GE Healthcare B105 / B125 / B155 Series

NIBP Cuff Will Not Inflate or Measurement Fails

On this page

Asset Type

Patient Monitor

Manufacturer

GE Healthcare

Model

B105 / B125 / B155 Series

What This Guide Helps With

Troubleshoots failed NIBP cycles, no cuff inflation, leaks, damaged hoses, cuff problems, connection issues, and externally verifiable measurement failures.

Step-by-Step Troubleshooting

1. Protect the Patient and Provide an Alternate Blood Pressure Method

If blood pressure measurement is clinically required and NIBP is unavailable or unreliable, ensure an alternate verified blood pressure method is available before troubleshooting.

Do not repeatedly cycle a malfunctioning cuff on a patient while diagnosing equipment.

Expected outcome: Clinical care continues without dependence on the affected NIBP function.

2. Confirm the Exact NIBP Complaint

Determine whether:

Expected outcome: The NIBP failure is clearly characterized.

3. Inspect the Cuff and Hose

Inspect the NIBP cuff, tubing, connectors, and hose for:

Expected outcome: The external pneumatic path is unobstructed, intact, and securely connected.

If correcting a kink or loose fitting restores normal measurements, proceed to final verification.

4. Verify Cuff Selection and Application

Coordinate with clinical staff to confirm the cuff is appropriate for the intended patient and is applied correctly.

A poorly positioned or inappropriate cuff can contribute to failed or unreliable measurements even when the monitor functions normally.

Expected outcome: A suitable cuff is correctly applied without obstruction of the tubing.

If proper cuff application corrects the issue, verify repeated successful measurements and troubleshooting can stop.

5. Reseat the NIBP Hose Connection

Disconnect and reconnect the external NIBP hose at accessible connection points.

Inspect the monitor-side connector for obvious physical damage or looseness.

Do not force incompatible fittings.

Expected outcome: The hose is fully seated and forms a secure pneumatic connection.

If reseating restores consistent inflation and measurement, proceed to final verification.

6. Substitute Known-Good NIBP Accessories

Use a compatible known-good cuff and hose.

Substitute one component at a time when practical to determine whether the failure follows the cuff or tubing.

Expected outcome: The known-good cuff and hose inflate and complete measurements normally.

If the original cuff or hose causes the failure, replace the defective accessory.

7. Perform an Off-Patient NIBP Test

Use approved NIBP test equipment or a suitable simulator according to facility procedures.

Observe whether:

Do not infer calibration accuracy from a simple inflation test.

Expected outcome: The monitor completes a controlled NIBP test consistently.

8. Check for an External Leak

Using approved test methods, determine whether an externally connected hose or cuff assembly leaks.

Do not disassemble the monitor's internal pneumatic system as part of routine external troubleshooting.

Expected outcome: The external pneumatic circuit maintains pressure as expected for the approved test.

If an external leak is identified, replace the affected accessory and retest.

9. Perform Final Functional Verification

After correction, verify:

Perform any required performance verification with approved NIBP test equipment.

Expected outcome: NIBP operation is repeatable and appropriate for clinical use.

If all checks pass, document and return the monitor to service.

10. Escalate Persistent NIBP Failure

If the monitor cannot inflate or complete measurements using known-good accessories and approved test equipment, stop external troubleshooting.

Do not proceed into internal pump, valve, transducer, pneumatic assembly, or board-level repair unless specifically authorized and trained.

Expected outcome: The monitor is removed from service and routed for qualified repair.

If the Problem Persists

External cuff, hose, fittings, application, and controlled test conditions have been ruled out. Remaining causes may involve internal pneumatic components, pressure measurement circuitry, calibration, configuration, or another service-level fault.

The monitor should be:

After repair, complete the required NIBP performance verification and confirm overall monitor and alarm operation before return to clinical use.

Knowing when a pneumatic failure requires service-level evaluation is proper troubleshooting.

Clinical Use Tip

Avoid repeated unsuccessful cuff cycles on a patient; move to an alternate verified blood pressure method while Clinical Engineering evaluates the monitor.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Nursing reported that the B155 monitor began inflating the NIBP cuff but repeatedly failed to complete measurements."

Cause

What was observed during troubleshooting.

Example:
"Clinical Engineering found a leak in the external NIBP hose and confirmed normal operation with a known-good hose."

Resolution

What action was taken.

Example:
"Replaced the damaged hose, verified repeated successful NIBP cycles with approved test equipment, checked deflation and alarms, and returned the monitor to service."

Helpful Details to Include (If Known)

Final Thought

Maintain an alternate blood pressure method, inspect the complete external pneumatic path before assuming an internal failure, and distinguish simple inflation from verified NIBP performance. Escalate unresolved pneumatic problems and document the confirmed cause and testing.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide