On this page
Asset Type
Manufacturer
Model
What This Guide Helps With
Troubleshooting failed oxygen-sensor calibration, missing or unstable FiO2 readings, and measurements inconsistent with the selected gas mixture or independent analyzer.
Step-by-Step Troubleshooting
1. Ensure Patient Safety First
Do not calibrate, remove, or troubleshoot the oxygen-measurement system while the Carestation 600 Series is supporting an active patient.
If the displayed FiO2 is missing, unstable, or inconsistent with the delivered gas mixture:
- Notify the anesthesia provider immediately.
- Establish approved independent oxygen monitoring.
- Transfer the patient to another verified anesthesia machine when reliable oxygen measurement cannot be confirmed.
- Follow the facility’s anesthesia-machine failure procedure.
Expected outcome: Patient care no longer depends on equipment with questionable oxygen measurement.
Continue troubleshooting only after the machine has been removed from active clinical use.
2. Confirm the Exact Failure Condition
Record the complete displayed message and determine whether the problem involves:
- O2 sensor calibration failure
- Calibration required
- FiO2 reading higher or lower than expected
- FiO2 reading that drifts or fluctuates
- O2 measurement unavailable
- Measured O2 greater than the machine’s expected range
- Failure occurring only during checkout
- Disagreement between the machine and an independent oxygen analyzer
Expected outcome: The failure is clearly defined before components are disturbed.
3. Confirm the Displayed Value Is Measured FiO2
Verify that staff are comparing the correct parameters.
Do not confuse:
- Selected fresh-gas oxygen concentration
- Total oxygen flow
- ecoFLOW information
- Inspired oxygen measured from the breathing system or airway gas module
A difference between the selected gas mixture and measured inspired oxygen can occur because of circuit volume, fresh-gas flow, patient uptake, leaks, sampling delay, or residual gas.
Expected outcome: The reported error is confirmed as an actual oxygen-measurement problem rather than a misunderstanding of displayed values.
4. Inspect the Breathing Circuit and Sampling Setup
Check the external breathing system for:
- Loose or disconnected inspiratory and expiratory limbs
- Incorrectly assembled circuit components
- Open ports or missing caps
- Loose Y-piece connections
- Excessive moisture
- Obstructed filters
- Damaged tubing
- Incorrectly installed absorber canister
- Leaks around the breathing-system assembly
A significant leak or improper circuit assembly can allow room air to enter the system and alter the measured oxygen concentration.
Expected outcome: The circuit is correctly assembled, dry, unobstructed, and leak-free.
If correcting a connection restores stable FiO2 measurement, complete the machine checkout and stop troubleshooting.
5. Check the Airway Gas Sampling Components
When FiO2 is measured through an installed airway gas module, inspect:
- Sampling line connection at both ends
- Sampling line for kinks, cracks, blockage, or moisture
- Water trap installation and fill level
- Water trap seating
- Sample-gas exhaust connection
- Gas module seating and status indicators
- Compatibility of the sampling accessories
Replace a visibly contaminated, wet, obstructed, or damaged disposable sampling component with an approved compatible component.
Expected outcome: The gas module receives a continuous, unobstructed respiratory-gas sample.
If the reading becomes stable after replacing an external sampling component, perform checkout and document the correction.
6. Verify Oxygen and Balance-Gas Supplies
Confirm that:
- The oxygen pipeline hose is fully connected.
- Pipeline pressure is within the facility’s acceptable range.
- The oxygen cylinder is adequately filled when used.
- Air and nitrous oxide supplies are correctly connected.
- No supply hoses are crossed.
- Supply-pressure alarms are absent.
- The selected balance gas matches the connected gas supplies.
Use the facility’s approved gas-pressure test equipment when verification is necessary.
Expected outcome: The machine has stable and correctly connected gas supplies.
If the oxygen concentration changes unexpectedly with a specific pipeline outlet, hose, or cylinder, remove that supply component from use and investigate it separately.
7. Allow the System and Sensor to Stabilize
Power the machine on in a suitable environment and allow the oxygen-measurement system to stabilize.
Ensure that:
- The machine is at normal room temperature.
- It has not just been moved from a very cold or hot environment.
- Condensation is not present.
- The breathing system has been exposed to the calibration gas long enough to reach a stable concentration.
Expected outcome: Environmental conditions and residual gas no longer interfere with calibration.
8. Expose the Oxygen Sensor to the Correct Calibration Gas
Follow the on-screen instructions for oxygen-sensor calibration.
For a room-air calibration:
- Disconnect the breathing circuit or configure the system exactly as directed by the machine.
- Ensure the sensor is exposed to uncontaminated room air.
- Keep supplemental oxygen, open oxygen cylinders, oxygen flush flow, and other enriched gas sources away from the sensor.
- Allow residual oxygen-rich gas to dissipate before starting calibration.
The Carestation 600 Series documentation identifies weekly oxygen-cell calibration as calibration to approximately 21% oxygen through exposure to room air.
Expected outcome: The machine recognizes a stable room-air oxygen concentration and completes calibration.
If calibration succeeds, continue to Step 11 for verification.
9. Repeat Calibration Once Under Controlled Conditions
If the first attempt fails:
- Cancel the calibration.
- Reconfirm that the sensor is exposed to room air.
- Remove nearby sources of supplemental oxygen.
- Confirm that the sampling path is dry and unobstructed.
- Allow additional stabilization time.
- Repeat the calibration once.
Do not repeatedly recalibrate a sensor that continues to fail. Repeated attempts can delay identification of a deteriorated sensor or internal measurement fault.
Expected outcome: A setup-related calibration failure is corrected without unnecessary component replacement.
10. Inspect Accessible Sensor Components
With the machine powered down and removed from service, inspect only externally accessible oxygen-sensor components permitted by facility procedure.
Check for:
- Loose sensor connection
- Sensor not fully seated
- Damaged connector contacts
- Cracked housing
- Moisture or contamination
- Evidence of incorrect installation
- Sensor age beyond the facility’s expected replacement interval
Do not clean, soak, disinfect, or open an oxygen cell unless specifically permitted by GE Healthcare documentation.
Expected outcome: The sensor and its accessible connection are dry, intact, and correctly installed.
If an approved compatible sensor is available, replacement should be performed only by qualified personnel following the applicable service procedure.
11. Compare the Reading With a Calibrated Independent Analyzer
After successful calibration:
- Configure a verified breathing circuit and test lung.
- Deliver a known oxygen concentration.
- Allow the circuit concentration to stabilize.
- Compare the Carestation FiO2 reading with a calibrated independent oxygen analyzer.
- Repeat at more than one oxygen concentration when required by facility procedure.
Expected outcome: The Carestation reading is stable and agrees with the independent analyzer within the manufacturer’s applicable tolerance.
Do not return the machine to service based only on a completed calibration message if the measured value remains inaccurate or unstable.
12. Run the Complete Preoperative Checkout
Perform the full machine checkout rather than only repeating the oxygen calibration.
Verify:
- Oxygen calibration passes
- Gas supplies are recognized
- Circuit leak testing passes
- Flow and pressure checks pass
- Gas monitoring is available
- FiO2 responds appropriately to changes in gas concentration
- No oxygen-related alarms or advisories remain
- Alarm limits can be set and tested
Expected outcome: The complete anesthesia system passes checkout without recurring oxygen-measurement faults.
If the checkout passes and independent testing confirms accurate measurement, return the device to service according to facility policy.
If the Problem Persists
If oxygen calibration continues to fail, FiO2 remains inaccurate, or the measurement is unstable after the gas supplies, circuit, sampling accessories, setup, environment, and accessible sensor connections have been checked, the common external causes have been ruled out.
The likely causes may include a depleted oxygen cell, failed airway gas module, damaged sensor interface, internal pneumatic problem, or internal electronic measurement fault.
The device should be:
- Removed from service
- Labeled Out of Service
- Sent for qualified bench evaluation or manufacturer-authorized repair
Do not adjust internal calibration values, enter unsupported service settings, or perform board-level repair without the appropriate GE Healthcare service documentation, training, and test equipment.
Knowing when to stop and escalate is proper troubleshooting.
Clinical Use Tip
Never rely on a questionable FiO2 value during patient care. Move the patient to another verified device or establish approved independent oxygen monitoring before troubleshooting the affected machine.
Work Order Documentation (CCR Method)
CCR = Complaint, Cause, Resolution
Complaint
What was reported by the clinical staff.
Example:
"Anesthesia staff reported that the Carestation 650 failed O2 sensor calibration and displayed an unstable inspired-oxygen reading."
Cause
What was observed during troubleshooting.
Example:
"Inspection found a moisture-contaminated sampling line and partially filled water trap restricting the airway gas sample."
Resolution
What action was taken.
Example:
"Replaced the sampling line and water trap, completed O2 calibration and full system checkout, and verified FiO2 against a calibrated independent oxygen analyzer."
Helpful Details to Include (If Known)
- Exact alarm or advisory message
- Whether the failure occurred during startup, checkout, or clinical use
- O2 pipeline pressure verified
- Cylinder supply tested
- Balance gas selected
- Breathing circuit inspected
- Leak-test result
- Sampling line condition
- Water trap condition
- Gas module status
- Sensor connection inspected
- Room-air calibration attempted
- Independent analyzer reading
- Delivered test-gas concentration
- Indicator lights or module messages
- Presence of moisture or contamination
- Final checkout result
- Final device status
Final Thought
Reliable oxygen measurement is essential to anesthesia safety. Begin with patient protection, confirm the gas supplies and external sampling path, calibrate only under controlled conditions, verify the result independently, and escalate persistent failures. Complete CCR documentation preserves both the troubleshooting logic and the final safety decision.
That is successful troubleshooting.