What This Guide Helps With
The system will not prepare or expose, exposure is inhibited, or the exposure control does not complete the expected sequence.
Step-by-Step Troubleshooting
1. Protect the Patient and Prevent Unintended Exposure
Do not repeatedly operate the exposure control around a patient while determining why the sequence is failing. Use a controlled test environment and follow facility radiation-safety procedures.
Expected outcome: Troubleshooting can proceed without unnecessary radiation exposure.
2. Confirm the Exact Sequence Failure
Determine whether:
- Prep does not begin
- Prep occurs but exposure does not occur
- The switch feels mechanically abnormal
- The system displays an inhibit
- The sequence fails only in one operating mode
- The problem is intermittent
Record any displayed status or inhibit message exactly.
Expected outcome: The failure point in the prep/expose sequence is established.
3. Verify Normal System Readiness
Confirm the AMX Navigate completes normal startup, has adequate battery charge, and is not reporting another condition that intentionally prevents exposure.
Expected outcome: The system reaches its normal ready state.
4. Verify the Detector Is Available
Confirm the selected detector is connected, recognized, ready, and appropriately selected for the examination. A detector communication problem may prevent normal exposure workflow.
Expected outcome: The intended detector is available for acquisition.
5. Verify Examination and Acquisition Setup
Confirm a valid examination or approved test workflow has been selected and that required acquisition selections are complete. Do not alter protected configuration parameters simply to bypass an inhibit.
Expected outcome: The system has the required normal workflow information to permit exposure.
6. Inspect the Prep/Expose Control
Inspect the hand switch or exposure control and its accessible cable for:
- Cracks
- Pinching
- Cuts
- Loose connector seating
- Bent or damaged connector areas
- Sticking buttons
- Abnormal tactile response
Expected outcome: The switch and cable are physically intact and securely connected.
7. Reseat Accessible Connections
With the system in a safe state, reseat the normal external exposure-control connection where permitted.
Expected outcome: The exposure control is securely connected. If the sequence is restored, continue to final verification.
8. Substitute a Known-Good Approved Exposure Control
If a compatible known-good hand switch or exposure control is available, substitute it without opening the system.
Expected outcome: If the known-good control restores operation, the original control or cable is isolated as the likely cause.
9. Check for Other Safety Inhibits
Verify that emergency-stop, collision, detector, generator-readiness, or other normal system status conditions are not preventing operation. Do not bypass a safety interlock.
Expected outcome: No active safety condition is blocking the sequence.
10. Perform an Approved Functional Exposure Test
Using facility-approved radiation-safety procedures, a suitable test object, and no patient, verify the prep and exposure sequence.
Expected outcome: Prep initiates correctly and the exposure completes normally. If successful, troubleshooting can stop.
11. Escalate an Unresolved Exposure Inhibit
If prep/expose remains unreliable after the switch, detector, workflow, system readiness, and accessible connections have been checked, remove the system from clinical use.
Expected outcome: A unit with unreliable exposure control is not used on patients.
If the Problem Persists
Common external causes have been ruled out. Remaining possibilities may involve exposure-control circuitry, generator readiness, safety-interlock logic, internal wiring, detector communication, system configuration, or other service-level faults.
The system should be:
- Removed from service
- Labeled Out of Service
- Sent for repair or bench evaluation
- Evaluated using appropriate GE Healthcare service documentation and approved X-ray test equipment
- Repaired or configured only by qualified personnel
Return-to-service testing should verify normal prep, exposure initiation, exposure termination, image acquisition, and applicable radiation-safety functions.
Knowing when to stop external troubleshooting is proper troubleshooting.
Clinical Use Tip
Never bypass an exposure inhibit or safety interlock to obtain an image; identify why the system is preventing exposure.
Work Order Documentation (CCR Method)
CCR = Complaint, Cause, Resolution
Complaint
What was reported by the clinical staff.
Example:
"Radiology reported the AMX Navigate would enter prep but would not complete an exposure when the hand switch was pressed."
Cause
What was observed during troubleshooting.
Example:
"Clinical Engineering found damage at the exposure-switch cable that caused intermittent operation."
Resolution
What action was taken.
Example:
"Replaced the approved exposure control, verified reliable prep and expose operation using a controlled non-patient test, and returned the unit to service."
Helpful Details to Include (If Known)
- Exact inhibit or status message
- Whether prep occurred
- Whether exposure occurred
- Hand-switch condition
- Cable and connector condition
- Detector availability
- Battery state
- Emergency-stop status
- Known-good switch result
- Functional exposure-test result
- Final device status
Final Thought
Exposure problems require disciplined troubleshooting because the system may be intentionally inhibiting radiation for a valid reason. Verify external controls, detector readiness, workflow, and safety conditions before escalating to generator or internal-system evaluation.
That is successful troubleshooting.