GE Healthcare B105 / B125 / B155 Series

Alarm Speaker, Alarm Light, or Audible Alarm Failure

On this page

Asset Type

Patient Monitor

Manufacturer

GE Healthcare

Model

B105 / B125 / B155 Series

What This Guide Helps With

Troubleshoots missing audible alarms, alarm-light problems, volume or configuration concerns, blocked sound paths, and alarm-output failures.

Step-by-Step Troubleshooting

1. Protect the Patient and Replace Unreliable Alarm Coverage

If the monitor cannot produce required audible or visual alarms, do not leave it in service for patient monitoring.

Move the patient to another verified monitor or establish an approved alternate alarm-surveillance method before troubleshooting.

Expected outcome: The patient is protected by a reliable alarm system independent of the affected monitor.

2. Confirm the Exact Alarm Complaint

Determine whether:

Do not create unsafe clinical alarm conditions merely to reproduce the complaint.

Expected outcome: The failed annunciation function is clearly identified.

3. Inspect the Monitor Exterior

Inspect the speaker openings, alarm-light area, housing, and controls for:

Remove only external obstructions that can be safely removed.

Expected outcome: Audible and visual alarm annunciators are not externally blocked or damaged.

If removing an external obstruction restores full alarm function, continue with formal verification.

4. Verify Alarm Volume and User-Accessible Settings

Review the current authorized alarm volume and related user-accessible settings.

Confirm that the reported failure is not caused by an unexpectedly low setting or another approved configuration state.

Do not disable required alarms, bypass minimum settings, or enter restricted service menus.

Expected outcome: Alarm configuration is appropriate for clinical use.

If correcting an authorized setting restores audible alarm output, continue to final verification.

5. Confirm the Monitor Is Not in a Temporary Alarm State

Determine whether alarm silence, pause, standby, or another operator-accessible state is active.

Coordinate with clinical staff when necessary to distinguish expected alarm behavior from actual failure.

Expected outcome: The monitor is in a normal alarm-enabled state for testing.

6. Perform a Controlled Alarm Test Off-Patient

Use an approved test method to create a safe alarm condition with the monitor off-patient or connected to appropriate simulation equipment.

Verify:

Expected outcome: Audible, visual, and on-screen alarm functions occur together as intended for the approved test.

If all alarm outputs work reliably and the original complaint is explained by an external setting or temporary state, troubleshooting can stop after documentation.

7. Evaluate Sound Quality and Intermittency

Listen for:

Do not conclude that any audible tone proves the alarm speaker is reliable.

Expected outcome: Alarm audio remains clear and consistent through repeated approved tests.

8. Verify Alarm Light Function

Observe the visual alarm indicator during an approved alarm test.

Check for:

Do not disassemble the alarm-light assembly during external troubleshooting.

Expected outcome: The visual alarm indicator functions reliably during the test condition.

9. Verify Networked Alarm Communication When Applicable

If the monitor normally sends alarm information to a central station, verify the complete monitoring path after local alarm testing.

A functioning central station does not compensate for a failed required local alarm output unless specifically addressed by facility policy and approved clinical workflow.

Expected outcome: Required local and remote alarm annunciation pathways are verified.

10. Perform Final Functional Verification

Before return to service, verify:

Expected outcome: All required alarm outputs operate reliably.

If every required alarm function passes, document and return the monitor to service.

11. Escalate Any Persistent Alarm-Output Failure

If the speaker, alarm light, or other required annunciator remains unreliable after settings and external conditions are ruled out, remove the monitor from service.

Do not return a monitor to patient use based on the assumption that staff will notice visual alarms or central-station alarms instead.

Expected outcome: A monitor with unresolved alarm annunciation failure is prevented from clinical use and routed for qualified repair.

If the Problem Persists

External obstruction, user-accessible settings, temporary alarm states, and controlled alarm testing have been evaluated. Remaining causes may involve the speaker, alarm-light assembly, internal audio circuitry, control electronics, software, configuration, or another service-level condition.

The monitor should be:

After repair, perform complete alarm-system verification, including audible, visual, on-screen, acknowledgment, and applicable central-station alarm behavior before return to clinical service.

Knowing that a monitor with unreliable alarm annunciation is unsafe even when parameter measurements remain accurate is proper troubleshooting.

Clinical Use Tip

A patient monitor with a failed local alarm speaker should be removed from clinical use rather than relying on staff to watch the screen continuously.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Nursing reported that the B155 monitor displayed alarms on screen but no audible alarm could be heard."

Cause

What was observed during troubleshooting.

Example:
"Clinical Engineering found the local alarm speaker remained silent during repeated approved alarm tests despite appropriate alarm settings."

Resolution

What action was taken.

Example:
"Removed the monitor from service and sent it for qualified repair after confirming the audible alarm failure and documenting that the visual alarm remained functional."

Helpful Details to Include (If Known)

Final Thought

Alarm reliability is a patient-safety function, not a convenience. Verify settings and external conditions first, test the complete alarm path under controlled conditions, immediately remove unreliable alarm equipment from service, and document exactly what was proven.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide