GE Healthcare Aisys CS2

O2 Sensor Calibration or FiO2 Measurement Error

On this page

Asset Type

Anesthesia Machine

Manufacturer

GE Healthcare

Model

Aisys CS2

What This Guide Helps With

Troubleshooting failed O2-cell calibration, unavailable oxygen measurement, or inaccurate FiO2 readings caused by setup, gas supply, sampling, sensor, or module issues.

Step-by-Step Troubleshooting

1. Ensure Patient Safety First

Do not calibrate, remove, exchange, or test oxygen-measurement components while the Aisys CS2 is supporting an active patient.

If the displayed FiO2 is missing, unstable, or inconsistent with the selected gas mixture:

Expected outcome: Patient care no longer depends on equipment with questionable oxygen measurement.

Continue troubleshooting only after the machine is removed from active patient use.

2. Confirm Which Oxygen Measurement Is Affected

Determine whether the reported problem involves:

Record the complete displayed alarm, advisory, or checkout message.

Expected outcome: The affected oxygen-measurement path is clearly identified before components are disturbed.

The Aisys CS2 may use a circuit O2 cell and, depending on configuration, an airway module capable of measuring oxygen and other respiratory gases.

3. Verify the Oxygen and Air Supplies

Confirm that:

Low, unstable, or incorrectly connected gas supplies can create an FiO2 value that does not match the selected fresh-gas settings.

Expected outcome: Gas supplies are stable, correctly connected, and free of related alarms.

If correcting the gas supply restores an accurate and stable FiO2 reading, complete checkout and return the machine to service.

4. Allow the Machine to Complete Startup

Perform a normal shutdown and restart while the machine is disconnected from patient use.

Allow the Aisys CS2 to:

Do not repeatedly cycle power if the same error immediately returns.

Expected outcome: A temporary startup or communication problem clears and the oxygen-measurement components are recognized.

5. Inspect the Circuit O2 Cell Externally

If the machine is equipped with a circuit O2 cell:

Do not open the sensor body or attempt internal repair.

Expected outcome: The O2 cell is correctly installed, clean, dry, and externally undamaged.

If reseating a loose connection restores normal operation, run the required checkout before returning the machine to service.

6. Inspect the Breathing Circuit and Gas Path

Confirm that:

Expected outcome: The oxygen sensor is exposed to a stable and representative gas mixture.

7. Inspect the Airway Gas Module and Sampling Components

If FiO2 is obtained through the airway gas module:

The airway module measures and monitors gases delivered to and exhaled by the patient; sampling problems can therefore affect the displayed oxygen value.

Expected outcome: Sampling is unobstructed, the module is recognized, and its oxygen value is stable.

8. Run the Circuit O2 Cell Test

From the Checkout menu, perform the appropriate Circuit O2 Cell test by following the on-screen instructions.

During the room-air portion:

GE’s Aisys CS2 training material specifically instructs allowing the reading to stabilize before completing the Circuit O2 Cell test.

Expected outcome: The Circuit O2 Cell test completes successfully.

If the test passes and the oxygen reading remains accurate during functional testing, stop troubleshooting.

9. Perform the Approved O2-Cell Calibration

If calibration is requested or the Circuit O2 Cell test does not pass:

Expected outcome: Calibration completes without an error and the displayed oxygen value agrees with the reference condition.

The Aisys CS2 checkout documentation directs calibration of the O2 cell when necessary following evaluation of the stabilized reading.

10. Compare the Reading With an Independent Analyzer

Using an approved, calibrated oxygen analyzer:

Do not rely solely on the selected fresh-gas setting as proof of delivered FiO2.

Expected outcome: The displayed oxygen concentration is stable, responsive, and within the facility’s approved performance tolerance.

11. Substitute Approved External Components When Available

If permitted by facility policy and manufacturer guidance:

Do not use unverified sensors or modules from equipment with unresolved problems.

Expected outcome: The faulty external sensor, module, or sampling component is isolated.

If replacement resolves the issue, complete the full checkout and document the component exchanged.

12. Complete a Full Checkout

After correction:

GE identifies the Circuit O2 Cell test as part of the Aisys CS2 checkout when that cell is present.

Expected outcome: All required tests pass and oxygen measurement remains accurate and stable.

If the Problem Persists

If calibration repeatedly fails, FiO2 remains inaccurate, or the oxygen reading is unavailable after external components, gas supplies, sampling parts, connections, and checkout conditions have been verified, common external causes have been ruled out.

The problem may involve an internally failed sensor circuit, module interface, wiring connection, signal-processing component, or gas-measurement assembly.

The device should be:

Do not proceed with undocumented internal adjustments or board-level repairs. Knowing when to stop and escalate is proper troubleshooting.

Clinical Use Tip

Never troubleshoot questionable FiO2 measurement on an active patient. Establish independent oxygen monitoring or move the patient to another verified anesthesia machine before testing, calibration, or component exchange.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Anesthesia staff reported that the Aisys CS2 displayed an unstable FiO2 value and would not complete the Circuit O2 Cell test."

Cause

What was observed during troubleshooting.

Example:
"Inspection found moisture contamination around the circuit O2 cell connection, preventing the reading from stabilizing during checkout."

Resolution

What action was taken.

Example:
"Removed the machine from service, dried and reseated the approved O2 cell, completed calibration, verified readings with an independent oxygen analyzer, and passed the full checkout."

Helpful Details to Include (If Known)

Final Thought

Accurate oxygen measurement is essential to safe anesthesia delivery. Begin with patient protection, verify gas supplies and external components, use stable calibration conditions, confirm performance independently, and escalate when the measurement cannot be trusted. Clear CCR documentation preserves both patient safety and repair history.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide