GE Healthcare Carestation 600 Series

O2 Sensor Calibration or FiO2 Measurement Error

On this page

Asset Type

Anesthesia Machine

Manufacturer

GE Healthcare

Model

Carestation 600 Series

What This Guide Helps With

Troubleshooting failed oxygen-sensor calibration, missing or unstable FiO2 readings, and measurements inconsistent with the selected gas mixture or independent analyzer.

Step-by-Step Troubleshooting

1. Ensure Patient Safety First

Do not calibrate, remove, or troubleshoot the oxygen-measurement system while the Carestation 600 Series is supporting an active patient.

If the displayed FiO2 is missing, unstable, or inconsistent with the delivered gas mixture:

Expected outcome: Patient care no longer depends on equipment with questionable oxygen measurement.

Continue troubleshooting only after the machine has been removed from active clinical use.

2. Confirm the Exact Failure Condition

Record the complete displayed message and determine whether the problem involves:

Expected outcome: The failure is clearly defined before components are disturbed.

3. Confirm the Displayed Value Is Measured FiO2

Verify that staff are comparing the correct parameters.

Do not confuse:

A difference between the selected gas mixture and measured inspired oxygen can occur because of circuit volume, fresh-gas flow, patient uptake, leaks, sampling delay, or residual gas.

Expected outcome: The reported error is confirmed as an actual oxygen-measurement problem rather than a misunderstanding of displayed values.

4. Inspect the Breathing Circuit and Sampling Setup

Check the external breathing system for:

A significant leak or improper circuit assembly can allow room air to enter the system and alter the measured oxygen concentration.

Expected outcome: The circuit is correctly assembled, dry, unobstructed, and leak-free.

If correcting a connection restores stable FiO2 measurement, complete the machine checkout and stop troubleshooting.

5. Check the Airway Gas Sampling Components

When FiO2 is measured through an installed airway gas module, inspect:

Replace a visibly contaminated, wet, obstructed, or damaged disposable sampling component with an approved compatible component.

Expected outcome: The gas module receives a continuous, unobstructed respiratory-gas sample.

If the reading becomes stable after replacing an external sampling component, perform checkout and document the correction.

6. Verify Oxygen and Balance-Gas Supplies

Confirm that:

Use the facility’s approved gas-pressure test equipment when verification is necessary.

Expected outcome: The machine has stable and correctly connected gas supplies.

If the oxygen concentration changes unexpectedly with a specific pipeline outlet, hose, or cylinder, remove that supply component from use and investigate it separately.

7. Allow the System and Sensor to Stabilize

Power the machine on in a suitable environment and allow the oxygen-measurement system to stabilize.

Ensure that:

Expected outcome: Environmental conditions and residual gas no longer interfere with calibration.

8. Expose the Oxygen Sensor to the Correct Calibration Gas

Follow the on-screen instructions for oxygen-sensor calibration.

For a room-air calibration:

The Carestation 600 Series documentation identifies weekly oxygen-cell calibration as calibration to approximately 21% oxygen through exposure to room air.

Expected outcome: The machine recognizes a stable room-air oxygen concentration and completes calibration.

If calibration succeeds, continue to Step 11 for verification.

9. Repeat Calibration Once Under Controlled Conditions

If the first attempt fails:

Do not repeatedly recalibrate a sensor that continues to fail. Repeated attempts can delay identification of a deteriorated sensor or internal measurement fault.

Expected outcome: A setup-related calibration failure is corrected without unnecessary component replacement.

10. Inspect Accessible Sensor Components

With the machine powered down and removed from service, inspect only externally accessible oxygen-sensor components permitted by facility procedure.

Check for:

Do not clean, soak, disinfect, or open an oxygen cell unless specifically permitted by GE Healthcare documentation.

Expected outcome: The sensor and its accessible connection are dry, intact, and correctly installed.

If an approved compatible sensor is available, replacement should be performed only by qualified personnel following the applicable service procedure.

11. Compare the Reading With a Calibrated Independent Analyzer

After successful calibration:

Expected outcome: The Carestation reading is stable and agrees with the independent analyzer within the manufacturer’s applicable tolerance.

Do not return the machine to service based only on a completed calibration message if the measured value remains inaccurate or unstable.

12. Run the Complete Preoperative Checkout

Perform the full machine checkout rather than only repeating the oxygen calibration.

Verify:

Expected outcome: The complete anesthesia system passes checkout without recurring oxygen-measurement faults.

If the checkout passes and independent testing confirms accurate measurement, return the device to service according to facility policy.

If the Problem Persists

If oxygen calibration continues to fail, FiO2 remains inaccurate, or the measurement is unstable after the gas supplies, circuit, sampling accessories, setup, environment, and accessible sensor connections have been checked, the common external causes have been ruled out.

The likely causes may include a depleted oxygen cell, failed airway gas module, damaged sensor interface, internal pneumatic problem, or internal electronic measurement fault.

The device should be:

Do not adjust internal calibration values, enter unsupported service settings, or perform board-level repair without the appropriate GE Healthcare service documentation, training, and test equipment.

Knowing when to stop and escalate is proper troubleshooting.

Clinical Use Tip

Never rely on a questionable FiO2 value during patient care. Move the patient to another verified device or establish approved independent oxygen monitoring before troubleshooting the affected machine.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Anesthesia staff reported that the Carestation 650 failed O2 sensor calibration and displayed an unstable inspired-oxygen reading."

Cause

What was observed during troubleshooting.

Example:
"Inspection found a moisture-contaminated sampling line and partially filled water trap restricting the airway gas sample."

Resolution

What action was taken.

Example:
"Replaced the sampling line and water trap, completed O2 calibration and full system checkout, and verified FiO2 against a calibrated independent oxygen analyzer."

Helpful Details to Include (If Known)

Final Thought

Reliable oxygen measurement is essential to anesthesia safety. Begin with patient protection, confirm the gas supplies and external sampling path, calibrate only under controlled conditions, verify the result independently, and escalate persistent failures. Complete CCR documentation preserves both the troubleshooting logic and the final safety decision.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide