GE Healthcare AMX Navigate

Image Has Detector Lines, Lag, Nonuniformity, or Calibration Artifacts

On this page

Asset Type

Mobile X-Ray System

Manufacturer

GE Healthcare

Model

AMX Navigate

What This Guide Helps With

Acquired images show lines, residual images, uneven response, fixed artifacts, or other detector-related image-quality abnormalities.

Step-by-Step Troubleshooting

1. Protect the Patient From Unnecessary Repeat Exposure

Do not repeatedly expose a patient to determine whether an image artifact is reproducible. Move to an alternate verified detector or system when clinically necessary.

Expected outcome: Image-quality troubleshooting does not create unnecessary patient dose.

2. Characterize the Artifact

Determine whether the artifact is:

Record whether the artifact remains in the same detector location.

Expected outcome: The artifact pattern is clearly described.

3. Inspect the Detector Surface

Check the detector and any protective cover for:

Expected outcome: No external material is creating an image artifact.

4. Verify Detector Positioning

Confirm the detector is correctly positioned and not being flexed, compressed, or improperly supported during testing.

Expected outcome: Mechanical positioning is not contributing to the image abnormality.

5. Verify Detector Communication and Battery Condition

Confirm the detector is stably connected and has adequate battery power.

Expected outcome: Communication or low-power instability is not contributing to the problem.

6. Acquire an Approved Test Image

Using appropriate test material and no patient, obtain an image under controlled conditions.

Expected outcome: The artifact is either reproduced or ruled out without involving a patient.

7. Rotate or Reposition the Detector

Change detector orientation relative to the test object while maintaining an appropriate controlled setup.

Expected outcome: The test helps localize the artifact source.

8. Compare With a Known-Good Detector

Use a compatible known-good detector on the same AMX Navigate.

Expected outcome: If the artifact disappears, the original detector is isolated as the likely source. If it remains, evaluate the system or acquisition path.

9. Perform Approved Calibration or Quality-Control Functions

If normal Clinical Engineering responsibility includes approved detector calibration or quality-control procedures, perform only those documented procedures using the proper conditions and test equipment.

Do not improvise calibration steps or alter protected service parameters.

Expected outcome: The detector completes the approved procedure and subsequent test images are uniform.

10. Verify Image Quality After Correction

Acquire a controlled non-patient test image and confirm that lines, lag, or nonuniformity are no longer present.

Expected outcome: Image quality is acceptable and repeatable. Troubleshooting can stop.

11. Escalate Persistent Image Artifacts

If the artifact follows a detector, remains after approved calibration, or cannot be isolated through external testing, remove the affected detector or system from service.

Expected outcome: Equipment producing unreliable diagnostic images is not returned to patient use.

If the Problem Persists

Common external and acquisition causes have been ruled out. Remaining possibilities may involve detector electronics, detector calibration, internal image-processing functions, generator consistency, software, or another service-level imaging component.

The affected equipment should be:

Return-to-service testing should confirm acceptable image quality over appropriate controlled test conditions.

Knowing when to stop external troubleshooting is proper troubleshooting.

Clinical Use Tip

Use a test object rather than a patient when determining whether an image artifact follows the detector.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Radiology reported a fixed vertical line appeared in the same location on images acquired with one AMX Navigate detector."

Cause

What was observed during troubleshooting.

Example:
"Clinical Engineering reproduced the artifact on a non-patient test image and confirmed it followed the detector when compared with a known-good detector."

Resolution

What action was taken.

Example:
"Removed the affected detector from service for repair and verified artifact-free imaging on the AMX Navigate using the known-good detector."

Helpful Details to Include (If Known)

Final Thought

Image artifacts should be isolated methodically without creating unnecessary repeat exposures. Confirm the detector, positioning, environment, and known-good comparison before assuming an internal imaging-system failure.

That is successful troubleshooting.