GE Healthcare Discovery MI

DICOM, PACS, or Worklist Transfer Fails

On this page

Asset Type

PET / CT System

Manufacturer

GE Healthcare

Model

Discovery MI

What This Guide Helps With

Troubleshoots worklist and image-transfer failures caused by connectivity, destination availability, patient data, queue status, network infrastructure, or configuration.

Step-by-Step Troubleshooting

1. Protect the Patient and Preserve Imaging Data

A DICOM or PACS failure may not require stopping acquisition immediately if local imaging remains reliable, but do not delete studies, recreate examinations unnecessarily, or repeat scans because images have not transferred.

Follow department downtime procedures when required.

Expected outcome: Patient imaging data are preserved and no unnecessary repeat exposure or acquisition occurs.

2. Identify Which Function Is Failing

Determine whether the issue involves:

Record any message or queue status.

Expected outcome: The affected DICOM workflow is clearly identified.

3. Confirm Local Scanner Operation

Verify that the Discovery MI can:

Expected outcome: Local acquisition and storage either function normally or reveal that the issue is broader than DICOM transfer.

4. Check External Network Connectivity

Inspect accessible network connections for:

Check whether other networked imaging equipment is experiencing similar problems.

Expected outcome: Physical network connectivity is intact or an external network problem is identified.

5. Verify Destination Availability

Coordinate with PACS, IT, or imaging informatics staff to determine whether the destination service is available.

Check for known:

Expected outcome: The remote destination is confirmed available or the infrastructure outage is identified.

If the destination is unavailable, do not make unnecessary scanner configuration changes. Follow downtime procedures.

6. Verify Patient and Study Information

Review the affected examination for obvious demographic or workflow problems such as incomplete patient identification or an examination that was manually created differently from normal workflow.

Do not alter patient identifiers merely to force transmission.

Expected outcome: Patient and study information are complete and consistent with department workflow.

7. Review the Transfer Queue or Status

Use normal operator-accessible functions to determine whether images are:

Avoid deleting queued studies unless the data have been safely preserved and department procedures specifically permit it.

Expected outcome: The transfer status identifies whether the scanner is attempting communication and where the process stops.

8. Compare With Another Study or Destination

When appropriate and without exposing a patient, determine whether:

Expected outcome: The issue is narrowed to the scanner, a particular DICOM service, or broader infrastructure.

9. Perform an Approved Restart if Appropriate

If local policy permits and patient data are protected, restart the affected workstation or communication function using approved procedures.

Do not alter DICOM network parameters without authorization and verified configuration information.

Expected outcome: Network services reconnect and the expected DICOM workflow resumes.

10. Verify End-to-End Transfer

Confirm the complete workflow:

Expected outcome: The full DICOM, PACS, or worklist path operates normally.

If the complete path is verified, troubleshooting can stop.

If the Problem Persists

If physical connections, local operation, destination availability, patient information, queue status, and approved restart procedures have been checked but DICOM communication still fails, the remaining cause may involve network infrastructure, server services, DICOM configuration, interface routing, workstation software, firewall or security controls, or another service-level condition.

The system should be:

Coordinate with PACS, IT, and imaging informatics as needed.

After correction, verify the complete transfer path before restoring normal workflow.

Knowing when to stop external troubleshooting and escalate is proper troubleshooting.

Clinical Use Tip

Never repeat a PET or CT acquisition simply because images failed to reach PACS; verify that the original study is preserved and troubleshoot the communication path first.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"PET / CT staff reported that completed Discovery MI studies remained in the send queue and were not appearing in PACS."

Cause

What was observed during troubleshooting.

Example:
"Clinical Engineering found that the scanner's accessible network connection had been disconnected during nearby equipment work."

Resolution

What action was taken.

Example:
"Clinical Engineering restored the network connection, verified queued studies transmitted successfully to PACS, confirmed receipt with imaging staff, and returned the system to normal workflow."

Helpful Details to Include (If Known)

Final Thought

Treat DICOM failures as an end-to-end communication problem. Preserve patient data, confirm local operation, verify physical network connectivity, remote services, study information, and queue status, and involve IT or PACS support when the fault is outside the scanner.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide