GE Healthcare Avance

Electronic Flow Control Malfunction

On this page

Asset Type

Anesthesia Machine

Manufacturer

GE Healthcare

Model

Avance

What This Guide Helps With

Troubleshooting unavailable, inaccurate, or unresponsive electronic fresh-gas flow control caused by supply, control, configuration, or gas-delivery system problems.

Step-by-Step Troubleshooting

1. Ensure Patient Safety First

Do not perform extended flow-control troubleshooting while the Avance is supporting an active patient.

If fresh-gas flow cannot be adjusted, displayed flow is questionable, or gas delivery is unstable:

Expected: Patient care no longer depends on an anesthesia machine with uncertain fresh-gas delivery.

If the patient has been safely transferred or the machine is not in use, continue troubleshooting.

2. Confirm the Exact Failure Condition

Record the complete displayed alarm or message and determine whether the problem involves:

Expected: The complaint is clearly defined and can be reproduced safely on a test circuit.

3. Check for Alternate O2 Operation

Inspect the Alternate O2 control and determine whether it has activated or was manually turned on.

Certain system failures may cause Alternate O2 to activate automatically. When active, normal electronic gas-control operation may be unavailable or altered.

Expected: Alternate O2 status matches the selected operating condition.

If the control was unintentionally active and normal electronic flow control returns after proper reset and checkout, document the finding and stop.

4. Verify AC Power and Startup Condition

Confirm that:

Electronic gas control depends on stable system power and functional control electronics.

Expected: The Avance remains powered from AC mains and completes startup normally.

If power is unstable, correct the external power problem and repeat the manufacturer-approved checkout.

5. Verify Pipeline Gas Supplies

Check the connected O2, Air, and N2O pipeline supplies as equipped.

Expected: Each required gas supply is correctly connected and available at the proper pressure.

If flow control returns after correcting a pipeline connection or pressure problem, complete a full checkout, document the repair, and stop.

6. Check Backup Cylinder Status

Inspect installed backup cylinders without interrupting any required gas supply.

Expected: Backup cylinders are correctly installed and are not creating an abnormal supply condition.

7. Inspect External Controls and Display Response

Operate the fresh-gas flow controls in a controlled test environment.

Expected: Each control input produces an immediate and consistent display response.

If the controls are physically damaged or intermittently unresponsive, remove the machine from service for repair.

8. Review Gas-Selection and Flow Settings

Confirm that the selected carrier gas, oxygen concentration, and total fresh-gas flow are appropriate for the test.

Expected: Displayed gas settings respond logically and remain stable after adjustment.

9. Inspect the Common Gas Outlet and Breathing-System Connections

Check the external gas-delivery path.

GE identifies selectable common-gas-outlet components as part of the anesthesia machine’s breathing-circuit gas-delivery system.

Expected: Fresh gas is directed through the intended outlet without an external restriction or misconnection.

10. Inspect Vaporizers and External Accessories

Confirm that:

Expected: External accessories do not interfere with gas delivery or system checkout.

If removing a suspect accessory restores normal operation, quarantine the accessory and repeat the complete checkout.

11. Perform the Approved Machine Checkout

With no patient connected:

The Avance breathing system is designed to support rapid control of fresh-gas flows, so delayed, unstable, or unavailable adjustment is not normal operation.

Expected: The system passes checkout, flow settings remain stable, and measured gas delivery agrees with the selected settings within applicable specifications.

If the machine passes all testing and the complaint cannot be reproduced, document the tests performed and return it to service only according to facility policy.

12. Compare With Known-Good External Components

When permitted, substitute known-good external items one at a time:

Repeat the relevant functional check after each substitution.

Expected: A defective external component is isolated without opening the machine.

If a substituted component resolves the issue, replace or quarantine the failed item, complete the full checkout, and stop.

If the Problem Persists

If power, medical-gas supplies, controls, accessories, settings, outlet configuration, and external connections have been verified, the fault is likely within the electronic gas mixer, control electronics, valves, sensors, wiring, or related gas-delivery assemblies.

The device should be:

Do not continue adjusting or replacing internal gas-control components without appropriate training, service documentation, and gas-delivery verification equipment.

Knowing when to stop is proper troubleshooting.

Clinical Use Tip

Never test uncertain electronic gas delivery on an active patient. Transfer the patient first and use independent oxygen, ventilation, and gas monitoring until another verified anesthesia machine is available.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Anesthesia staff reported that the Avance electronic fresh-gas flow setting would not respond and Alternate O2 activated during setup."

Cause

What was observed during troubleshooting.

Example:
"Pipeline supplies and external connections tested normally, but electronic flow adjustment remained unavailable during the approved machine checkout."

Resolution

What action was taken.

Example:
"Removed the Avance from service, labeled it Out of Service, provided a replacement machine, and forwarded the unit for electronic gas-delivery system evaluation."

Helpful Details to Include (If Known)

Final Thought

Electronic flow-control complaints require immediate attention because fresh-gas and oxygen delivery may be unreliable. Protect the patient, verify external supplies and controls logically, and escalate unresolved gas-delivery faults for qualified repair. Detailed CCR documentation supports safe follow-up and prevents repeated troubleshooting.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide