GE Healthcare Carestation 600 Series

PEEP High or Elevated Expiratory Pressure

On this page

Asset Type

Anesthesia Machine

Manufacturer

GE Healthcare

Model

Carestation 600 Series

What This Guide Helps With

Troubleshooting elevated PEEP or incomplete exhalation caused by circuit restrictions, valve problems, scavenging issues, moisture, accessories, or incorrect ventilation settings.

Step-by-Step Troubleshooting

1. Ensure Patient Safety First

Do not perform extended troubleshooting while the Carestation 600 Series is supporting an active patient.

If expiratory pressure remains elevated during a procedure:

Expected: Patient ventilation no longer depends on equipment with unresolved elevated expiratory pressure.

Once the patient is safely supported by alternate equipment, continue troubleshooting in a controlled Clinical Engineering environment.

2. Confirm the Reported Condition

Review the alarm history and determine whether the issue involves:

Record the selected PEEP, measured PEEP, peak pressure, ventilation mode, breathing rate, and circuit configuration.

Expected: The reported condition can be reproduced and distinguished from a temporary clinical event or incorrect setting.

If the pressure returns to the selected baseline and the problem cannot be reproduced, complete the full system checkout before returning the device to service.

3. Verify Ventilator and Fresh-Gas Settings

Confirm that:

Use a manufacturer-compatible test lung rather than evaluating the condition on a patient.

Expected: Expiratory pressure returns to the selected PEEP level between breaths.

If correcting the settings resolves the issue, run the required checkout and stop troubleshooting.

4. Inspect the Patient Breathing Circuit

Trace the entire breathing circuit and check for:

Remove unnecessary accessories and test with a known-good circuit and test lung.

Expected: Gas flows freely through the inspiratory and expiratory limbs, and pressure returns normally during exhalation.

If replacing the circuit or accessory resolves the problem, document the failed component and stop.

5. Check the Expiratory Path

Inspect the externally accessible expiratory components for:

Do not force components into position. Follow facility infection-control procedures when handling contaminated breathing-system parts.

Expected: The expiratory path is correctly assembled, unobstructed, dry, and securely seated.

If reseating or replacing an approved external component resolves the issue, complete the system checkout and stop.

6. Inspect the Absorber and Breathing-System Assembly

Confirm that:

Use a known-good absorber assembly when available.

Expected: The breathing system passes checkout and allows unrestricted exhalation.

If the problem follows a particular absorber or external assembly, remove that component from service.

7. Evaluate the Scavenging System

Disconnect the machine from active patient use before changing scavenging connections.

Inspect the anesthetic-gas scavenging system for:

Temporarily evaluate the machine using the approved facility method for isolating the external scavenging connection. Do not vent anesthetic gas into an occupied workspace.

Expected: Expiratory pressure remains stable when connected to a properly functioning scavenging system.

If the elevated pressure disappears when the facility scavenging connection is isolated, refer the terminal, hose, or disposal system for correction.

8. Check the Manual Ventilation Configuration

Place the machine in the appropriate nonclinical test configuration and inspect:

Confirm that the adjustable pressure-limiting valve is not incorrectly set during manual ventilation testing.

Expected: Pressure releases normally through the intended pathway in both manual and mechanical test configurations.

If the problem occurs only in one configuration, document that distinction for bench evaluation.

9. Compare With Known-Good External Components

Using compatible and approved parts, substitute one component at a time:

Avoid changing multiple components simultaneously because doing so may prevent identification of the actual cause.

Expected: The faulty or restrictive external component is identified.

If a substituted component resolves the problem, remove the suspect component from service and complete the required functional testing.

10. Restart and Perform the Complete System Checkout

With the machine removed from patient use:

Do not return the machine to service solely because the alarm cleared after restarting.

Expected: The machine passes checkout without pressure, flow, valve, leak, or ventilation failures.

If the issue is resolved and all required tests pass consistently, document the findings and return the device according to facility policy.

If the Problem Persists

If elevated expiratory pressure continues after settings, circuits, accessories, breathing-system assembly, absorber components, and scavenging connections have been ruled out, the cause is likely internal.

Possible internal areas may include the expiratory valve mechanism, ventilator control system, internal pneumatic pathways, pressure sensing, or electronic control components.

The anesthesia machine should be:

Do not continue repeated clinical testing or perform unauthorized internal adjustments. Knowing when to stop and escalate is proper troubleshooting.

Clinical Use Tip

Never troubleshoot persistent elevated PEEP on an active patient. Move the patient to verified equipment first, then reproduce the issue with an approved test lung in a controlled environment.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Anesthesia staff reported that airway pressure remained above the selected PEEP and did not return to baseline during mechanical ventilation."

Cause

What was observed during troubleshooting.

Example:
"A water-filled filter in the expiratory circuit restricted exhaled gas flow and caused elevated expiratory pressure."

Resolution

What action was taken.

Example:
"Replaced the restricted filter, installed a known-good breathing circuit, completed the system checkout, and verified that measured PEEP returned to the selected value."

Helpful Details to Include (If Known)

Final Thought

Elevated expiratory pressure can create a serious ventilation hazard. Begin with patient safety, then evaluate settings, circuit restrictions, breathing-system components, and scavenging before suspecting an internal failure. Careful testing and complete CCR documentation support both safe escalation and reliable repair.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide