GE Healthcare AMX Navigate

Emergency Stop, Bumper, or Anti-Collision Sensor Prevents Movement

On this page

Asset Type

Mobile X-Ray System

Manufacturer

GE Healthcare

Model

AMX Navigate

What This Guide Helps With

Powered movement remains inhibited because an emergency stop, bumper, collision input, or related safety function appears active.

Step-by-Step Troubleshooting

1. Protect the Patient and Preserve the Safety Function

Do not bypass, defeat, tape down, jumper, or otherwise disable an emergency-stop, bumper, or collision-protection device to restore movement.

Expected outcome: The transport safety system remains functional throughout troubleshooting.

2. Confirm the Exact Movement Inhibit

Determine whether:

Expected outcome: The affected safety function is identified as closely as possible.

3. Verify Emergency-Stop Position

Inspect the emergency-stop control and verify it is in its normal released position according to the device's normal operating design.

Expected outcome: The emergency stop is not intentionally engaged.

4. Restart Normally if Appropriate

After confirming it is safe and the emergency stop has been released, perform a normal system restart if required by standard operation.

Expected outcome: A temporary latched safety condition clears and normal drive operation returns. If so, continue to final verification.

5. Inspect Bumpers and Collision Surfaces

Inspect accessible bumpers and anti-collision surfaces for:

Expected outcome: All external collision surfaces are unobstructed and return freely to their normal position.

6. Remove External Obstructions

Remove only obvious external debris or objects that are preventing a bumper or sensor surface from returning to normal.

Expected outcome: The affected safety surface moves freely and the movement inhibit clears.

7. Compare Movement Directions

In a controlled open area, determine whether forward, reverse, left, and right movement behave differently.

Expected outcome: The test identifies whether the inhibit is associated with a particular collision-protection zone or direction.

8. Check for Recent Impact

Ask whether the unit struck a wall, doorway, bed, or other object before the problem began. Inspect externally for displacement or damage.

Expected outcome: A potential mechanical cause is identified without opening the safety system.

9. Verify Safety-Device Operation

After external correction, gently verify that applicable bumpers or safety surfaces respond as expected during an approved functional test. Do not perform any test that risks uncontrolled movement.

Expected outcome: Normal movement is available when the safety device is released, and movement is appropriately inhibited when the device is actuated.

10. Verify Controlled Transport

Test forward, reverse, turning, and stopping in an open area.

Expected outcome: Powered transport is reliable and the safety functions operate normally. Troubleshooting can stop.

11. Escalate a Persistent Safety Inhibit

If the emergency-stop, bumper, or collision system remains active, intermittent, physically damaged, or fails to inhibit motion correctly, remove the unit from service.

Expected outcome: The system is not used with a defeated or unreliable collision-protection function.

If the Problem Persists

Common external obstructions and control-position causes have been ruled out. Remaining possibilities may involve bumper switches, collision sensors, internal wiring, alignment, drive-control safety logic, or another service-level safety circuit.

The device should be:

After repair, verify emergency-stop operation, all applicable collision-protection functions, powered movement, steering, and stopping before return to service.

Knowing when to stop external troubleshooting is proper troubleshooting.

Clinical Use Tip

Never trade collision protection for mobility; a mobile X-ray system must have both predictable powered movement and fully functional safety interlocks.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Radiology reported the AMX Navigate would power on normally but powered drive remained inhibited after transport through a crowded patient room."

Cause

What was observed during troubleshooting.

Example:
"Clinical Engineering found an external object lodged against a lower bumper, preventing the bumper from returning to its normal position."

Resolution

What action was taken.

Example:
"Removed the obstruction, verified free bumper movement, confirmed proper movement inhibit when actuated, and completed normal forward, reverse, steering, and stopping tests."

Helpful Details to Include (If Known)

Final Thought

Emergency-stop and collision systems are safety devices, not nuisances to work around. Verify simple mechanical causes first, never bypass the protection, and escalate whenever the safety function or powered movement remains unreliable.

That is successful troubleshooting.