On this page
Asset Type
Manufacturer
Model
What This Guide Helps With
Troubleshoots touchscreen, visual alarm, or audible alarm failures caused by contamination, obstruction, settings, accessories, software state, or device-level faults.
Step-by-Step Troubleshooting
1. Protect the Patient and Remove Unreliable Alarm Equipment From Dependence
If the touchscreen prevents safe control of monitoring or if the alarm light or audible alarm is unreliable, move the patient to another verified monitor before troubleshooting.
A patient monitor with an untrustworthy alarm output must not remain the sole source of alarm notification.
Expected outcome: The patient is monitored by equipment with verified controls and alarm annunciation.
2. Identify the Exact Failed Function
Determine whether the complaint involves:
- Touchscreen completely unresponsive.
- Specific touchscreen area not responding.
- Delayed or erratic touch response.
- Alarm light not illuminating.
- Audible alarm absent, weak, distorted, or intermittent.
- More than one function failing simultaneously.
Confirm whether the failure is continuous or intermittent.
Expected outcome: The affected user-interface or alarm function is clearly identified.
3. Inspect the Exterior for Physical or Liquid Damage
Check the display, bezel, alarm indicator area, speaker openings, and enclosure for:
- Cracks.
- Impact marks.
- Liquid intrusion.
- Cleaning residue.
- Heavy contamination.
- Obstruction.
- Loose external accessories pressing against the display.
Remove the device from service immediately if liquid intrusion or significant damage is suspected.
Expected outcome: No external condition is compromising safe operation. If cleaning or removing an obstruction restores normal function, continue to final verification.
4. Clean the Touchscreen Using Approved Methods
If touch response is poor, clean and dry the screen according to approved facility and manufacturer-compatible practices.
Remove gloves, protective coverings, adhesive film, or external objects that may interfere with normal touch input when appropriate.
Expected outcome: Touch response is normal across the usable display. If so, verify the rest of the monitor and stop troubleshooting.
5. Check Normal User-Accessible Alarm Settings
Verify alarm volume and other accessible alarm settings have not been intentionally reduced or changed in a way that explains the complaint.
Confirm the monitor is not in an expected operational state that suppresses or modifies audible annunciation.
Do not bypass alarm protections or enter unauthorized service menus.
Expected outcome: Alarm settings are appropriate for clinical use. If an accessible setting caused the complaint, correct it and perform a full alarm test.
6. Perform a Normal Restart When Clinically Safe
With the patient transferred to alternate monitoring, perform a normal restart of the CARESCAPE ONE.
Observe touchscreen operation, startup indicators, and any abnormal system messages.
Expected outcome: The device initializes normally and the affected function returns. If a restart resolves the condition, complete repeated functional and alarm verification before return to service.
7. Verify Touchscreen Operation
Test normal accessible screen controls across multiple regions of the display without entering restricted service functions.
Look for dead zones, unintended selections, repeated taps, or inconsistent operation.
Expected outcome: Touch input accurately and consistently responds across normal clinical controls. Any persistent dead zone or erratic input requires removal from service.
8. Verify Visual and Audible Alarms
Using an approved simulator or controlled test condition, generate appropriate monitor alarms and verify:
- Alarm message appears.
- Alarm light functions.
- Audible annunciation is present and clear.
- Alarm priority behavior is appropriate according to approved documentation.
- Alarm silence or acknowledgement controls function normally.
Do not rely solely on a startup sound as proof that the alarm system is functional.
Expected outcome: Visual and audible alarm annunciation functions consistently. If all alarm outputs are verified, troubleshooting can stop.
9. Check Interaction With the Host Monitoring System
If the CARESCAPE ONE is docked to a B450, B650, or B850, confirm whether alarm behavior is local, host-dependent, or affected only while docked.
Verify alarms at every intended point of annunciation rather than assuming host communication compensates for a failed local alarm.
Expected outcome: Alarm behavior is understood and confirmed across the intended monitoring configuration.
10. Escalate Any Persistent Control or Alarm Failure
If the touchscreen remains unreliable, the alarm light does not function, or audible alarms cannot be verified after external checks and normal restart, stop troubleshooting.
Expected outcome: The monitor is removed from service and is not returned to patient care until the failed safety function is corrected and verified.
If the Problem Persists
Common external causes have been ruled out. The remaining problem may involve touchscreen hardware, display assembly, alarm indicator hardware, speaker or audio circuitry, internal connections, software, firmware, or another service-level fault.
The CARESCAPE ONE should be:
- Removed from service.
- Labeled Out of Service.
- Sent for repair or bench evaluation.
- Evaluated using appropriate GE Healthcare documentation and approved test equipment.
- Repaired or configured only by qualified personnel.
After repair, verify touchscreen control, display operation, visual alarms, audible alarms, parameter monitoring, docking, communication, and all applicable return-to-service safety checks before clinical use.
Knowing when to stop external troubleshooting is proper troubleshooting.
Clinical Use Tip
A monitor with a failed alarm light or audible alarm is not reliable patient-support equipment even if waveforms and numerics still appear normal.
Work Order Documentation (CCR Method)
CCR = Complaint, Cause, Resolution
Complaint
What was reported by the clinical staff.
Example:
"Nursing reported the CARESCAPE ONE displayed alarms visually but produced no audible alarm tone."
Cause
What was observed during troubleshooting.
Example:
"Clinical Engineering confirmed the audible alarm remained absent during controlled alarm testing despite appropriate user-accessible volume settings."
Resolution
What action was taken.
Example:
"The monitor was removed from service, labeled Out of Service, and sent for qualified bench repair after the alarm failure was reproduced and documented."
Helpful Details to Include (If Known)
- Exact failed function.
- Whether failure was intermittent.
- Touchscreen areas affected.
- Physical or liquid damage observed.
- Cleaning residue or screen obstruction.
- Alarm volume setting observed.
- Whether a normal restart changed the condition.
- Visual alarm behavior.
- Audible alarm behavior.
- Host-monitor alarm behavior when docked.
- Final device disposition or post-repair verification.
Final Thought
Touchscreen and alarm failures directly affect safe monitor operation, so patient dependence must end before testing begins. Rule out simple external and setting-related causes, verify alarms under controlled conditions, and remove the device from service whenever a critical control or annunciation function remains unreliable.
That is successful troubleshooting.