GE Healthcare Carestation 600 Series

Bellows Not Rising or Ventilator Not Cycling

On this page

Asset Type

Anesthesia Machine

Manufacturer

GE Healthcare

Model

Carestation 600 Series

What This Guide Helps With

Troubleshooting absent bellows movement or ventilation caused by circuit leaks, improper assembly, gas-supply problems, settings, connections, or external breathing-system faults.

Step-by-Step Troubleshooting

1. Ensure Patient Safety First

Do not perform extended troubleshooting while the Carestation 600 Series is supporting an active patient.

If the bellows stops moving or mechanical ventilation is not occurring during a procedure:

Expected: Patient ventilation no longer depends on equipment with an unresolved ventilator problem.

Continue troubleshooting only after the machine has been safely removed from patient use.

2. Confirm the Exact Failure Condition

Determine whether:

Record the selected ventilation mode, respiratory rate, tidal volume or pressure setting, fresh-gas flow, and observed bellows behavior.

Expected: The failure is clearly defined before components are changed.

3. Verify the Machine Is in Mechanical Ventilation Mode

Confirm that:

A selector left in bag mode can prevent expected ventilator and bellows operation.

Expected: The machine enters mechanical ventilation and the bellows begins cycling.

If ventilation starts normally, complete a full checkout before returning the machine to service and stop troubleshooting.

4. Verify Pipeline and Cylinder Gas Availability

Check the displayed supply pressures and confirm that the required medical-gas pipeline connections are secure.

Inspect for:

The ventilator requires an adequate drive-gas supply to operate correctly.

Expected: Gas pressures are within the facility and manufacturer-approved operating range, with no active supply alarms.

If gas supply is restored and ventilation operates normally, complete the machine checkout and stop.

5. Inspect the Patient Breathing Circuit

Remove the machine from patient use and inspect the complete breathing circuit.

Confirm that:

Because the Carestation 600 Series uses an ascending bellows, a major breathing-circuit disconnection or leak may prevent the bellows from filling normally.

Expected: The circuit is complete, unobstructed, and capable of holding pressure.

Replace damaged or questionable disposable components and repeat the ventilation test.

6. Check the Adjustable Pressure-Limiting Valve and Bag Assembly

Verify that the adjustable pressure-limiting valve moves normally and is not visibly damaged or obstructed.

Inspect the reservoir bag for:

Although the adjustable pressure-limiting valve primarily affects manual ventilation, abnormalities in the bag circuit can help identify a broader breathing-system leak or assembly problem.

Expected: Manual ventilation with a test lung produces pressure and the reservoir bag fills and empties normally.

If manual ventilation is also unsuccessful, focus on a breathing-system leak, obstruction, or assembly problem before suspecting an internal ventilator failure.

7. Inspect the Bellows Assembly Externally

With the machine off patient and depressurized, visually inspect the bellows housing.

Look for:

Do not remove or internally rebuild the bellows assembly unless authorized, trained, and following approved service documentation.

Expected: The bellows is correctly seated, physically intact, and able to move freely within its housing.

If visible damage is found, remove the machine from service and send it for repair.

8. Verify the Breathing-System Components Are Properly Installed

Inspect externally accessible breathing-system components that may have been removed for cleaning or service.

Confirm that:

A missing, loose, or incorrectly installed component can create a significant leak and prevent the bellows from rising.

Expected: All breathing-system components are present, correctly oriented, and securely installed.

If reseating a component resolves the problem, run the complete system checkout and stop.

9. Inspect the Flow Sensors and Associated Connections

Check the inspiratory and expiratory flow sensors for:

Allow wet components to dry according to approved facility and manufacturer procedures. Replace visibly damaged sensors with compatible known-good components when permitted.

Expected: Flow sensors are clean, dry, correctly installed, and recognized without related alarms.

If the ventilator begins cycling normally, perform the required calibration or checkout before returning the machine to service.

10. Check the Scavenging Connection for Severe Restriction

Inspect the anesthetic gas scavenging system connection and hose.

Confirm that:

A significant evacuation-system restriction or excessive suction can disrupt breathing-system pressure and bellows operation.

Expected: Scavenging is connected correctly and does not interfere with ventilation.

If disconnecting the machine from a questionable external scavenging source restores operation during controlled testing, investigate the facility evacuation system before returning the machine to use.

11. Power-Cycle the Machine Safely

After documenting active messages and removing the device from patient use:

Do not repeatedly restart a machine that displays a persistent ventilator, control, or hardware fault.

Expected: The system starts without ventilator-related faults and successfully completes checkout.

If normal ventilation returns after restart, perform repeated operational testing to confirm that the problem is not intermittent.

12. Perform the Full System Checkout

Use an approved test lung and perform the complete manufacturer-required checkout.

Verify:

The Carestation 600 Series user reference documentation identifies the platform as an integrated anesthesia delivery and patient-monitoring system; a successful system checkout is necessary before clinical use.

Expected: The machine completes checkout and provides stable, repeatable ventilation without abnormal alarms.

Do not return the machine to service based only on a single successful breath or brief test.

If the Problem Persists

If the bellows still does not rise or the ventilator does not cycle after gas supplies, settings, breathing-circuit connections, removable components, flow sensors, scavenging, and system checkout have been evaluated, common external causes have been ruled out.

The remaining problem may involve an internal leak, drive-gas fault, ventilator valve, control electronics, internal pneumatic component, or bellows-base assembly.

The device should be:

Do not proceed with internal pneumatic or electronic disassembly without appropriate training, authorization, test equipment, and current service documentation.

Knowing when to stop and escalate is proper troubleshooting.

Clinical Use Tip

Never attempt prolonged bellows or ventilator troubleshooting while the machine is connected to an active patient. Establish alternate ventilation first and troubleshoot only when patient care is safely supported.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Anesthesia staff reported that the Carestation 600 Series bellows did not rise and the ventilator would not cycle in mechanical ventilation mode."

Cause

What was observed during troubleshooting.

Example:
"Inspection found the expiratory flow sensor incorrectly seated after breathing-system cleaning, creating a significant system leak."

Resolution

What action was taken.

Example:
"Correctly installed the flow sensor, completed the full system checkout, verified stable bellows movement and ventilation with a test lung, and returned the machine to service."

Helpful Details to Include (If Known)

Final Thought

Bellows and ventilator problems require immediate attention because they directly affect ventilation. Protect the patient first, verify external gas, circuit, accessory, and assembly conditions logically, and escalate when reliable operation cannot be confirmed. Accurate CCR documentation preserves the failure history and supports safe repair decisions.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide