GE Healthcare B105 / B125 / B155 Series

Network or Central Station Communication Failure

On this page

Asset Type

Patient Monitor

Manufacturer

GE Healthcare

Model

B105 / B125 / B155 Series

What This Guide Helps With

Troubleshoots loss of central monitoring, network disconnection, cabling, infrastructure, configuration, location-specific faults, and communication-path problems.

Step-by-Step Troubleshooting

1. Protect the Patient and Maintain Alarm Surveillance

If the bedside monitor is not communicating with the central station or required networked alarm destination, ensure staff understand that central surveillance may be unavailable.

Provide an alternate verified monitoring and alarm-observation method as required by clinical workflow.

Do not assume bedside monitoring alone provides equivalent central alarm coverage.

Expected outcome: Patient monitoring and alarm surveillance remain clinically appropriate while communication is unavailable.

2. Confirm the Exact Communication Failure

Determine whether:

Expected outcome: The scope and direction of the communication failure are understood.

3. Verify Bedside Monitor Operation

Confirm the monitor itself is powered and locally displaying the expected patient parameters and alarms.

A local monitoring failure should be addressed before network troubleshooting.

Expected outcome: The bedside monitor is functioning normally apart from network or central-station communication.

4. Inspect External Network Connections

For wired connections, inspect accessible network cables, wall connections, couplers, and monitor-side connections for:

Reseat external cables.

For wireless configurations, confirm that the expected wireless communication state is present without making unauthorized network changes.

Expected outcome: The physical communication path is securely connected and free of obvious damage.

If reseating or replacing a damaged external cable restores communication, proceed to end-to-end verification.

5. Use a Known-Good Cable or Port Comparison

When authorized, substitute a known-good compatible network cable.

If infrastructure policy permits, compare operation at a verified working network connection in the same approved clinical environment.

Do not move network devices between ports arbitrarily or alter switch configurations.

Expected outcome: The fault is separated between the monitor/cable and the network connection.

If the monitor communicates normally through a known-good external path, escalate the original port or infrastructure issue to the appropriate network team.

6. Determine Whether the Problem Is Device-Specific or Area-Wide

Check whether nearby networked monitors are communicating normally.

A failure affecting multiple monitors suggests infrastructure, central-station, network-service, or system-level causes rather than simultaneous bedside monitor failures.

Expected outcome: The scope of the outage is identified.

If multiple devices are affected, follow the facility escalation process for monitoring-network or central-station outages.

7. Verify Approved Network Configuration

Review only configuration fields that Clinical Engineering is authorized to inspect.

Confirm that expected network identity, location, bed assignment, central-station association, or other required configuration has not obviously been changed.

Do not guess configuration values or copy settings from another monitor without verifying they are correct for that device and location.

Expected outcome: The monitor has the approved configuration required for its network role.

If an authorized configuration correction restores communication, proceed to end-to-end verification.

8. Check Central Station Recognition

At the receiving system, determine whether the bedside device:

Coordinate with the appropriate clinical systems, IT, or network support team when central-station or infrastructure access is outside Clinical Engineering scope.

Expected outcome: The communication failure is localized to the bedside device, network path, or receiving system.

9. Perform an End-to-End Communication Test

After any correction, verify the complete path from bedside monitor to required receiving station.

Confirm:

Expected outcome: The intended bedside-to-central communication path is fully restored.

If all required central monitoring functions are verified, troubleshooting can stop.

10. Escalate an Unresolved Communication Failure

If cabling, approved configuration, and local device operation are normal but communication remains unavailable, stop external troubleshooting and escalate appropriately.

Do not perform unauthorized switch, server, VLAN, network-security, or protected central-station configuration changes.

Expected outcome: The unresolved issue is assigned to the appropriate Clinical Engineering, clinical systems, IT, network, or manufacturer support resource.

If the Problem Persists

External connections, device operation, basic approved configuration, and local comparisons have been checked. Remaining causes may involve monitor network hardware, central-station configuration, network infrastructure, switch or port configuration, server services, security controls, or other system-level communication functions.

The affected monitor should be:

Infrastructure issues should be escalated to the responsible support team while clinical staff maintain an alternate surveillance plan.

Return the device to network-dependent service only after the complete communication and alarm path is verified.

Knowing when a failure belongs to infrastructure rather than the bedside monitor is proper troubleshooting.

Clinical Use Tip

After restoring connectivity, verify the correct patient or bed appears at the intended central station; a network link alone does not prove the monitoring path is correct.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Nursing reported that the B125 monitor was operating normally at bedside but was no longer visible at the central station."

Cause

What was observed during troubleshooting.

Example:
"Clinical Engineering found the external network cable had a damaged locking tab and the connection was intermittently opening."

Resolution

What action was taken.

Example:
"Replaced the network cable, verified stable bedside connectivity, confirmed correct waveforms and alarm communication at the central station, and returned the monitor to service."

Helpful Details to Include (If Known)

Final Thought

Maintain a safe alarm-surveillance plan, verify the physical and logical communication path before assuming a monitor failure, separate device-specific problems from infrastructure outages, and confirm end-to-end central monitoring before return to service.

That is successful troubleshooting.