GE Healthcare B105 / B125 / B155 Series

Invasive Pressure Channel Missing, Drifting, or Will Not Zero

On this page

Asset Type

Patient Monitor

Manufacturer

GE Healthcare

Model

B105 / B125 / B155 Series

What This Guide Helps With

Troubleshoots missing invasive pressure channels, zeroing problems, unstable baselines, transducer cables, connections, setup, and external pressure-monitoring components.

Step-by-Step Troubleshooting

1. Protect the Patient and Maintain Pressure Monitoring

If invasive pressure monitoring is clinically required and the channel is absent, drifting, or cannot be zeroed reliably, ensure another verified monitoring method is available before technical troubleshooting.

Do not disconnect or manipulate an active invasive pressure system without coordination with appropriate clinical staff.

Expected outcome: Patient monitoring remains continuous while the affected pressure channel is evaluated safely.

2. Confirm the Exact Pressure Complaint

Determine whether:

Identify which invasive pressure channel is affected.

Expected outcome: The failure pattern and affected signal path are clearly identified.

3. Inspect the External Transducer Cable

Inspect the reusable pressure cable and connectors for:

Reseat accessible monitor and transducer cable connections.

Expected outcome: The cable is intact and fully seated.

If reseating restores a stable channel, continue to functional verification.

4. Verify the Disposable Transducer Setup With Clinical Staff

Coordinate with clinical personnel to verify the patient-side pressure setup is assembled and positioned appropriately.

Check externally for:

Clinical personnel should manage the patient-connected fluid path.

Expected outcome: The pressure system is appropriately configured for the clinical zeroing process.

If correcting the clinical setup allows a stable zero and waveform, troubleshooting can stop after verification.

5. Verify Channel Configuration

Confirm that the pressure parameter is enabled and assigned appropriately for the connected input.

Avoid changing clinical labels or configuration unnecessarily. Record existing configuration before authorized adjustments.

Expected outcome: The connected invasive pressure input is configured and available for display.

If the channel was disabled or incorrectly assigned and authorized correction restores it, verify function and stop troubleshooting.

6. Attempt a Proper Zero With Clinical Coordination

With the transducer appropriately positioned and open to atmosphere according to clinical procedure, initiate the normal zero operation.

Observe whether zeroing:

Do not attempt to compensate for drift by repeatedly zeroing a faulty system.

Expected outcome: The channel accepts a stable zero under appropriate conditions.

7. Substitute a Known-Good Pressure Cable

Use a compatible known-good reusable invasive pressure cable when available.

If the issue resolves, retest the original cable to confirm the failure follows it before replacement.

Expected outcome: A known-good cable provides stable channel recognition and zeroing.

If the problem follows the cable, replace the defective cable and proceed to final verification.

8. Use an Approved Pressure Simulator

Off-patient, connect an approved invasive pressure simulator or test setup using known-good accessories.

Verify:

Do not perform unauthorized calibration adjustments.

Expected outcome: The monitor accurately and consistently responds to the controlled pressure input within the applicable approved test procedure.

9. Compare Another Pressure Input When Available

If the monitor configuration provides another appropriate invasive pressure input, use a controlled off-patient comparison according to facility procedures.

This can help determine whether the problem follows a cable/transducer or remains associated with one monitor input.

Expected outcome: The fault can be localized to an external component or a particular monitor channel without internal disassembly.

10. Perform Final Functional Verification

After correction, verify:

Expected outcome: The invasive pressure channel performs consistently and is ready for clinical use.

If all checks pass, document and return the monitor to service.

11. Escalate Persistent Pressure-Channel Failure

If the channel remains missing, will not zero, or drifts when connected to known-good accessories and approved pressure test equipment, stop external troubleshooting.

Expected outcome: The monitor is removed from service for qualified evaluation.

If the Problem Persists

External transducers, reusable cables, configuration, zeroing conditions, and controlled simulator inputs have been evaluated. Remaining possibilities may involve the invasive pressure input circuitry, connector interface, parameter hardware, calibration, configuration, or another service-level fault.

The monitor should be:

After repair, verify all affected invasive pressure channels, zero function, pressure response, alarms, and required overall monitor operation before return to service.

Knowing when repeated zeroing is masking rather than solving a pressure-monitoring problem is proper troubleshooting.

Clinical Use Tip

Do not manipulate the patient-connected fluid path solely for equipment troubleshooting; coordinate zeroing and transducer setup changes with the responsible clinical team.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Nursing reported that the B155 arterial pressure channel intermittently disappeared and would not consistently zero."

Cause

What was observed during troubleshooting.

Example:
"Clinical Engineering found an intermittent reusable invasive pressure cable and confirmed stable operation with a known-good cable."

Resolution

What action was taken.

Example:
"Replaced the defective pressure cable, verified channel detection, stable zero and simulated pressure response, tested applicable alarms, and returned the monitor to service."

Helpful Details to Include (If Known)

Final Thought

Preserve patient safety and the sterile clinical pressure setup, verify transducers, cables, positioning, zeroing conditions, and configuration before suspecting internal hardware, and escalate any channel that cannot remain stable under controlled testing.

That is successful troubleshooting.