GE Healthcare MAC 5

Error: Host Software Image Is Not Digitally Signed

On this page

Asset Type

Electrocardiograph (EKG) Machine

Manufacturer

GE Healthcare

Model

MAC 5

What This Guide Helps With

This guide helps troubleshoot the Host Software Image Is Not Digitally Signed error on the GE Healthcare MAC 5. This error prevents the software upgrade or installation from continuing because the host software image is not recognized as a valid digitally signed GE HealthCare software package.

The issue may be caused by an incorrect software image, incomplete software package, unofficial or unsupported file, corrupted file, renamed or modified file, wrong software package for the device, or software obtained from an unapproved source.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Error and Behavior

Translate the Error Into the Real-World Problem

Check the Upgrade Source

Confirm the Software Came From an Approved Source

Check for File Corruption or Incomplete Copy

Verify the File Location

Retry the Upgrade With the Correct Image

Do Not Attempt Workarounds

If the Problem Persists

If the error continues after verifying the USB drive or shared folder, file location, and software source, the software image should be considered invalid for this upgrade.

External causes such as incorrect media placement, wrong file location, incomplete copy, unsupported USB media, shared-folder access issue, and unapproved software source should be ruled out before escalation.

Remove the MAC 5 from service if the software upgrade is required before clinical use.

Label the device Out of Service if it cannot be returned safely to normal operation.

Send the device for repair or bench evaluation, and contact GE HealthCare Service Support to obtain the correct digitally signed host software image or confirm the correct upgrade package for the device.

Clinical Use Tip

Never troubleshoot software upgrade errors on a device being used for patient ECG acquisition.

Move the patient to a backup ECG machine first if testing is clinically needed.

Maintain therapy and testing continuity by using another verified ECG machine until the MAC 5 upgrade issue is resolved and the unit passes functional verification.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"GE MAC 5 displayed Host Software Image Is Not Digitally Signed during attempted host software upgrade. Upgrade would not continue."

Cause

What was observed during troubleshooting.

Example:
"Host software image failed digital signature validation. External media and file location checked; software package suspected to be unsigned, incorrect, corrupted, modified, or not from approved GE HealthCare source."

Resolution

What action was taken.

Example:
"Stopped upgrade attempt. Verified upgrade media and file placement. Removed device from service pending correct digitally signed host software image from GE HealthCare Service Support."

Helpful Details to Include (If Known)

Final Thought

This error should be treated as a software integrity protection message. Clinical Engineering should verify the external media, file location, and software source, but should not attempt to bypass digital signature validation. If the correct signed image is not available, escalate to GE HealthCare Service Support and document the failure clearly.

That is successful troubleshooting.