GE Healthcare MAC 5

Error: Copying Software Upgrade Files Unsuccessful

On this page

Asset Type

Electrocardiograph (EKG) Machine

Manufacturer

GE Healthcare

Model

MAC 5

What This Guide Helps With

This guide helps troubleshoot Copying Software Upgrade Files Unsuccessful on a GE Healthcare MAC 5 when the software image or digital signature on the USB flash drive does not have sufficient permissions to copy the files. This condition can stop a USB software upgrade before it can continue and may leave the device unavailable until the upgrade media issue is corrected.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Error and Behavior

Confirm the USB Flash Drive Is the Intended Upgrade Media

Check That the Required Upgrade Files Are Present

Check File Permissions on the USB Flash Drive

Test Copying the Files Manually on a Workstation

Verify the USB Drive Is Not Write-Protected

Reapply Proper File Permissions

Try a Known-Good USB Flash Drive

Restart the MAC 5 and Retry the Upgrade

Stop When Resolved

If the Problem Persists

If the error continues after verifying file permissions, confirming the software image and digital signature are present, testing a known-good USB flash drive, and restarting the device, external causes have been ruled out.

At that point, the issue may involve corrupted upgrade files, incompatible upgrade media, USB access problems, or an internal device software or storage issue.

Remove the MAC 5 from service, label it Out of Service, and send it for repair, bench evaluation, or vendor-supported software service.

Clinical Use Tip

Never troubleshoot software upgrade errors on an active patient. Move the patient to a backup ECG machine first and maintain testing continuity until the MAC 5 upgrade issue is resolved and the device passes functional checks.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"MAC 5 displayed Copying Software Upgrade Files Unsuccessful during USB software upgrade, and upgrade files would not copy from the USB flash drive."

Cause

What was observed during troubleshooting.

Example:
"Software image or digital signature files on the USB flash drive did not have sufficient permissions for the device to copy the upgrade files. External USB file permission issue suspected or confirmed."

Resolution

What action was taken.

Example:
"Removed device from patient use, verified USB upgrade contents, checked file access, corrected read/write permissions on the software image and digital signature files, and retried upgrade. Device returned to service after upgrade process continued successfully and functional checks passed."

Helpful Details to Include (If Known)

Final Thought

Software upgrade errors should be handled carefully because they can affect device availability and reliability. For Clinical Engineering, the correct approach is to confirm the displayed error, verify the USB upgrade files, check permissions, rule out bad media, and escalate when the device still cannot copy valid files. Clear CCR documentation helps show whether the issue was corrected externally or requires repair support.

That is successful troubleshooting.