What This Guide Helps With

High pressure or high PEEP alarms occur because of circuit obstruction, water, filters, valves, settings, patient factors, or breathing-system problems.

Step-by-Step Troubleshooting

1. Protect the Patient and Address Ventilation Immediately

If a high airway pressure or PEEP high alarm occurs during patient care, clinical staff must assess the patient and airway first.

Do not troubleshoot the anesthesia machine while the patient depends on unreliable ventilation. Provide alternate ventilation if needed.

Expected outcome: Patient ventilation is stabilized before equipment troubleshooting begins.

2. Confirm the Alarm Under Controlled Conditions

After removing the machine from patient use, connect an appropriate test lung and attempt to reproduce the condition.

Record:

Expected outcome: The equipment-related condition is either reproduced or separated from patient-specific factors.

3. Inspect the Patient Circuit for Obstruction

Check the entire external breathing circuit for:

Expected outcome: The circuit is open and unobstructed. If removing an obstruction restores normal pressure, troubleshooting can stop after verification.

4. Inspect Filters and Accessories

Remove or replace questionable filters, adapters, humidification components, sampling adapters, or other added accessories using compatible known-good components.

Do not omit clinically required accessories when returning the machine to service.

Expected outcome: No external accessory creates excessive resistance. If pressure normalizes after replacement, the external cause is confirmed.

5. Verify Ventilator Settings

Review the configured:

Do not change clinically prescribed settings on an active patient as an equipment troubleshooting method.

Expected outcome: The test settings are appropriate and do not inherently explain the alarm. Incorrect test settings can be corrected and reassessed.

6. Check the Breathing-System Assembly

Verify externally removable breathing-system components are properly seated and assembled.

Inspect accessible expiratory-path components for obvious contamination, improper installation, or obstruction without deep disassembly.

Expected outcome: The breathing system is correctly assembled with no visible restriction.

7. Compare With a Known-Good Circuit and Test Lung

Replace the external breathing circuit and test lung with known-good compatible components.

Expected outcome: If the alarm disappears, the original circuit or accessory was responsible. If high pressure or PEEP persists, continue troubleshooting.

8. Observe Expiratory Pressure Return

During controlled ventilation, observe whether airway pressure returns appropriately during expiration.

Persistent elevated expiratory pressure with a known-good external setup may indicate a breathing-system, valve, sensor, or control problem requiring escalation.

Expected outcome: Pressure returns appropriately during expiration. If not, remove the machine from service.

9. Complete Final Functional Verification

After correcting an external cause, verify:

Expected outcome: The original alarm does not recur under controlled testing and the machine passes checkout. Troubleshooting can stop.

10. Escalate Persistent High Pressure or High PEEP

If the alarm persists with known-good circuits, accessories, and appropriate settings, stop external troubleshooting.

Expected outcome: The Carestation 750 is removed from service for qualified service evaluation.

If the Problem Persists

Common external causes have been ruled out. Remaining categories may include expiratory-path restriction, internal valve function, pressure measurement, ventilator control, breathing-system interfaces, or other service-level faults.

The Carestation 750 should be:

After repair, airway pressure, PEEP, ventilation, alarms, and breathing-system performance must be verified before return to service.

Knowing when persistent elevated pressure requires escalation protects the patient and the equipment.

Clinical Use Tip

During an actual high-pressure alarm, patient, airway, and circuit causes must be evaluated clinically before the machine itself is assumed to be defective.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Anesthesia staff reported repeated high airway pressure alarms from the Carestation 750 during pre-use testing."

Cause

What was observed during troubleshooting.

Example:
"Clinical Engineering found a kinked breathing hose creating an external obstruction."

Resolution

What action was taken.

Example:
"The breathing circuit was replaced, airway pressure returned to normal during test-lung ventilation, alarms were verified, and checkout passed."

Helpful Details to Include (If Known)

Final Thought

High airway pressure should be approached from the patient and external circuit inward. Confirm obstruction, accessories, settings, and breathing-system assembly before considering an internal fault, then verify stable ventilation and alarm behavior before return to service.

That is successful troubleshooting.