GE Healthcare B105 / B125 / B155 Series

SpO2 Sensor Not Detected or No Saturation Reading

On this page

Asset Type

Patient Monitor

Manufacturer

GE Healthcare

Model

B105 / B125 / B155 Series

What This Guide Helps With

Troubleshoots missing SpO2 readings, sensor detection problems, damaged cables, poor application, low signal conditions, and external connection faults.

Step-by-Step Troubleshooting

1. Protect the Patient and Maintain Oxygenation Monitoring

If SpO2 monitoring is clinically required and the displayed saturation is absent or unreliable, provide another verified method of monitoring before troubleshooting.

Do not leave a patient dependent on an unreliable SpO2 channel.

Expected outcome: The patient's oxygenation is being monitored reliably while the affected device is evaluated.

2. Confirm the Exact SpO2 Complaint

Determine whether:

Record any displayed status or alarm message without assuming its cause.

Expected outcome: The specific SpO2 failure pattern is identified.

3. Inspect and Reseat the SpO2 Connections

Inspect the sensor, extension/interconnect cable if present, and monitor connection for:

Disconnect and reconnect each accessible external connection securely.

Expected outcome: The SpO2 signal path is physically intact and fully connected.

If reseating a loose connection restores normal readings, proceed to final verification.

4. Verify Sensor Compatibility

Confirm that the attached sensor and any interface cable are approved and compatible with the monitor's installed SpO2 technology.

Do not assume that connectors which physically fit are electrically or technologically compatible.

Expected outcome: The connected SpO2 accessories are appropriate for the monitor.

If an incompatible accessory was installed, replace it with an approved compatible accessory and verify operation.

5. Check Sensor Application and Measurement Site

Coordinate with clinical staff to verify appropriate sensor placement and adequate contact with the patient.

Consider external factors such as:

Clinical staff should select or reposition the patient measurement site as appropriate.

Expected outcome: The sensor has appropriate patient contact and a usable physiologic signal.

If proper application restores a stable waveform and saturation value, troubleshooting can stop after verification.

6. Substitute a Known-Good Sensor

Use a compatible known-good SpO2 sensor to determine whether the problem follows the original sensor.

If an intermediate cable is used, substitute that component separately when available.

Change one component at a time.

Expected outcome: A known-good sensor and cable are detected and produce normal monitoring.

If the failure follows the original sensor or cable, replace the defective accessory.

7. Compare With an Approved SpO2 Simulator When Available

For bench evaluation, connect approved test equipment appropriate for the installed SpO2 technology.

Verify the monitor can detect a valid signal and display expected monitoring information.

Expected outcome: The monitor responds normally to a controlled known-good SpO2 input.

If simulator testing passes, focus on patient-side sensors, cables, application, and physiologic/environmental conditions.

8. Inspect the Monitor SpO2 Input

Inspect the accessible input connection for debris, contamination, looseness, or physical damage.

Do not probe or repair the internal connector without approved service procedures.

Expected outcome: The monitor input appears clean, intact, and mechanically secure.

9. Perform Final Functional Verification

After correction, verify:

Expected outcome: SpO2 monitoring and alarms function reliably.

If all checks pass, document and return the monitor to service.

10. Escalate an Unresolved SpO2 Failure

If multiple compatible known-good sensors and cables fail, or approved simulator testing indicates the monitor cannot acquire SpO2, stop external troubleshooting.

Expected outcome: The monitor is removed from service for qualified evaluation.

If the Problem Persists

Accessory compatibility, connections, patient application, known-good sensors, and accessible monitor connections have been checked. Remaining causes may involve the SpO2 acquisition subsystem, input interface, installed parameter hardware, configuration, or another service-level condition.

The monitor should be:

After repair, verify SpO2 acquisition, sensor detection, alarm operation, and required overall monitor functions before return to service.

Knowing when the problem no longer follows an external SpO2 accessory is proper troubleshooting.

Clinical Use Tip

When comparing SpO2 accessories, verify technology compatibility first; a physically connectable sensor is not automatically compatible with the monitor.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Nursing reported that the B125 monitor intermittently displayed no SpO2 reading and lost sensor detection."

Cause

What was observed during troubleshooting.

Example:
"Clinical Engineering found the SpO2 sensor cable was damaged near the strain relief and failed when flexed."

Resolution

What action was taken.

Example:
"Replaced the defective compatible sensor, verified a stable SpO2 signal and saturation display, tested alarms, and returned the monitor to service."

Helpful Details to Include (If Known)

Final Thought

Maintain reliable oxygenation monitoring, verify compatibility and the complete external sensor path before suspecting the monitor, and use controlled substitutions to separate accessory failures from equipment faults. Escalate unresolved acquisition problems and document the verified findings.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide