GE Healthcare Avance

PEEP High or Elevated Expiratory Pressure

On this page

Asset Type

Anesthesia Machine

Manufacturer

GE Healthcare

Model

Avance

What This Guide Helps With

Troubleshooting elevated PEEP or incomplete exhalation caused by circuit restrictions, valve problems, scavenging issues, moisture, accessories, or incorrect ventilator settings.

Step-by-Step Troubleshooting

1. Ensure Patient Safety First

Do not perform extended troubleshooting while the Avance is supporting an active patient.

If expiratory pressure remains elevated during a procedure:

Expected: Patient ventilation no longer depends on equipment with unresolved elevated expiratory pressure.

If the patient is safely supported by another device, continue troubleshooting in a controlled Clinical Engineering setting.

2. Confirm the Reported Condition

Review the alarm history and determine whether the issue involves:

Record the selected PEEP, measured PEEP, airway-pressure waveform, ventilation mode, respiratory rate, and tidal volume.

Expected: The complaint is clearly defined and can be reproduced using a test lung.

3. Verify Ventilator Settings

Check that:

Return settings to a known test configuration when appropriate.

Expected: Expiratory pressure returns to the selected PEEP level.

If correcting the settings resolves the issue, complete operational testing and stop.

4. Inspect the Patient Circuit

Disconnect the circuit from clinical use and inspect the full breathing path for:

Replace questionable components with known-good compatible accessories.

Expected: Gas can flow freely through the inspiratory and expiratory limbs.

If replacing the circuit resolves the issue, complete a system checkout and stop.

5. Check the Expiratory Limb and Accessories

Pay particular attention to the expiratory side of the circuit.

Remove or replace, one item at a time:

A saturated filter or restricted accessory may allow inspiration while slowing exhalation.

Expected: Pressure falls promptly during expiration and stabilizes at the selected PEEP.

6. Inspect the Breathing-System Valves

With the machine removed from patient use, visually inspect the inspiratory and expiratory check-valve areas for:

Do not disassemble internal valve mechanisms beyond normal operator-removable components.

Expected: Visible valve disks are correctly seated and move freely during controlled testing.

If contamination or incorrect assembly is corrected, repeat the system checkout.

7. Inspect the Absorber and Breathing-System Assembly

Confirm that:

Expected: The breathing system is assembled correctly with no visible obstruction.

8. Check the Adjustable Pressure-Limiting Valve

If the issue occurs during manual or bag ventilation:

Do not attempt internal repair of the valve mechanism during field troubleshooting.

Expected: Pressure releases appropriately during manual ventilation.

9. Evaluate the Scavenging System

Inspect the anesthetic-gas scavenging path for:

Temporarily isolate the machine from the facility scavenging connection only in an approved test environment and according to facility procedures.

Expected: Scavenging does not create backpressure or interfere with breathing-system exhaust.

10. Check the Bellows Movement

Connect a verified test lung and observe the bellows during ventilation.

Look for:

The Avance uses an ascending bellows assembly, and abnormal filling or movement may indicate a breathing-system, exhaust, or ventilator problem.

Expected: The bellows moves smoothly and refills consistently between breaths.

Do not open the bellows housing or perform internal pneumatic adjustments outside authorized service procedures.

11. Compare Ventilator and Manual Modes

Test the machine with a test lung in both:

Interpret the result:

Expected: The failure is narrowed to an external component, operating mode, or internal system.

12. Run the Approved System Checkout

After correcting external issues:

Expected: The machine completes testing and maintains the selected PEEP without unintended pressure buildup.

If testing passes and the complaint cannot be reproduced, document all findings and return the device to service according to facility policy.

If the Problem Persists

If elevated expiratory pressure remains after settings, circuits, filters, check valves, absorber components, scavenging, and visible breathing-system components have been checked, common external causes have been ruled out.

The problem may involve an internal:

The device should be:

Do not continue clinical use based only on intermittent normal operation. Knowing when to stop and escalate is proper troubleshooting.

Clinical Use Tip

Never troubleshoot elevated PEEP on an active patient. Incomplete exhalation can cause unintended pressure buildup. Move the patient to verified equipment before manipulating circuits, valves, or scavenging connections.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Anesthesia staff reported that the GE Healthcare Avance displayed elevated PEEP and airway pressure did not return to the selected baseline during ventilation."

Cause

What was observed during troubleshooting.

Example:
"Testing found a moisture-saturated filter in the expiratory limb that restricted exhaled gas flow and caused pressure to remain elevated between breaths."

Resolution

What action was taken.

Example:
"Replaced the restricted filter, inspected the breathing circuit, completed the system checkout, and verified stable ventilation with measured PEEP matching the selected setting."

Helpful Details to Include (If Known)

Final Thought

Elevated expiratory pressure requires immediate attention because it may affect ventilation and patient safety. Start with settings, circuits, accessories, valves, and scavenging before suspecting internal failure. Escalate appropriately and document the complaint, confirmed cause, corrective action, and final operating status.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide