On this page
Asset Type
Manufacturer
Model
What This Guide Helps With
The main battery does not appear to charge, discharges quickly, or provides insufficient runtime for powered transport and imaging workflow.
Step-by-Step Troubleshooting
1. Protect Clinical Operations
Do not allow the unit to become stranded during patient transport or imaging. If battery operation is unreliable, move clinical work to another verified mobile X-ray system.
Expected outcome: Patient care does not depend on an unreliable battery.
2. Confirm the Exact Battery Complaint
Determine whether:
- The battery does not charge at all
- Charging starts but stops
- Charge indication appears inaccurate
- Runtime is short
- Powered drive fails before imaging functions
- The unit shuts down unexpectedly
- The issue occurs only when unplugged
Expected outcome: The battery-related symptom is clearly defined.
3. Verify the AC Power Source
Inspect the power cord and plug for damage. Connect the unit to a known-good approved receptacle. Verify the receptacle according to facility practice if AC power is questionable.
Expected outcome: Reliable AC power is available to the system.
4. Inspect Power-Cord and Inlet Condition
Inspect accessible power connections for:
- Loose fit
- Burn marks
- Heat damage
- Bent contacts
- Cracked insulation
- Evidence of arcing
- Strain-relief damage
Remove the unit from service if unsafe electrical damage is present.
Expected outcome: The AC power path is physically intact.
5. Verify Charging Indication
With the unit connected to known-good AC power, observe normal charging indicators or battery status.
Expected outcome: The system recognizes AC input and indicates charging as expected.
6. Allow Normal Charging Opportunity
Confirm the unit has actually been connected to AC power long enough under normal operating conditions to establish whether the battery accepts a charge. Avoid judging battery condition immediately after a heavily discharged unit is plugged in.
Expected outcome: Battery state increases normally or the inability to charge is confirmed.
7. Check for Abnormal Heat, Odor, or Battery-Related Damage
Inspect accessible areas for unusual heat, swelling, odor, discoloration, or other signs of battery or charging-system distress.
Expected outcome: No unsafe physical condition is present. If abnormal heat, odor, or damage exists, disconnect the unit when safe and remove it from service.
8. Compare Stationary and Powered-Drive Runtime
When safe, observe whether battery depletion is unusually rapid during powered transport compared with stationary operation. Excessive mechanical resistance may increase drive demand.
Expected outcome: The test helps distinguish general battery weakness from excessive transport load.
9. Check Wheels and Casters for Excessive Resistance
Inspect for debris, binding, damaged casters, or dragging components that could force the drive system to work harder.
Expected outcome: The unit rolls freely without obvious external resistance.
10. Perform a Controlled Battery Functional Check
After charging, operate the system under an approved non-patient workflow and observe whether battery status remains stable during normal transport and system operation.
Expected outcome: Battery operation is stable and sufficient for normal workflow. If so, troubleshooting can stop.
11. Escalate Persistent Charging or Runtime Failure
If a known-good AC source, power cord, external mechanical condition, and normal charging opportunity do not correct the problem, remove the unit from service for battery and charging-system evaluation.
Expected outcome: A system with unreliable battery capacity is not returned to mobile clinical use.
If the Problem Persists
Common external causes have been ruled out. Remaining possibilities may involve battery capacity, internal charging electronics, battery monitoring, high internal current draw, drive-system load, wiring, or service-level power management.
The device should be:
- Removed from service
- Labeled Out of Service
- Sent for repair or bench evaluation
- Evaluated using appropriate GE Healthcare documentation and approved test equipment
- Repaired or configured only by qualified personnel
After repair, verify AC charging, battery indication, startup, controlled powered movement, and appropriate battery-supported operation before return to service.
Knowing when to stop external troubleshooting is proper troubleshooting.
Clinical Use Tip
Return the mobile X-ray unit to approved charging power when available so unexpected battery depletion does not interrupt bedside imaging.
Work Order Documentation (CCR Method)
CCR = Complaint, Cause, Resolution
Complaint
What was reported by the clinical staff.
Example:
"Radiology reported the AMX Navigate battery discharged rapidly and powered transport runtime was significantly shorter than expected."
Cause
What was observed during troubleshooting.
Example:
"Clinical Engineering found heavy debris around a caster that increased rolling resistance and placed additional demand on the powered-drive system."
Resolution
What action was taken.
Example:
"Removed the external obstruction, verified free caster movement, charged the system, and confirmed stable battery-supported transport and normal system operation."
Helpful Details to Include (If Known)
- Battery indication at start
- AC charging indication
- Receptacle tested
- Power-cord condition
- Charging duration
- Shutdown behavior
- Runtime complaint
- Wheel and caster condition
- Abnormal heat or odor
- Battery status before and after testing
- Final functional result
- Final device status
Final Thought
Verify AC supply, physical connections, charging indications, and unnecessary mechanical load before condemning the battery. Battery-supported transport must be predictable before a mobile X-ray system is returned to clinical service.
That is successful troubleshooting.