GE Healthcare Discovery MI

Exposure or Scan Will Not Start or Is Aborted

On this page

Asset Type

PET / CT System

Manufacturer

GE Healthcare

Model

Discovery MI

What This Guide Helps With

Troubleshoots scans that will not begin or abort unexpectedly because of readiness, interlocks, patient setup, acquisition settings, connections, or infrastructure.

Step-by-Step Troubleshooting

1. Protect the Patient and Stop Repeated Scan Attempts

Do not repeatedly attempt CT exposure or PET / CT acquisition while the reason for an aborted or inhibited scan is unknown.

Safely pause the examination, maintain patient communication, and move the patient to an alternate verified system if timely continuation cannot be assured.

Expected outcome: The patient is protected from unnecessary repeat attempts and the scanner is available for controlled troubleshooting.

2. Confirm When the Scan Fails

Determine whether the scan:

Record any displayed message and the point in the workflow where the failure occurs.

Expected outcome: The exact failure point is known.

3. Verify Overall System Readiness

Confirm that the system has completed startup and that required PET and CT subsystems indicate ready.

Check for unresolved:

Expected outcome: The scanner is fully initialized and no general readiness condition is blocking acquisition.

4. Check Patient Positioning and Travel Path

Verify that the patient, table, cables, monitoring equipment, injector tubing, positioning aids, and other accessories are arranged so they cannot interfere with table or gantry operation.

Make sure no object is contacting the gantry or obstructing travel.

Expected outcome: The patient setup permits safe, unobstructed positioning and acquisition.

5. Verify Safety and Exposure Controls

Check accessible emergency-stop, exposure-control, and normal safety-interlock status.

Do not bypass or defeat an interlock to make an exposure proceed.

Inspect external exposure-control hardware and accessible cables for damage or loose connections where applicable.

Expected outcome: Safety controls are in their normal operating state and accessible exposure-control components are intact.

6. Review Study and Protocol Setup

Confirm that the selected protocol and patient examination are complete and appropriate for the intended scan.

Check operator-accessible items such as:

Do not modify restricted protocol or service parameters unless authorized.

Expected outcome: The examination is correctly prepared and no incomplete setup is preventing acquisition.

7. Determine Whether the Failure Is Protocol-Specific

When safe and clinically appropriate, compare the behavior using an approved test workflow or another known-good standard configuration.

Do not expose a patient solely for troubleshooting.

Expected outcome: The technician can determine whether the fault affects the entire acquisition system or only a particular workflow.

If an operator-level setup issue is identified and corrected, proceed to final verification.

8. Perform an Approved Controlled Restart

If no unsafe condition exists and external checks are normal, perform an approved normal restart according to department and manufacturer procedures.

Avoid repeated restart attempts if the same abort recurs.

Expected outcome: The scanner initializes normally and acquisition readiness is restored.

9. Perform Final Functional Verification

Complete an approved nonclinical or quality-control verification as appropriate.

Confirm:

Expected outcome: The system completes a representative acquisition normally.

If required functional and safety testing passes, troubleshooting can stop and the scanner may be returned to service.

If the Problem Persists

If readiness, patient setup, safety controls, external exposure controls, protocol setup, connections, and approved restart procedures are normal but scanning still fails or aborts, the cause may involve internal acquisition electronics, safety interlocks, motion synchronization, generator or exposure systems, detector communication, PET acquisition hardware, workstation communication, or another service-level condition.

The system should be:

Do not bypass exposure interlocks or repeatedly generate exposures to reproduce a persistent fault.

After repair, perform all required radiation, acquisition, image-quality, functional, and safety verification before return to clinical use.

Knowing when to stop external troubleshooting and escalate is proper troubleshooting.

Clinical Use Tip

Avoid unnecessary repeat CT exposure while investigating scan aborts; move the patient to another verified imaging system when reliable acquisition cannot be promptly confirmed.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"PET / CT staff reported that the Discovery MI would prepare the examination but the scan would not begin."

Cause

What was observed during troubleshooting.

Example:
"Clinical Engineering found the external scan-control connection loose at the operator area."

Resolution

What action was taken.

Example:
"Clinical Engineering secured the connection, verified normal system readiness, completed an approved functional acquisition, and returned the scanner to service after required checks passed."

Helpful Details to Include (If Known)

Final Thought

A failed or aborted scan should be approached by protecting the patient, identifying exactly where the workflow stops, and checking readiness, positioning, interlocks, controls, and study setup before assuming internal acquisition failure. Verify the correction without unnecessary patient exposure and document the result clearly.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide