GE Healthcare Avance

System Checkout or Leak Test Failure

On this page

Asset Type

Anesthesia Machine

Manufacturer

GE Healthcare

Model

Avance

What This Guide Helps With

Troubleshooting failed checkout or leak testing caused by circuit connections, breathing-system assembly, vaporizers, absorbers, sampling components, or test setup.

Step-by-Step Troubleshooting

1. Ensure Patient Safety First

Do not perform a system checkout, leak test, or extended troubleshooting while the Avance is connected to an active patient.

If a leak or checkout failure occurs during clinical use:

Expected: Patient care is no longer dependent on equipment with an unresolved leak or checkout failure.

If the device is safely removed from patient use, continue troubleshooting.

2. Confirm the Exact Test Failure

Review the displayed checkout results and record:

The Avance low-pressure leak check evaluates the gas path before the breathing system, while other checkout portions evaluate the breathing circuit and ventilation system.

Expected: The failed portion of the checkout is clearly identified before components are disturbed.

3. Verify the Test Setup

Confirm that:

An incorrect setup can create a failure even when the machine is functioning normally.

Expected: The checkout is repeated using the correct configuration.

If the test passes, document the setup error and return the device to service after completing the full checkout.

4. Inspect the Breathing Circuit

Check the inspiratory and expiratory limbs for:

Replace questionable disposable components with known-good compatible items.

Expected: The breathing circuit is complete, securely connected, and free of visible damage.

Repeat the failed test. If it passes, stop troubleshooting.

5. Check the Patient Y-Piece and Test Occlusion

Inspect the Y-piece and the device used to occlude the patient connection.

Confirm that:

GE identifies properly sized, undamaged plugs as necessary for isolating breathing-system sections during leak testing.

Expected: The patient connection remains completely sealed during the test.

6. Inspect the Bag, Bag Hose, and Bag/Vent Connections

Check:

Substitute a known-good reservoir bag and hose when available.

Expected: The manual breathing path is securely assembled and does not introduce leakage.

Repeat the checkout. If the test passes, replace the failed accessory and stop.

7. Verify the Breathing-System Module Is Fully Seated

Inspect the externally removable breathing-system assembly.

Confirm that:

Do not force the assembly or remove internal covers.

Expected: The breathing system sits flush and locks securely into position.

8. Check the Absorber Canister Assembly

Inspect the carbon-dioxide absorber assembly for:

Reseat the absorber according to normal operator-accessible procedures.

Expected: The absorber is securely installed with clean, intact sealing surfaces.

Repeat the leak test. If the test passes, document the seating or absorber issue.

9. Inspect Sampling and Gas-Monitoring Connections

Check the gas-sampling path for:

Temporarily replace suspect sampling accessories with known-good compatible parts.

Expected: All gas-monitoring connections are secure and properly sealed.

10. Inspect Vaporizers and Mounting Positions

Ensure all installed vaporizers are:

If permitted by facility procedure, remove and reseat one vaporizer at a time while the machine is out of service. Repeat the appropriate test after each change.

Do not open or internally service a vaporizer.

The machine checkout may identify leakage within the vaporizer or low-pressure gas path before the breathing system.

Expected: Vaporizers are securely mounted and do not cause the test failure.

11. Check External Gas and Auxiliary Connections

Inspect accessible connections for:

Use a known-good pipeline source when a supply problem is suspected.

Expected: External gas connections are secure and are not creating an unintended leak path.

12. Power-Cycle and Repeat the Full Checkout

After correcting any external issue:

Do not bypass or accept a failed mandatory checkout solely because ventilation appears functional.

Expected: All required checkout tests pass without recurring messages or abnormal leakage.

If the checkout passes repeatedly, complete a functional verification and return the device to service according to facility policy.

If the Problem Persists

If the Avance continues to fail checkout after the circuit, absorber, sampling components, vaporizers, test setup, and external connections have been verified, common external causes have been ruled out.

The problem may involve an internal valve, manifold, pressure sensor, pneumatic connection, breathing-system seal, or ventilation component.

The device should be:

Do not repeatedly clear or bypass a failed checkout for clinical use. Knowing when to stop and escalate is proper troubleshooting.

Clinical Use Tip

Never perform leak testing or system checkout on an active patient. Move the patient to verified equipment first, then troubleshoot the Avance in a controlled environment.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Clinical staff reported that the GE Healthcare Avance failed the pre-use system checkout with a breathing-system leak message."

Cause

What was observed during troubleshooting.

Example:
"Inspection found the absorber canister was not fully seated, creating leakage during the checkout."

Resolution

What action was taken.

Example:
"Reseated and locked the absorber canister, inspected the circuit connections, and completed the full system checkout successfully."

Helpful Details to Include (If Known)

Final Thought

A failed anesthesia-machine checkout must be handled with patient safety first and a logical examination of external leak paths. Careful documentation and timely escalation help prevent unsafe return to service.

That is successful troubleshooting.