GE Healthcare Discovery MI

Cooling, Fan, or Overtemperature Fault Occurs

On this page

Asset Type

PET / CT System

Manufacturer

GE Healthcare

Model

Discovery MI

What This Guide Helps With

Troubleshoots cooling, airflow, fan, and temperature faults caused by blocked ventilation, room conditions, facility cooling, contamination, or service-level cooling problems.

Step-by-Step Troubleshooting

1. Protect the Patient and Respond to Overtemperature Safely

If the scanner reports an overtemperature condition, produces unusual heat, smells hot, or has abnormal fan noise, stop clinical operation and safely remove the patient when appropriate.

If smoke, burning odor, visible electrical damage, or excessive heat is present, isolate the equipment according to emergency procedures and do not continue troubleshooting under power.

Expected outcome: The patient is safe and no overheating equipment remains in clinical use.

2. Confirm the Exact Cooling Condition

Determine whether the complaint involves:

Record when the fault appears and whether it clears after the system is idle.

Expected outcome: The cooling-related symptom is clearly identified.

3. Check Room Temperature and HVAC Operation

Assess the imaging room and nearby equipment area for obvious environmental problems.

Check for:

Coordinate with Facilities when room cooling is abnormal.

Expected outcome: The room environment is appropriate or an external HVAC problem is identified.

If restoring room cooling resolves the scanner condition and required verification passes, troubleshooting can stop.

4. Inspect Accessible Scanner Ventilation

Inspect externally accessible air inlets, outlets, and surrounding clearance.

Look for:

Do not remove internal covers solely to inspect fans.

Expected outcome: External ventilation paths are unobstructed.

5. Listen for Abnormal Fan Operation

From outside the equipment, note whether cooling fans sound:

Do not reach into vents or moving equipment.

Expected outcome: No obvious abnormal mechanical fan condition is present.

If grinding, binding, or other abnormal mechanical noise is identified, stop and escalate.

6. Check for Recent Environmental or Power Events

Determine whether the fault followed:

Expected outcome: Any external event contributing to the cooling problem is identified or ruled out.

7. Allow Approved Cool-Down if Appropriate

If the manufacturer-approved workflow permits and no hazardous condition exists, allow the system to remain idle or shut down normally while the environmental issue is corrected.

Do not repeatedly restart an overheating system.

Expected outcome: The system returns to a normal temperature state without recurring warning after the external cause is corrected.

8. Verify Normal Operation

After correcting an external cooling condition, confirm:

Expected outcome: The system remains thermally stable during verification.

If applicable return-to-service testing passes, troubleshooting can stop and the scanner may be returned to service.

If the Problem Persists

If room conditions, ventilation, external obstruction, recent power or HVAC events, and approved cool-down procedures have been addressed but a cooling or overtemperature fault persists, the cause may involve an internal fan, cooling circuit, temperature sensor, power supply, heat exchanger, detector cooling requirement, or another service-level condition.

The system should be:

Do not bypass thermal protection or operate the scanner repeatedly to determine how long it runs before overheating.

After service, verify stable operating temperature and all affected scanner functions before return to clinical use.

Knowing when to stop external troubleshooting and escalate is proper troubleshooting.

Clinical Use Tip

An overtemperature warning should never be treated as a nuisance alarm; unreliable thermal control can affect scanner availability and image acquisition reliability.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Imaging staff reported that the Discovery MI displayed an overtemperature warning and became unavailable for scanning."

Cause

What was observed during troubleshooting.

Example:
"Clinical Engineering found equipment stored against an external ventilation opening and restricting airflow."

Resolution

What action was taken.

Example:
"Clinical Engineering cleared the ventilation area, allowed the system to return to normal operating condition, verified that the warning did not recur during functional testing, and returned the scanner to service."

Helpful Details to Include (If Known)

Final Thought

Cooling problems should begin with patient safety and the environment. Verify room HVAC, external airflow, ventilation clearance, and obvious fan behavior before assuming an internal cooling failure. Do not bypass thermal protections, and escalate recurring faults appropriately.

That is successful troubleshooting.