Look for external conditions that could prevent normal exhalation:

Temporarily evaluate the machine with a verified scavenging setup according to facility procedures.

Expected outcome: Exhaled gas exits normally, and pressure returns toward baseline between breaths.

9. Check Manual Ventilation Operation

With a test lung installed, place the system in manual ventilation and verify:

The APL valve controls pressure during manual ventilation and is adjustable up to 70 cmH₂O on the Aisys CS2.

Expected outcome: Manual ventilation produces controlled pressure that decreases normally during exhalation.

If pressure remains trapped in manual ventilation, remove the machine from service.

10. Inspect Accessible Breathing-System Components

With the device powered down and removed from clinical use, visually inspect user-removable breathing-system components for:

Reseat components only as permitted by the operator documentation and Clinical Engineering procedures.

Do not disassemble internal valves, manifolds, or ventilator assemblies during initial troubleshooting.

Expected outcome: All accessible components are clean, intact, and correctly installed.

11. Restart and Perform the Required Checkout

Reconnect the machine to verified electrical and gas supplies, then:

GE instructs users to follow the Aisys CS2 reference documentation for complete operation and checkout requirements.

Expected outcome: The machine completes checkout and ventilates the test lung without an unexpected high-airway-pressure alarm.

If the checkout passes and the alarm cannot be reproduced, document all findings and return the device according to facility policy.

If the Problem Persists

If the alarm continues with verified settings, a known-good circuit, minimal accessories, a verified test lung, and an unobstructed scavenging setup, common external causes have been ruled out.

The condition may involve an internal:

The device should be:

Do not continue replacing external parts or repeatedly cycling the machine when airway pressure cannot be reliably controlled. Knowing when to stop and escalate is proper troubleshooting.

Clinical Use Tip

Never troubleshoot a high-airway-pressure condition by repeatedly ventilating an active patient. The anesthesia provider must first evaluate patient, airway, positioning, and clinical causes. Equipment testing should continue only with a verified test lung after the patient is safely supported elsewhere.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Clinical staff reported repeated high-airway-pressure alarms during mechanical ventilation on the GE Healthcare Aisys CS2."

Cause

What was observed during troubleshooting.

Example:
"Inspection found a moisture-saturated bacterial filter restricting flow at the patient connection and causing peak pressure to reach the configured limit."

Resolution

What action was taken.

Example:
"Replaced the restricted filter, installed a known-good circuit, completed the system checkout, and verified normal pressure and ventilation using a test lung."

Helpful Details to Include (If Known)

Final Thought

High airway pressure may result from the patient, airway, breathing circuit, settings, accessories, scavenging system, or the anesthesia machine itself. Patient safety comes first, followed by controlled external checks and testing with known-good components. When pressure remains abnormal after external causes are eliminated, remove the Aisys CS2 from service and escalate for repair. Clear CCR documentation preserves the reported condition, findings, corrective action, and final equipment status.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide