GE Healthcare Aisys CS2

Scavenging / AGSS Restriction or Evacuation Fault

On this page

Asset Type

Anesthesia Machine

Manufacturer

GE Healthcare

Model

Aisys CS2

What This Guide Helps With

Troubleshooting restricted, inadequate, or excessive waste-gas evacuation caused by hoses, connections, suction settings, receiving systems, or external scavenging components.

Step-by-Step Troubleshooting

1. Ensure Patient Safety First

Do not perform extended scavenging troubleshooting while the Aisys CS2 is supporting an active patient.

If scavenging performance becomes questionable during a procedure:

Expected outcome: Patient care continues safely without relying on equipment with unresolved scavenging performance.

Continue troubleshooting only after the machine has been removed from active patient use.

2. Confirm the Exact Failure Condition

Determine whether the reported issue involves:

Record the complete alarm, advisory, or observed behavior.

Expected outcome: The failure condition is clearly identified before components are changed.

3. Inspect the Scavenging Hose and Connections

Check the scavenging transfer hose from the anesthesia machine to the facility waste-gas disposal connection.

Verify that the hose is:

Straighten or reconnect the hose as needed.

Expected outcome: Waste gas has an open and properly connected path to the disposal system.

If normal scavenging operation returns, complete an operational checkout and stop troubleshooting.

4. Check the Facility Evacuation Connection

Confirm that the scavenging hose is connected to the correct wall outlet or waste anesthetic gas disposal connection.

Inspect the outlet for:

When permitted, test the machine using another verified scavenging outlet or approved test connection.

Expected outcome: The facility evacuation source is available and appropriate for the anesthesia machine.

If the problem follows the wall outlet, notify Facilities or the department responsible for the waste-gas disposal system.

5. Inspect the Scavenging Interface Externally

Inspect all externally accessible scavenging interface components.

Look for:

Do not disassemble internal valves or regulated assemblies unless authorized by the manufacturer’s service documentation.

Expected outcome: External scavenging components are clean, correctly assembled, and able to move normally.

6. Evaluate the Reservoir Response

With the machine off patient and connected to a verified test circuit, operate the system under controlled conditions.

Observe the scavenging reservoir:

Do not intentionally create unsafe pressure within the breathing system.

Expected outcome: The reservoir responds to waste-gas flow without remaining overfilled or continuously collapsed.

7. Check Adjustable Scavenging Flow Controls

If the installed scavenging configuration includes an externally adjustable flow indicator or control, verify that it is set according to facility practice and the approved operating instructions.

Check for:

Do not adjust facility vacuum regulators without authorization.

Expected outcome: Evacuation flow is sufficient to remove waste gas without applying excessive negative pressure.

8. Inspect Connected Accessories

Check accessories that may affect scavenging performance, including:

Remove unnecessary adapters and replace questionable accessories with verified compatible components.

Expected outcome: No external accessory is restricting, diverting, or incorrectly applying suction to the scavenging pathway.

9. Rule Out a Breathing-System Restriction

Inspect the breathing circuit and externally accessible breathing-system components for conditions that may resemble a scavenging fault.

Check for:

Expected outcome: Elevated pressure or poor gas evacuation is not being caused by a separate breathing-circuit restriction.

10. Disconnect the Facility Evacuation for Comparison

With the machine off patient and in a controlled test environment, disconnect the scavenging hose from the facility disposal connection when permitted by facility policy.

Observe whether:

Limit the test duration and prevent unnecessary release of anesthetic vapor.

Expected outcome: A significant improvement when disconnected suggests excessive facility suction or a downstream disposal-system problem.

11. Restart and Complete the System Checkout

After correcting external connections or replacing accessories:

Expected outcome: The machine completes checkout without scavenging-related abnormalities.

If the checkout passes and evacuation remains stable, return the device to service according to departmental policy.

If the Problem Persists

If the scavenging hose, facility outlet, external interface, reservoir, accessories, breathing circuit, and operating setup have been verified, the fault may involve an internal scavenging valve, interface assembly, pressure-relief component, obstruction, or other internal pneumatic problem.

The device should be:

Do not continue making adjustments when proper evacuation or breathing-system pressure cannot be verified. Knowing when to stop and escalate is proper troubleshooting.

Clinical Use Tip

Never troubleshoot unresolved scavenging or breathing-system pressure problems on an active patient. Move the patient to another verified anesthesia machine before performing extended testing.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Anesthesia staff reported that the Aisys CS2 scavenging reservoir remained full and waste-gas evacuation appeared inadequate."

Cause

What was observed during troubleshooting.

Example:
"The AGSS transfer hose was found kinked behind the anesthesia machine, restricting flow to the facility evacuation outlet."

Resolution

What action was taken.

Example:
"Repositioned and inspected the hose, verified normal reservoir movement, completed system checkout and leak testing, and returned the machine to service."

Helpful Details to Include (If Known)

Final Thought

Scavenging faults can involve the anesthesia machine, connecting hose, facility evacuation system, or breathing-circuit setup. Patient safety, external-first troubleshooting, controlled testing, appropriate escalation, and complete CCR documentation are essential.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide